Abstract

In bone density examinations, a change in the measured BMD occurs owing to the differences between the measured areas. To address this problem, we aimed to develop a new auxiliary device that could be rotated by $15^{\circ}$ by fixing the ankle to the distal femur neck. Dual energy X-ray absorptiometry (DXA) of BMD examinations were performed once a year, but 10 patients were examined over three sessions to analyze the area for measuring the femur neck BMD. The goal of this test was to determine the device's reliability, and the results were expressed in terms of the standard deviation of measurements. After performing bone density measurements using the new auxiliary device on 10 normal patients, with three measurements for each patient, the obtained standard deviation was 0.03. The standard deviation of the measured BMD was 0.19 when using the currently existing auxiliary device, while the standard deviation of the measured BMD was 0.03 when using the new auxiliary device. By using the new auxiliary device, the standard deviation could be reduced by ~80%. Accurate rotation of the femur neck was possible in all examinations, and the standard deviation of BMD measurements could be reduced by up to 80% compared with the measurements performed using the currently existing auxiliary device. We hope that this advantageous new design can be used as a standard auxiliary device for measuring the femur neck BMD.

Keywords

References

  1. W. J. Lee and N. W. Kim, Korea Society of Radiological Science 28, 227 (2005).
  2. D. Y. Kim, Korean J. Nucl. Med. 38, 275 (2004).
  3. B. L. Riggs and L. J. Melton, Am. J. Med. 75, 899 (1983). https://doi.org/10.1016/0002-9343(83)90860-4
  4. X. Cheng, J. Li, Y. Lu, J. Keyak, and T. Lang, Bone J. 40, 169 (2014).
  5. J. S. Choi, G. C. Ahn, and C. S. Lee, Jouranl of Korean Society of Spine Surgery 10, 75 (2003). https://doi.org/10.4184/jkss.2003.10.2.75
  6. Y. H. Kang and K. H. Jo, Korea Society of Radiological Science 29, 229 (2006).
  7. International Society for Clinical Densitometry (ISCD).
  8. A. Naimark, J. Kossoff, and A. Schepsis, AJR Am. J. Roentgenol. 133, 889 (1979). https://doi.org/10.2214/ajr.133.5.889
  9. H. F. Newton-John and D. B. Morgan, Clin. Orthop. Relat. Res. 71, 229 (1970).
  10. B. E. Nordin, Calcif. Tissue. Int. 40, 57 (1987). https://doi.org/10.1007/BF02555705
  11. Y. H. Lee, I. J. Lee, and H. J. Yang, Korea Society of Radiological Science 35, 1 (2012).
  12. J. C. H. Goh and S. L. Low, Calcif. Tissue. Int. 57, 340 (1995). https://doi.org/10.1007/BF00302069