Abstract
AAPM CT performance for special medical equipment quality control checks using a standard phantom for evaluation, using the evaluator's subjective assessment as to minimize errors due computerized assessment program to evaluate their usefulness. Phantom for evaluation AAPM CT Performance Phantom: was used, the default shooting conditions are the same as quality control checks. And, we use IMAGE J to evaluate the program. Quantitative evaluation with CT attenuation coefficient and the noise measurement, the uniformity measurement, the slice thickness measurement, contrast resolution of the measurement, a phantom image of the spatial resolution determined by the evaluation program is evaluated as self-extracting the result after processing the image, CT uniformity measurement for the evaluation that was smaller and the standard deviation of a video image processing more uniform slice thickness measurements it is difficult to evaluate due to the difference of the ratio of the measured value of the phantom image. Contrast resolution was measured cylindrical diameter 6th evaluate the shape of a circle obtained a mean value and a standard deviation of diameters, the spatial resolution of the group of source, including acceptance criteria automatically extracted result as a result of both the number of the extracted circularIt appeared. Evaluate the source image and video processing, and video to qualitative evaluation by gross were processed video image is shown excellent results. If the evaluators in order to minimize the errors of subjective judgment based on the results of the above should be done with a quantitative evaluation and qualitative evaluation utilizes a computerized assessment program is considered that further evaluation be made more efficient.
AAPM CT 성능 평가용 표준 팬텀을 이용한 특수의료장비 품질관리 검사 시, 평가자의 주관적인 평가로 인한 오류를 최소화 하고자 전산화된 평가 프로그램을 이용하여 유용성을 평가하고자 한다. 평가 팬텀으로 AAPM CT Performance Phantom을 사용하였고, 기본 촬영 조건은 품질관리검사와 동일하며, 평가프로그램으로 IMAGE J를 사용하였다. 정량적인 평가로 CT 감약계수와 노이즈측정, 균일도측정, 슬라이스 두께 측정, 대조도 분해능 측정, 공간 분해능 측정의 팬텀 영상을 평가프로그램을 이용하여 영상처리를 한 후 자동추출 된 결과로 평가 하였으며, CT 감약계수, 노이즈, 균일도 측정은 영상처리를 한 영상의 표준편차가 작아 더 균일하다고 평가하였고, 슬라이스 두께 측정은 팬텀영상의 측정값과의 비율 차로 인해 평가에 어려움이 있었다. 대조도 분해능은 원통형의 지름을 6회 측정하여 지름의 평균값과 표준편차를 구해 원의 형태를 평가하였으며, 공간 분해능은 합격기준을 포함하는 원의 그룹을 자동 추출한 결과 원의 개수를 모두 추출한 결과로 나타났다. 정성적인 평가로 원본영상과 영상 처리한 영상을 육안적으로 비교 평가 하였는데 영상처리 된 영상이 우수한 결과를 나타내었다. 위의 결과 등을 바탕으로 평가자의 주관적인 판단의 오류를 최소화하기 위해서는 정량적인 평가와 정성적인 평가가 함께 이루어져야 하고 전산화된 평가프로그램을 활용한다면 보다 더 효율적인 평가가 이루어 질 것이라고 사료된다.