DOI QR코드

DOI QR Code

Comparison of Transcatheter Atrial Septal Defect Closure in Children, Adolescents and Adults: Differences, Challenges and Short-, Mid- and Long-Term Results

  • Saritas, Turkay (Istanbul Medipol University, Medicine Faculty, Department of Pediatric Cardiology) ;
  • Yucel, Ilker Kemal (Dr. Siyami Ersek Thoracic and Cardiovascular Surgery Center and Research Hospital, Department of Pediatric Cardiology) ;
  • Demir, Ibrahim Halil (Dr. Siyami Ersek Thoracic and Cardiovascular Surgery Center and Research Hospital, Department of Pediatric Cardiology) ;
  • Demir, Fadli (Dr. Siyami Ersek Thoracic and Cardiovascular Surgery Center and Research Hospital, Department of Pediatric Cardiology) ;
  • Erdem, Abdullah (Istanbul Medipol University, Medicine Faculty, Department of Pediatric Cardiology) ;
  • Celebi, Ahmet (Dr. Siyami Ersek Thoracic and Cardiovascular Surgery Center and Research Hospital, Department of Pediatric Cardiology)
  • Received : 2015.08.27
  • Accepted : 2016.01.12
  • Published : 2016.11.30

Abstract

Background and Objectives: This study aims to compare the characteristics, effectiveness and results of transcatheter closure of atrial septal defect between children, adolescents, and adults. Subjects and Methods: In this study, 683 patients who underwent atrial septal defect closure in the last 10 years were divided into three groups: children (age <12), adolescents (age 12 to 16), and adults (age >16) as group 1, group 2 and group 3, respectively. Results: The average defect size and incidence of complex atrial septal defect were higher in group 3 (p=0.0001 and 0.03 respectively). While the average size of the devic was higher in adults ($22.6{\pm}6.4mm$ vs. $18.5{\pm}4.9mm$; p=0.0001), the ratio of the device size/total septum was higher in both children and adolescents (Group 1 and 2). In the child and adolescent groups and patients with only complex atrial septal defect, the use of techniques, other than standard deployment, was similar in all three groups (p=0.86 and 0.41, respectively). The ratio of the residual shunt was similar in all three groups. Major complications were seen in 5 cases (4 cases with migration, and 1 case with dislocation) in group 3 and 1 case (migration) in group 1. Conclusion: Depending on the complexity of the defect and age of the patient, transcatheter closure of atrial septal defect might have certain difficulties and complications. Patients must be evaluated in detail to avoid major complications and possible problems during the procedure.

Keywords

References

  1. Kim NK, Park SJ, Choi JY. Transcatheter closure of atrial septal defect: does age matter? Korean Circ J 2011;41:633-8. https://doi.org/10.4070/kcj.2011.41.11.633
  2. Park SJ, Kim NK, Kim JO, Yoo BW, Choi JY, Sul JH. Morphologic Characteristics and RelatingFactors to the Need of Technical Modification in Transcatheter Closure of Large Atrial Septal Defect (>25mm). Korean Circ J 2010;40:191-6. https://doi.org/10.4070/kcj.2010.40.4.191
  3. Bishnoi RN, Everett AD, Ringel RE, et al. Device closure of secundum atrial septal defects in infants weighing less than 8 kg. Pediatr Cardiol 2014;35:1124-31. https://doi.org/10.1007/s00246-014-0905-7
  4. Lairakdomrong K, Srimahachota S, Lertsapcharoen P, Chaipromprasit J, Boonyaratavej S, Kaewsukkho P. Clinical results of large secundum atrial septal defect closure in adult using percutaneous transcatheter Cocoon atrial septal occluder. J Med Assoc Thai 2013;96:1127-34.
  5. Cardenas L, Panzer J, Boshoff D, Malekzadeh-Milani S, Ovaert C. Transcatheter closure of secundum atrial defect in small children. Catheter Cardiovasc Interv 2007;69:447-52. https://doi.org/10.1002/ccd.21030
  6. Dalvi B, Pinto R, Gupta A. Device closure of large atrial septal defects requiring devices > or =20 mm in small children weighing <20 kg. Catheter Cardiovasc Interv 2008;71:679-86. https://doi.org/10.1002/ccd.21450
  7. Zhang H, Chen Q, Chen LW, Cao H, Zhang GC, Chen DZ. Intraoperative device closure of atrial septal defects in the older population. J Cardiothorac Surg 2011;6:123. https://doi.org/10.1186/1749-8090-6-123
  8. Gabriels C, De Meester P, Pasquet A, et al. A different view on predictors of pulmonary hypertension in secundum atrial septal defect. Int J Cardiol 2014;176:833-40. https://doi.org/10.1016/j.ijcard.2014.08.009
  9. Lakhdhar R, Drissa M, Drissa H. Natural history of atrial septal defect in the sixth decade : study of 5 cases. Tunis Med 2013;91:243-7.
  10. Beitzke A, Zobel G, Nagel B, Koestenberger M. Transcatheter closure of an atrial septal defect in a newborn with aortic stenosis. Acta Paediatr 2009;98:582-3. https://doi.org/10.1111/j.1651-2227.2008.01149.x
  11. Du ZD, Hijazi ZM, Kleinman CS, Silverman NH, Larntz K; Amplatzer Investigators. Comparison between transcatheter and surgical closure of secundum atrial septal defect in children and adults: results of a multicenter nonrandomized trial. J Am Coll Cardiol 2002;39:1836-44. https://doi.org/10.1016/S0735-1097(02)01862-4
  12. Knirsch W, Dodge-Khatami A, Valsangiacomo-Buechel E, Weiss M, Berger F. Challenges encountered during closure of atrial septal defects. Pediatr Cardiol 2005;26:147-53. https://doi.org/10.1007/s00246-004-0958-0
  13. Cooke JC, Gelman JS, Harper RW. Chiari network entanglement and herniation into the left atrium by an atrial septal defect occluder device. J Am Soc Echocardiogr 1999;12:601-3. https://doi.org/10.1016/S0894-7317(99)70009-X
  14. Masseli J, Bertog S, Stanczak L, et al. Transcatheter closure of multiple interatrial communications. Catheter Cardiovasc Interv 2013;81:825-36. https://doi.org/10.1002/ccd.24329
  15. Turkay S, Abdullah E, Celal A, et al. Multiple transcatheter interventions in the same session in congenital cardiopathies. J Cardiovasc Dis Res 2010;1:181-90. https://doi.org/10.4103/0975-3583.74261
  16. Erdem A, Saritas T, Zeybek C, et al. Transthoracic echocardiographic guidance during transcatheter closure of atrial septal defects in children and adults. Int J Cardiovasc Imaging 2013;29:53-61.
  17. Geva T, Martins JD, Wald RM. Atrial septal defects. Lancet 2014;383:1921-32. https://doi.org/10.1016/S0140-6736(13)62145-5
  18. El-Said H, Hegde S, Foerster S, et al. Device therapy for atrial septal defects in a multicenter cohort: acute outcomes and adverse events. Catheter Cardiovasc Interv 2015;85:227-33. https://doi.org/10.1002/ccd.25684
  19. Diab KA, Cao QL, Bacha EA, Hijazi ZM. Device closure of atrial septal defects with the Amplatzer septal occluder: safety and outcome in infants. J Thorac Cardiovasc Surg 2007;134:960-6. https://doi.org/10.1016/j.jtcvs.2007.06.018
  20. Mahadevan VS, Gomperts N, Haberer K, et al. Transcatheter closure of atrial septal defects with multiple devices in adults: procedural and clinical outcomes. Int J Cardiol 2009;133:359-63. https://doi.org/10.1016/j.ijcard.2008.01.027
  21. Butera G, Romagnoli E, Saliba Z, et al. Percutaneous closure of multiple defects of the atrial septum: procedural results and longterm follow-up. Catheter Cardiovasc Interv 2010;76:121-8. https://doi.org/10.1002/ccd.22435
  22. Awad SM, Garay FF, Cao QL, Hijazi ZM. Multiple Amplatzer septal occluder devices for multiple atrial communications: immediate and long-term follow-up results. Catheter Cardiovasc Interv 2007;70:265-73. https://doi.org/10.1002/ccd.21145
  23. Kawamura A, Nishiyama N, Kawakami T, Fukuda K. Retrieval of embolized Amplatzer septal occluder using ablation catheter and triple-loop snare. Cardiovasc Interv Ther 2014;29:350-3. https://doi.org/10.1007/s12928-013-0238-0
  24. Mashman WE, King SB, Jacobs WC, Ballard WL. Two cases of late embolization of Amplatzer septal occluder devices to the pulmonary artery following closure of secundum atrial septal defects. Catheter Cardiovasc Interv 2005;65:588-92. https://doi.org/10.1002/ccd.20429
  25. Goel PK, Kapoor A, Batra A, Khanna R. Transcatheter retrieval of embolized AMPLATZER Septal Occluder. Tex Heart Inst J 2012;39:653-6.
  26. Dhaliwal RS, Singh H, Swami N, Srivastava V. Removal of displaced and impacted ASD device after 4 years. Thorac Cardiovasc Surg 2009;57:233-5. https://doi.org/10.1055/s-2008-1038727
  27. Szkutnik M, Lenarczyk A, Kusa J, Bialkowski J. Symptomatic tachyand bradyarrhythmias after transcatheter closure of interatrial communications with Amplatzer devices. Cardiol J 2008;15:510-6.
  28. Nehgme RA, Huddleston AR, Cheatham JP. Progression to late complete atrioventricular block following amplatzer device closure of atrial septal defect in a child. Pediatr Cardiol 2009;30:367-70. https://doi.org/10.1007/s00246-008-9322-0
  29. Rohit MK, Puri K, Vadivelu R. Reversible complete atrioventricular block after percutaneous ASD device closure in a child <15 kg. Indian Heart J 2014;66:366-9. https://doi.org/10.1016/j.ihj.2014.03.013
  30. Amoozgar H, Ahmadipoor M, Amirghofran AA. Complete heart block following transcatheter closure of atrial septal defect due to growth of inflammatory tissue. Pediatr Cardiol 2014;35:1301-3. https://doi.org/10.1007/s00246-014-0986-3

Cited by

  1. The Significance of Crochetage on the R wave of an Electrocardiogram for the Early Diagnosis of Pediatric Secundum Atrial Septal Defect vol.39, pp.5, 2016, https://doi.org/10.1007/s00246-018-1857-0
  2. Hemolysis and its outcome following percutaneous closure of cardiac defects among children and adolescents: a prospective study vol.45, pp.1, 2016, https://doi.org/10.1186/s13052-019-0728-5
  3. Procedural Predictors and Outcomes of Percutaneous Secundum Atrial Septal Defect Closure in Children Aged &lt;6 Years vol.85, pp.9, 2016, https://doi.org/10.1253/circj.cj-20-1023
  4. Association between the Use of Diuretics and Size Reduction in Pediatric Atrial Septal Defect vol.51, pp.None, 2021, https://doi.org/10.4070/kcj.2021.0076