DOI QR코드

DOI QR Code

플레이트의 소성변형 과정이 재건술에서 플레이트 안정성에 미치는 영향

On the Stability of the Permanently Bent Mini-plate in Reconstructive Surgery

  • 박시명 (한국과학기술연구원 의공학연구소 바이오닉스연구단) ;
  • 이득희 (한국과학기술연구원 의공학연구소 바이오닉스연구단) ;
  • 노건우 (한국과학기술연구원 의공학연구소 바이오닉스연구단)
  • Park, Si Myung (Center for Bionics, Biomedical Research Institute, Korea Institute of Science and Technology(KIST)) ;
  • Lee, Deukhee (Center for Bionics, Biomedical Research Institute, Korea Institute of Science and Technology(KIST)) ;
  • Noh, Gunwoo (Center for Bionics, Biomedical Research Institute, Korea Institute of Science and Technology(KIST))
  • 투고 : 2016.02.12
  • 심사 : 2016.04.20
  • 발행 : 2016.09.01

초록

Conventional bent plate used in mandibular reconstruction surgery needs safety verification since its mechanical properties are changed due to the plastic deformation during the bending process. In this study we investigate stability of the plastically deformed plate and the plate with the same shape without plastic deformation through the finite element analysis(FEA). First we simulate the process of plate bending to fit the defect in patient. Then, the other plate is modelled to represent a customized plate with the same shape of the plastically deformed one, but without any residual stresses from plastic deformation. After binding these plates to the mandible, we conduct the masticatory simulation. Finally, we compare the resulting Von Mises stress of the customized plate and of the bent plate. The bent plate shows much higher stress than the customized one due to the residual stresses form the bending process. The study shows that plastic deformation in the plate may decrease the safety of the reconstruction surgery.

키워드

참고문헌

  1. Gardner, M.J., Brophy, R.H., et al., 2005, The Mechanical behavior of Locking Compression Plates Compared with Dynamic Compression Plates in A Cadaver Radius Model, J. Orthop Trauma., 19, pp.597-603. https://doi.org/10.1097/01.bot.0000174033.30054.5f
  2. Kornoer, J., Diederichs, G., et al., 2004, A Biomechanical Evaluation of Methods of Distal Humerus Fracture Fixation Using Locking Compression Plates Versus Conventional Reconstruction Plates, J. Orthop. Trauma., 18, pp.286-293. https://doi.org/10.1097/00005131-200405000-00004
  3. Kowalski, M.J., Schemitsch, E.H., et al., 1996, A Comparative Biomechanical Evaluation of a Noncontacting Plate and Currently Used Devices for Tibial Fixation, J. Trauma., 40, pp.5-9. https://doi.org/10.1097/00005373-199601000-00002
  4. Stoffel, K., Stachowiak, G., et al., 2012, Oblique Screws at the Plate Ends Increase the Fixation Strength in Synthetic Bone Test Medium, J. Orthop. Trauma., 18, pp.611-6.
  5. Aquilina, P., Chamoli U., et al., 2012, Finite Element Analysis of Three Patterns of Internal Fixation of Fractures of the Mandibular Condyle, British Journal of Oral and Maxillofacial Surgery, 3883, pp.6-11.
  6. Cook, R.D., Malkus, D.S., et al., 2002, Concepts and Applications of Finite Element Analysis, John Wiley & Sons. Inc.
  7. DASSAULT SYSTEMS SIMULIA, "http://www.3ds.com/".
  8. Mesnard, M., Ramos, A., et al., 2011, Biomechanical Analysis Comparing Natural and Alloplastic Temporomandibular Joint Replacement Using a Finite Element Model, J. Oral Maxillo. Fac. Sung., 1008e1017, pp.69. https://doi.org/10.1016/j.joms.2010.02.019
  9. Mesnard, M., Famos, A., et al., 2010, Muscular and Articular Forces Exerted on the Mandible, 1st International Conference on Biodental Engineering, 199e204.
  10. AOFoundation, "https://www2.aofoundation.org".
  11. Bathe, K.-J. and Gunwoo, N., 2012, Insight into an Implicit Time Integration Scheme for Structural Dynamics, Computers and Structures, 98-99, pp.1-6. https://doi.org/10.1016/j.compstruc.2012.01.009
  12. Gunwoo, N. and Bathe, K.-J., 2013, An Explicit Time Integration Scheme for the Analysis of Wave Propagations, Computers and Structures, 129, pp.178-193. https://doi.org/10.1016/j.compstruc.2013.06.007
  13. Zenner, H. and Renner, F., 2002, Cyclic Material behavior of Magnesium Die Castings and Extrusions, International Journal of Fatigue, 24, pp.1255-1260. https://doi.org/10.1016/S0142-1123(02)00042-7