DOI QR코드

DOI QR Code

An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files

  • Gehlot, Paras Mull (Department of Conservative Dentistry and Endodontics, JSS University, JSS Dental College and Hospital) ;
  • Manjunath, Vinutha (Department of Conservative Dentistry and Endodontics, JSS University, JSS Dental College and Hospital) ;
  • Manjunath, Mysore Krishnaswamy (Department of Conservative Dentistry and Endodontics, JSS University, JSS Dental College and Hospital)
  • 투고 : 2015.07.21
  • 심사 : 2015.10.20
  • 발행 : 2016.02.29

초록

Objectives: The purpose of this in vitro study was to evaluate the accuracy of working length (WL) determination of four electronic apex locators (EALs), namely, Root ZX (RZX), Elements diagnostic unit and apex locator (ELE), SybronEndo Mini Apex locator (MINI) and Propex pixi (PIXI) using Stainless steel (SS) and nickel-titanium (NiTi) hand files. The null hypothesis was that there was no difference between canal length determination by SS and NiTi files of 4 EALs. Materials and Methods: Sixty extracted, single rooted human teeth were decoronated and the canal orifice flared. The actual length (AL) was assessed visually, and the teeth were embedded in an alginate model. The electronic length (EL) measurements were recorded with all four EALs using SS and NiTi files at '0.5' reading on display. The differences between the AL and EL were compared. Results: The results obtained with each EAL with SS and NiTi files were compared with AL. A paired sample t test showed that there was a statistical significant difference between EAL readings with SS and NiTi files for RZX and MINI (p < 0.05). The accuracy of RZX, ELE, MINI and PIXI within ${\pm}0.5 mm$ of AL with SS/NiTi files were 93.3%/70%, 90%/91.7%, 95%/68.3%, and 83.3%/83.3%, respectively. Conclusions: The results of this study indicate that Root ZX was statistically more accurate with NiTi files compared to SS files, while MINI was statistically more accurate with SS files compared to NiTi files. ELE and PIXI were not affected by the alloy type of the file used to determine WL.

키워드

참고문헌

  1. Ingle JI, Backland LK. Endodontics. 5th ed. Loma Linda, California: BC Decker Inc.; 2002. Chapter 10. Endodontic cavity preparation; p510-525.
  2. Kim E, Lee SJ. Electronic apex locator. Dent Clin North Am 2004;48:35-54. https://doi.org/10.1016/j.cden.2003.10.005
  3. Sunada I. New method for measuring the length of the root canal. J Dent Res 1962;41:375-387. https://doi.org/10.1177/00220345620410020801
  4. Nekoofar MH, Ghandi MM, Hayes SJ, Dummer PM. The fundamental operating principles of electronic root canal length measurement devices. Int Endod J 2006;39:595-609. https://doi.org/10.1111/j.1365-2591.2006.01131.x
  5. Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod 1994;20:111-114. https://doi.org/10.1016/S0099-2399(06)80053-1
  6. Gordon MP, Chandler NP. Electronic apex locators. Int Endod J 2004;37:425-37. https://doi.org/10.1111/j.1365-2591.2004.00835.x
  7. Bonilla M, Sayin TC, Schobert B, Hardigan P. Accuracy of a new apex locator in ex-vivo teeth using scanning electron microscopy. Endod Pract 2014;14-20. Available from: http://www.endopracticeus.com/clinical-articles/accuracy-of-a-new-apex-locator-in-ex-vivo-teeth-usingscanning-electron-microscopy/ (updated 2015 Sept 19).
  8. Dentsply Maillefer: Propex pixi apex locator (User manual). Available from: http://dentsplymea.com/sites/default/files/DFU%20(English).pdf (updated 2015 Dec 27).
  9. Thomas AS, Hartwell GR, Moon PC. The accuracy of the Root ZX electronic apex locator using stainless-steel and nickel-titanium files. J Endod 2003;29:662-663. https://doi.org/10.1097/00004770-200310000-00012
  10. Siu C, Marshall JG, Baumgartner JC. An in vivo comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by using rotary nickel-titanium files. J Endod 2009;35:962-965. https://doi.org/10.1016/j.joen.2009.04.025
  11. Higa RA, Adorno CG, Ebrahim AK, Suda H. Distance from file tip to the major apical foramen in relation to the numeric meter reading on the display of three different electronic apex locators. Int Endod J 2009;42:1065-1070. https://doi.org/10.1111/j.1365-2591.2009.01629.x
  12. Huang L. An experimental study of the principle of electronic root canal measurement. J Endod 1987;13:60-64. https://doi.org/10.1016/S0099-2399(87)80156-5
  13. Pilot TF, Pitts DL. Determination of impedance changes at varying frequencies in relation to root canal file position and irrigant. J Endod 1997;23:719-724. https://doi.org/10.1016/S0099-2399(97)80341-X
  14. Kuttler Y. Microscopic investigation of root apexes. J Am Dent Assoc 1955;50:544-552. https://doi.org/10.14219/jada.archive.1955.0099
  15. Guise GM, Goodell GG, Imamura GM. In vitro comparison of three electronic apex locators. J Endod 2010;36:279-281. https://doi.org/10.1016/j.joen.2009.09.016
  16. D'Assuncao FL, de Albuquerque DS, Salazar-Silva JR, de Queiroz Ferreira LC, Bezerra PM. The accuracy of root canal measurements using the Mini Apex Locator and Root ZX-II: An evaluation in vitro. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104:e50-e53.
  17. Plotino G, Grande NM, Brigante L, Lesti B, Somma F. Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex. Int Endod J 2006;39:408-414. https://doi.org/10.1111/j.1365-2591.2006.01095.x
  18. Ebrahim AK, Yoshioka T, Kobayashi C, Suda H. The effects of file size, sodium hypochlorite and blood on the accuracy of Root ZX apex locator in enlarged root canals: an in vitro study. Aust Dent J 2006;51:153-157. https://doi.org/10.1111/j.1834-7819.2006.tb00419.x
  19. ElAyouti A, Lost C. A simple mounting model for consistent determination of the accuracy and repeatability of apex locators. Int Endod J 2006;39:108-112. https://doi.org/10.1111/j.1365-2591.2006.01052.x
  20. Ounsi HF, Naaman A. In vitro evaluation of the reliability of the Root ZX electronic apex locator. Int Endod J 1999;32:120-123. https://doi.org/10.1046/j.1365-2591.1999.00202.x
  21. Barthelemy J, Gregor L, Krejci I, Wataha J, Bouillaguet S. Accuracy of electronic apex locater-controlled handpieces. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009;107:437-441. https://doi.org/10.1016/j.tripleo.2008.11.010
  22. Altenburger MJ, Cenik Y, Schirrmeister JF, Wrbas KT, Hellwig E. Combination of apex locator and endodontic motor for continuous length control during root canal treatment. Int Endod J 2009;42:368-374. https://doi.org/10.1111/j.1365-2591.2008.01535.x
  23. Sadeghi S, Abolghasemi M. The accuracy of the Raypex5 electronic apex locator using stainless-steel hand K-file versus nickel-titanium rotary Mtwo file. Med Oral Patol Oral Cir Bucal 2010;15:e788-e790.
  24. Walia HM, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of Nitinol root canal files. J Endod 1988;14:346-351. https://doi.org/10.1016/S0099-2399(88)80196-1
  25. Thompson SA. An overview of nickel-titanium alloys used in dentistry. Int Endod J 2000;33:297-310. https://doi.org/10.1046/j.1365-2591.2000.00339.x
  26. Bergmans L, Van Cleynenbreugel J, Wevers M, Lambrechts P. Mechanical root canal preparation with NiTi rotary instruments: rationale, performance and safety. Status report for the American Journal of Dentistry. Am J Dent 2001;14:324-333.
  27. Goldberg F, Marroquin BB, Frajlich S, Dreyer C. In vitro evaluation of the ability of three apex locators to determine the working length during retreatment. J Endod 2005;31:676-678. https://doi.org/10.1097/01.don.0000155226.03483.ff
  28. Weiger R, John C, Geigle H, Lost C. An in vitro comparison of two modern apex locators. J Endod 1999;25:765-768. https://doi.org/10.1016/S0099-2399(99)80128-9
  29. D'Assuncao FL, Albuquerque DS, Salazar-Silva JR, Dos Santos VC, Sousa JC. Ex vivo evaluation of the accuracy and coefficient of repeatability of three electronic apex locators using a simple mounting model: a preliminary report. Int Endod J 2010;43:269-274. https://doi.org/10.1111/j.1365-2591.2009.01674.x
  30. Mull JP, Manjunath V, Manjunath M. Comparison of accuracy of two electronic apex locators in the presence of various irrigants: an in vitro study. J Conserv Dent 2012;15:178-182. https://doi.org/10.4103/0972-0707.94585
  31. Stober EK, de Ribot J, Mercade M, Vera J, Bueno R, Roig M, Duran-Sindreu F. Evaluation of the Raypex 5 and the Mini Apex Locator: an in vivo study. J Endod 2011;37:1349-1352. https://doi.org/10.1016/j.joen.2011.06.021
  32. Nekoofar MH, Sadeghi K, Sadighi Akha E, Namazikhah MS. The accuracy of the Neosono Ultima EZ apex locator using files of different alloys: an in vitro study. J Calif Dent Assoc 2002;30:681-684.
  33. TIBTECH innovations: Properties table of Stainless steel, Metals and other Conductive materials. Available from: http://www.tibtech.com/conductivity.php (updated 2015 Mar 20).
  34. Lee SJ, Nam KC, Kim YJ, Kim DW. Clinical accuracy of a new apex locator with an automatic compensation circuit. J Endod 2002;28:706-709. https://doi.org/10.1097/00004770-200210000-00007
  35. Jung IY, Yoon BH, Lee SJ, Lee SJ. Comparison of the reliability of "0.5" and "APEX" mark measurements in two frequency-based electronic apex locators. J Endod 2011;37:49-52. https://doi.org/10.1016/j.joen.2010.08.048
  36. Hor D, Krusy S, Attin T. Ex vivo comparison of two electronic apex locators with different scales and frequencies. Int Endod J 2005;38:855-859. https://doi.org/10.1111/j.1365-2591.2005.01004.x
  37. Venturi M, Breschi L. A comparison between two electronic apex locators: an ex vivo investigation. Int Endod J 2007;40:362-373. https://doi.org/10.1111/j.1365-2591.2006.01229.x
  38. Meredith N, Gulabivala K. Electrical impedance measurements of root canal length. Endod Dent Traumatol 1997;13:126-131. https://doi.org/10.1111/j.1600-9657.1997.tb00025.x
  39. Kim PJ, Kim HG, Cho BH. Evaluation of electrical impedance ratio measurements in accuracy of electronic apex locators. Restor Dent Endod 2015;40:113-122. https://doi.org/10.5395/rde.2015.40.2.113
  40. Shen Y, Zhou HM, Zheng YF, Peng B, Haapasalo M. Current challenges and concepts of the thermomechanical treatment of nickel-titanium instruments. J Endod 2013;39:163-172. https://doi.org/10.1016/j.joen.2012.11.005

피인용 문헌

  1. Root ZX Electronic Foramen Locator: An Ex Vivo Study of Its Three Models’ Precision and Reproducibility vol.2017, pp.1687-8736, 2017, https://doi.org/10.1155/2017/5893790
  2. Determination of the Accuracy of 5 Electronic Apex Locators in the Function of Different Employment Protocols vol.43, pp.10, 2016, https://doi.org/10.1016/j.joen.2017.03.039
  3. A clinical evaluation of two electronic apex locators and conventional radiography in working length determination in primary molar and its influence on children's behavioral responses vol.38, pp.2, 2020, https://doi.org/10.4103/jisppd.jisppd_28_20
  4. Effect of solvent use on postoperative pain in root canal retreatment: a randomized, controlled clinical trial vol.24, pp.1, 2016, https://doi.org/10.1007/s00784-019-02948-3
  5. The accuracy of electronic apex locators for determining working length: An in vitro study with artificial teeth vol.47, pp.2, 2016, https://doi.org/10.1111/aej.12455