국내 방사선 측정장비 보유 현황 및 교정 현황 조사

Report of Present Status of Calibration for Domestic Radiation Measurements Instruments

  • 임상욱 (고신대학교 의과대학 방사선종양학교실) ;
  • 최진호 ;
  • 안소현 (가천대학교 의과대학 방사선종양학교실) ;
  • 조광환 (순천향대학교 의과대학 방사선종양학교실) ;
  • 이상훈 (관동대학교 의과대학 방사선종양학교실) ;
  • 이레나 (이화여자대학교 의과대학 방사선종양학교실) ;
  • 조삼주 (연세대학교 의과대학 방사선종양학교실)
  • Lim, Sangwook (Department of Radiation Oncology, Kosin University College of Medicine) ;
  • Choi, Jinho (Department of Radiation Oncology, Sheikh Khalifa Specialty Hospital Ras Al Khaimah) ;
  • An, Sohyun (Department of Radiation Oncology, Gachon University of Medicine and Science) ;
  • Cho, Kwang Hwan (Department of Radiation Oncology, Soonchunhyang University College of Medicine) ;
  • Lee, Sang Hoon (Department of Radiation Oncology, Kwandong University College of Medicine) ;
  • Lee, Rena (Department of Radiation Oncology, School of Medicine, Ewha Womans University) ;
  • Cho, Sam Ju (Department of Radiation Oncology, Yonsei University College of Medicine)
  • 투고 : 2016.03.15
  • 심사 : 2016.03.28
  • 발행 : 2016.03.31

초록

치료장비의 정확한 정도관리를 위해서는 선량측정기의 주기적 교정이 중요하다. 국내 방사선종양학과에서 사용하는 교정대상 측정기는 이온전리함, 온도계, 기압계 및 서베이미터 등이 있다. 현재 국내 교정기관에서 권고하고 있는 교정 유효 기간은 6개월에서 1년이나 최근 제조사 및 외국 교정기관에서는 2년의 유효기간을 제시하고 있다. 따라서 측정기의 특성에 따른 가장 합리적인 교정주기의 설정이 필요한 실정이다. 본 연구에서는 ILAC-G24 지침서(2) 등을 기반으로 국내 방사선종양학과에서 사용하는 측정기들의 최적의 교정주기를 설정하기 위한 기초단계로 불확도, 측정기의 사용 빈도, 사용 방법, 장비의 안정도 등을 조사하여 향후 측정장비의 적절한 유효 기간 도출의 기본 자료로 활용하고자 한다. 조사 방법은 온라인 설문조사도구인 SurveyMonkey를 이용하였으며 조상대상은 전국의 18개의 대학병원으로 이중 15개 기관에서 응답하였다. 설문조사 질의 내용은 작성자 정보, 온도계, 기압계, 서베이미터, 기준이온전리함, 소조사면 측정장비 및 근접치료용 우물형 전리함으로 구성하였고 총 64문항으로 모두 응답시 약 50분간의 시간이 소요된다. 조사대상의 과반수 기관에서 온도계와 기압계의 교정을 실시하지 않는 것으로 조사되었다. 반면 서베이미터는 6개월에서 1년 정도로 교정주기가 비교적 짧았다. 본 조사를 바탕으로 측정장치별 최적의 교정주기를 결정할 수 있을 것으로 사료된다.

Periodical calibrations of radiation detectors are important for accurate quality assurance of therapeutic linac. The measuring instruments such as ion-chamber, thermometer, barometer, and survey meter should be calibrated periodically. Period of calibration for these instruments is suggested 6 month to one year in Korea and two years in other countries nowadays. Therefore, the determination of reasonable period for calibration is needed. In this study, we plan to utilize the results of these survey; frequent in use, how to use and stability of instruments, to determine the optimized period of calibration for the instruments in the departments of radiation oncology in Korea based on the ILAC-G24. The SurveyMonkey web-based survey tool was used and the objects of survey were 18 department of radiation oncology in university hospitals, and 15 departments were answered. The 64 questionnaires which supposed to be answered in 50 minutes were classified as the information of candidates, the thermometer, the barometer, the surveymeter, and the ion-chamber. The thermometers and the barometers were not under periodical calibration for more than half of candidates. The periods of calibration of surveymeters were 6 month to 1 year. We expect that the calibration period can be determined based on these survey results.

키워드

참고문헌

  1. Gary A. Ezzell et al: IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med. Phys, 36(11), 5359-5373 (2009) https://doi.org/10.1118/1.3238104
  2. Alber M, Broggi S, De Wagter C: Guidelines for the verification of IMRT. Brussels, Belgium: ESTRO (2008)
  3. Thomas J. LoSasso, Ph.D: IMRT Delivery System QA. Intensity modulated radiation therapy-The state of the art (Palta JR, Mackie TR, editors). Madison, WI: Med. Phys; 561-591 (2003)
  4. ICRU 62: Prescribing, Recording, and Reporting Photon Beam Therapy (2001)
  5. Shaw, R. Kline, M. Gillin, L. et. al: Radiation Therapy Oncology Group: radiosurgery quality assurance guidelines Int J Radiat Oncol Biol Phys. 27(5), 1231-1239 (1993) https://doi.org/10.1016/0360-3016(93)90548-A
  6. NASA Handbook NHB: Metrology and Ca14)libration Provisions Guidelines, Jet Propulsion. Laboratory 5300.4(1A) (2000)
  7. P.J. Chiaro, Jr: Calibration Interval Thecnical Basis Document. Instrumentation and Controls Division (1998)
  8. ANSI N323A: American National Standard Radiation Protection Instrumentation Test and Calibration, Portable Survey Instruments. IEEE, USA (1997)
  9. IAEA Report NO. 16: Calibration of Radiation Protection Monitoring Instruments. (2000)
  10. Notification of Nuclear Safety and Security Commission: Technological standards for radiation safety of medical field (No. 2015-005).
  11. ILAC-G24: Guidelines for the determination of calibration intervals of measuring instruments. International Laboratory Accreditation Cooperation (2007)