DOI QR코드

DOI QR Code

Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants

  • Mohammed, Hnd Hadi (Department of Prosthodontics, College of Dentistry, Wonkwang University) ;
  • Lee, Jin-Han (Department of Prosthodontics, College of Dentistry, Wonkwang University) ;
  • Bae, Ji-Myung (Department of Dental Biomaterials, College of Dentistry, Wonkwang University) ;
  • Cho, Hye-Won (Department of Prosthodontics, College of Dentistry, Wonkwang University)
  • Received : 2016.01.04
  • Accepted : 2016.02.16
  • Published : 2016.02.29

Abstract

PURPOSE. The purpose of this study was to evaluate the effects of abutment screw length and cyclic loading on the removal torque (RTV) in external hex (EH) and internal hex (IH) implants. MATERIALS AND METHODS. Forty screw-retained single crowns were connected to external and internal hex implants. The prepared titanium abutment screws were classified into 8 groups based on the number of threads (n = 5 per group): EH 12.5, 6.5, 3.5, 2.5 and IH 6.5, 5, 3.5, 2.5 threads. The abutment screws were tightened with 20 Ncm torque twice with 10-minute intervals. After 5 minutes, the initial RTVs of the abutment screws were measured with a digital torque gauge (MGT12). A customized jig was constructed to apply a load along the implant long axis at the central fossa of the maxillary first molar. The post-loading RTVs were measured after 16,000 cycles of mechanical loading with 50 N at a 1-Hz frequency. Statistical analysis included one-way analysis of variance and paired t-tests. RESULTS. The post-loading RTVs were significantly lower than the initial RTVs in the EH 2.5 thread and IH 2.5 thread groups (P<.05). The initial RTVs exhibited no significant differences among the 8 groups, whereas the post-loading RTVs of the EH 6.5 and EH 3.5 thread groups were higher than those of the IH 3.5 thread group (P<.05). CONCLUSION. Within the limitations of this study, the external hex implants with short screw lengths were more advantageous than internal hex implants with short screw lengths in torque maintenance after cyclic loading.

Keywords

References

  1. Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the implant-abutment interface after fatigue testing. J Prosthet Dent 2001;85:268-75. https://doi.org/10.1067/mpr.2001.114266
  2. Jorneus L, Jemt T, Carlsson L. Loads and design of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992;7:353-9.
  3. Bianco G, Di Raimondo R, Luongo G, Paoleschi C, Piccoli P, Piccoli C, Rangert B. Osseointegrated implant for single tooth replacement: A retrospective multicenter study on routine use in private practice. Clin lmplant Dent Relat Res 2000; 2:152-8. https://doi.org/10.1111/j.1708-8208.2000.tb00006.x
  4. Theoharidou A, Petridis HP, Tzannas K, Garefis P. Abutment screw loosening in single-implant restorations: a systematic review. Int J Oral Maxillofac Implants 2008;23:681-90.
  5. Richter EJ. In vivo horizontal bending moments on implants. Int J Oral Maxillofac Implants 1998;13:232-44.
  6. Bickford JH. An introduction to the design and behavior of bolted joints. New York: Marcel Dekker; 1995. p. 515-64.
  7. Tsuge T, Hagiwara Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent Mater J 2009;28:373-81. https://doi.org/10.4012/dmj.28.373
  8. Delben JA, Gomes EA, Barao VA, Assuncao WG. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws. Int J Oral Maxillofac Implants 2011;26:251-56.
  9. Stuker RA, Teixeira ER, Beck JC, da Costa NP. Preload and torque removal evaluation of three different abutment screws for single standing implant restorations. J Appl Oral Sci 2008; 16:55-8. https://doi.org/10.1590/S1678-77572008000100011
  10. Khraisat A, Hashimoto A, Nomura S, Miyakawa O. Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system. J Prosthet Dent 2004;91:326-34. https://doi.org/10.1016/j.prosdent.2004.01.001
  11. Shigley JE. Mechanical engineering design. 3rd ed. New York: McGraw Hill; 1977. p. 240-5.
  12. Weiss EI, Kozak D, Gross MD. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent 2000;84:194-9. https://doi.org/10.1067/mpr.2000.108069
  13. Delben JA, Barao VA, Dos Santos PH, Assuncao WG. Influence of abutment type and esthetic veneering on preload maintenance of abutment screw of implant-supported crowns. J Prosthodont 2014;23:134-9. https://doi.org/10.1111/jopr.12065
  14. Khraisat A, Abu-Hammad O, Dar-Odeh N, Al-Kayed AM. Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading. Clin Implant Dent Relat Res 2004;6:157-64. https://doi.org/10.1111/j.1708-8208.2004.tb00223.x
  15. Al Jabbari YS, Fournelle R, Ziebert G, Toth J, Iacopino AM. Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 1: Characterization of adhesive wear and structure of retaining screws. J Prosthodont 2008;17:168-80. https://doi.org/10.1111/j.1532-849X.2007.00265.x
  16. Shin HM, Huh JB, Yun MJ, Jeon YC, Chang BM, Jeong CM. Influence of the implant-abutment connection design and diameter on the screw joint stability. J Adv Prosthodont 2014; 6:126-32. https://doi.org/10.4047/jap.2014.6.2.126
  17. Gracis S, Michalakis K, Vigolo P, Vult von Steyern P, Zwahlen M, Sailer I. Internal vs. external connections for abutments/reconstructions: A systematic review. Clin Oral Implants Res 2012;23:202-16. https://doi.org/10.1111/j.1600-0501.2012.02556.x
  18. Feitosa PC, de Lima AP, Silva-Concilio LR, Brandt WC, Neves AC. Stability of external and internal implant connections after a fatigue test. Eur J Dent 2013;7:267-71. https://doi.org/10.4103/1305-7456.115407
  19. Kim KS, Han JS, Lim YJ. Settling of abutments into implants and changes in removal torque in five different implant-abutment connections. Part 1: Cyclic loading. Int J Oral Maxillofac Implants 2014;29:1079-84. https://doi.org/10.11607/jomi.3383
  20. Binon PP. Implants and components: entering the new millennium. Int J Oral Maxillofac Implants 2000;15:76‑94.
  21. Sopwith DG. The distribution of load in screw threads. Proc Inst Mech Eng 1948;159:373-83. https://doi.org/10.1243/PIME_PROC_1948_159_030_02
  22. Kim BJ, Yeo IS, Lee JH, Kim SK, Heo SJ, Koak JY. The effect of screw length on fracture load and abutment strain in dental implants with external abutment connections. Int J Oral Maxillofac Implants 2012;27:820-3.
  23. Choi JH, Yang JH, Cho WP, Lee JB. The influence of abutment screw length and repeated tightening on screw loosening in dental implant. J Korean Acad Prosthodont 2006;44:432-42.
  24. Yeo IS, Lee JH, Kang TJ, Kim SK, Heo SJ, Koak JY, Park JM, Lee SY. The effect of abutment screw length on screw loosening in dental implants with external abutment connections after thermocycling. Int J Oral Maxillofac Implants 2014;29:59-62. https://doi.org/10.11607/jomi.3125
  25. Park SD, Lee Y, Kim YL, Yu SH, Bae JM, Cho HW. Microleakage of different sealing materials in access holes of internal connection implant systems. J Prosthet Dent 2012; 108:173-80 https://doi.org/10.1016/S0022-3913(12)60143-7
  26. Winkler S, Ring K, Ring JD, Boberick KG. Implant screw mechanics and the settling effect: Overview. J Oral Implantol 2003;29:242-5. https://doi.org/10.1563/1548-1336(2003)029<0242:ISMATS>2.3.CO;2
  27. Graf H. Bruxism. Dent Clin North Am 1969;13:659-65.
  28. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-36.
  29. Maeda, Y, Satoh T, Sogo M. In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication. J Oral Rehabil 2006;33:75-8. https://doi.org/10.1111/j.1365-2842.2006.01545.x
  30. Kim ES, Shin SY. Influence of the implant abutment types and the dynamic loading on initial screw loosening. J Adv Prosthodont 2013;5:21-8. https://doi.org/10.4047/jap.2013.5.1.21

Cited by

  1. Effect of Angle and Type of Customized Abutment (Castable & Cast-to) on Torque Loss and Fracture Resistance After Cyclic Loading vol.12, pp.None, 2016, https://doi.org/10.2174/1874210601811120987
  2. Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading vol.4, pp.None, 2016, https://doi.org/10.1186/s40729-018-0149-z
  3. Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study vol.5, pp.None, 2016, https://doi.org/10.1186/s40729-018-0154-2
  4. Analysis of the Influence of Parafunctional Loads on the Bone-Prosthesis System: A Non-Linear Finite Element Analysis vol.14, pp.6, 2016, https://doi.org/10.4236/jbise.2021.146019
  5. Analysis of Torque Maintenance and Fracture Resistance after Fatigue in Retention Screws Made of Different Metals for Screw-Retained Implant-Borne Prosthesis Joints vol.2021, pp.None, 2021, https://doi.org/10.1155/2021/9693239
  6. Screw stability of CAD‐CAM titanium and zirconia abutments on different implants: An in vitro study vol.23, pp.3, 2021, https://doi.org/10.1111/cid.13001
  7. Reverse Torque Values of Abutment Screws After Dynamic Loading: Effects of Sealant Agents and the Taper of Conical Connections vol.47, pp.4, 2016, https://doi.org/10.1563/aaid-joi-d-19-00228
  8. Effect of different tightening protocols on the probability of survival of screw-retained implant-supported crowns vol.126, pp.None, 2022, https://doi.org/10.1016/j.jmbbm.2021.105019