Stability of the prosthetic screws of three types of craniofacial prostheses retention systems

  • Lanata-Flores, Antonio Gabriel (Division of Oral and Maxillofacial Surgery, Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas) ;
  • Sigua-Rodriguez, Eder Alberto (Division of Oral and Maxillofacial Surgery, Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas) ;
  • Goulart, Douglas Rangel (Division of Oral and Maxillofacial Surgery, Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas) ;
  • Bomfim-Azevedo, Veber Luiz (Division of Oral and Maxillofacial Surgery, Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas) ;
  • Olate, Sergio (Department of Oral and Maxillofacial Surgery, Universidad de La Frontera) ;
  • de Albergaria-Barbosa, Jose Ricardo (Division of Oral and Maxillofacial Surgery, Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas)
  • Received : 2016.06.02
  • Accepted : 2016.09.01
  • Published : 2016.12.31

Abstract

Objectives: This study aimed to evaluate the stability of prosthetic screws from three types of craniofacial prostheses retention systems (bar-clip, ball/O-ring, and magnet) when submitted to mechanical cycling. Materials and Methods: Twelve models of acrylic resin were used with implants placed 20 mm from each other and separated into three groups: (1) bar-clip (Sistema INP, São Paulo, Brazil), (2) ball/O-ring (Sistema INP), and (3) magnet (Metalmag, São Paulo, Brazil), with four samples in each group. Each sample underwent a mechanical cycling removal and insertion test (f=0.5 Hz) to determine the torque and the detorque values of the retention screws. A servo-hydraulic MTS machine (810-Flextest 40; MTS Systems, Eden Prairie, MN, USA) was used to perform the cycling with 2.5 mm and a displacement of 10 mm/s. The screws of the retention systems received an initial torque of 30 Ncm and the torque values required for loosening the screw values were obtained in three cycles (1,080, 2,160, and 3,240). The screws were retorqued to 30 Ncm before each new cycle. Results: The sample was composed of 24 screws grouped as follows: bar-clip (n=8), ball/O-ring (n=8), and magnet (n=8). There were significant differences between the groups, with greater detorque values observed in the ball/O-ring group when compared to the bar-clip and magnet groups for the first cycle. However, the detorque value was greater in the bar-clip group for the second cycle. Conclusion: The results of this study indicate that all prosthetic screws will loosen slightly after an initial tightening torque, also the bar-clip retention system demonstrated greater loosening of the screws when compared with ball/O-ring and magnet retention systems.

Keywords

References

  1. Bacchi A, Regalin A, Bhering CL, Alessandretti R, Spazzin AO. Loosening torque of Universal Abutment screws after cyclic loading: influence of tightening technique and screw coating. J Adv Prosthodont 2015;7:375-9. https://doi.org/10.4047/jap.2015.7.5.375
  2. Jemt T, Lekholm U, Grondahl K. 3-year followup study of early single implant restorations ad modum Brånemark. Int J Periodontics Restorative Dent 1990;10:340-9.
  3. Binon PP. Evaluation of the effectiveness of a technique to prevent screw loosening. J Prosthet Dent 1998;79:430-2. https://doi.org/10.1016/S0022-3913(98)70157-X
  4. Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the implant-abutment interface after fatigue testing. J Prosthet Dent 2001;85:268-75. https://doi.org/10.1067/mpr.2001.114266
  5. Karakoca S, Aydin C, Yilmaz H, Bal BT. Retrospective study of treatment outcomes with implant-retained extraoral prostheses: survival rates and prosthetic complications. J Prosthet Dent 2010;103:118-26. https://doi.org/10.1016/S0022-3913(10)60015-7
  6. Visser A, Raghoebar GM, van Oort RP, Vissink A. Fate of implantretained craniofacial prostheses: life span and aftercare. Int J Oral Maxillofac Implants 2008;23:89-98.
  7. Sencimen M, Bal HE, Demirogullari M, Kocaoglu M, Dogan N. Auricular episthesis retained by an attachment system (2 case reports). Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008; 105:e28-34.
  8. Hooper SM, Westcott T, Evans PL, Bocca AP, Jagger DC. Implantsupported facial prostheses provided by a maxillofacial unit in a U.K. regional hospital: longevity and patient opinions. J Prosthodont 2005;14:32-8. https://doi.org/10.1111/j.1532-849X.2005.00004.x
  9. Aydin C, Karakoca S, Yilmaz H, Yilmaz C. Implant-retained auricular prostheses: an assessment of implant success and prosthetic complications. Int J Prosthodont 2008;21:241-4.
  10. Weiss EI, Kozak D, Gross MD. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent 2000;84:194-9. https://doi.org/10.1067/mpr.2000.108069
  11. Cardoso M, Torres MF, Lourenco EJ, de Moraes Telles D, Rodrigues RC, Ribeiro RF. Torque removal evaluation of prosthetic screws after tightening and loosening cycles: an in vitro study. Clin Oral Implants Res 2012;23:475-80. https://doi.org/10.1111/j.1600-0501.2011.02165.x
  12. Fromentin O, Picard B, Tavernier B. In vitro study of the retention and mechanical fatigue behavior of four implant overdenture studtype attachments. Pract Periodontics Aesthet Dent 1999;11:391-7; quiz 398.
  13. Wichmann MG, Kuntze W. Wear behavior of precision attachments. Int J Prosthodont 1999;12:409-14.
  14. Cho WR, Huh YH, Park CJ, Cho LR. Effect of cyclic loading and retightening on reverse torque value in external and internal implants. J Adv Prosthodont 2015;7:288-93. https://doi.org/10.4047/jap.2015.7.4.288
  15. de Sousa AA, Mattos BS. Magnetic retention and bar-clip attachment for implant-retained auricular prostheses: a comparative analysis. Int J Prosthodont 2008;21:233-6.
  16. Moura e Costa AJ, Lasserre FS, Westphalen VP, Deonizio MD, da Silva Neto UX, de Sousa MH. Delayed tooth replantation: case report. Rev Clin Pesq Odontol 2004;1:41-3.
  17. Patterson EA, Johns RB. Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants. Int J Oral Maxillofac Implants 1992;7:26-33.
  18. Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol 2002;28:67-73. https://doi.org/10.1563/1548-1336(2002)028<0067:TRBIPA>2.3.CO;2
  19. Spazzin AO, Henrique GE, Nobilo MA, Consani RL, Correr-Sobrinho L, Mesquita MF. Effect of retorque on loosening torque of prosthetic screws under two levels of fit of implant-supported dentures. Braz Dent J 2010;21:12-7. https://doi.org/10.1590/S0103-64402010000100002