The main purpose of this article is a complex hyperthermia applicator system design for treatment of head and neck region. The applicator system is composed of four waveguides with a stripline horn aperture and circular water bolus. The specific absorption rate (SAR) and temperature distribution from this applicator in various numerical phantom models was investigated. For used targets, the treatment planning based on the optimization process made through the SEMCAD X software is added to show the steering possibilities of SAR and thereby temperature distribution. Using treatment planning software, we proved that the SAR and temperature distribution can be effectively controlled (by amplitude and phase changing) improving the SAR and temperature target coverage approximately by 20 %. For the proposed applicator system analysis and quantitative evaluation of two parameters 25 % iso-SAR and $41^{\circ}C$ iso-temperature contours in the treatment area with the respect to sensitive structures in treatment area were defined. To verify our simulation results, the real measurement of reflectivity coefficient as well as the temperature distribution in a homogenous phantom were performed.

키워드

참고문헌

  1. A. Argiris, M. V. Karamouzis, D. Raben, and R. L. Ferris, "Head and neck cancer," The Lancet, vol. 371, no. 9625, pp. 1695-1709, May 2008. https://doi.org/10.1016/S0140-6736(08)60728-X
  2. L. P. Kowalski, "Head and neck cancer," BMC Proceedings, vol. 7, no. Suppl 2, p. K12, Apr. 2013.
  3. J. Vrba, "Medical Applications of Microwaves" (In Czech), 1st ed., CTU in Prague, Czech Republic, 2003.
  4. J. van der Zee, D. Gonzalez, G. C. van Rhoon, J. D. van Dijk, W. L. van Putten, and A. A. Hart, "Comparison of radiotherapy alone with radiotherapy plus hyperthermia in locally advanced pelvic tumours: a prospective, randomised, multicentre trial," The Lancet, vol. 355, no. 9210, pp. 1119-1125, Apr. 2000. https://doi.org/10.1016/S0140-6736(00)02059-6
  5. R. Colombo, L. F. Do Pozzo, A. Salonia, P. Rigatti, Z. Leib, J. Baniel, E. Caldarera, and M. Pavone-Macaluso, "Multicentric study comparing intravesical chemotherapy alone and with local microwave hyperthermia for prophylaxis of recurrence of superficial transitional cell carcinoma," J. Clin. Oncol., vol. 21, no. 23, pp. 4270-4276, Dec. 2003. https://doi.org/10.1200/JCO.2003.01.089
  6. M. M. Paulides, J. F. Bakker, E. Neufeld, J. van der Zee, P. P. Jansen, P. C. Levendag, and G. C. van Rhoon, "The HYPERcollar: A novel applicator for hyperthermia in the head and neck.," International Journal of Hyperthermia, vol. 23, no. 7, pp. 567-576, Nov. 2007. https://doi.org/10.1080/02656730701670478
  7. M. M. Paulides, J. F. Bakker, M. Linthorst, J. van der Zee, Z. Rijnen, E. Neufeld, P. M. T. Pattynama, P. P. Jansen, P. C. Levendag, and G. C. van Rhoon, "The clinical feasibility of deep hyperthermia treatment in the head and neck: new challenges for positioning and temperature measurement," Physics in Medicine and Biology, vol. 55, no. 9, pp. 2465-2480, May 2010. https://doi.org/10.1088/0031-9155/55/9/003
  8. P. Wust, M. Seebass, J. Nadobny, P. Deuflhard, G. Monicrh, and R. Felix, "Simulation studies promote technological development of radiofrequency phased array hyperthermia," International Journal of Hyperthermia, vol. 25, no. 7, pp. 517-528, Nov. 2009. https://doi.org/10.3109/02656730903287832
  9. J. Vrba, "Evanescent mode applicators for subcutaneous hyperthermia," Biomedical Engineering, IEEE Transactions on, vol. 40, no. 5, pp. 397-407, 1993. https://doi.org/10.1109/10.243422
  10. B. Vrbova, J. Vrba, "Microwave thermotherapy in cancer treatment: Evaluation of homogeneity of sar distribution," Progress in Electromag-netics Research- PIER, vol. 129, pp. 181-195, 2012. https://doi.org/10.2528/PIER12032304
  11. A. Boag, Y. Leviatan, and A. Boag, "Analysis and optimization of waveguide multiapplicator hyperthermia systems," IEEE Transactions on Biomedical Engineering, vol. 40, no. 9, pp. 946-952, 1993. https://doi.org/10.1109/10.245616
  12. M. M. Paulides, J. F. Bakker, L. W. Hofstetter, W. C. M. Numan, R. Pellicer, E. W. Fiveland, M. Tarasek, G. C. Houston, G. C. van Rhoon, D. T. B. Yeo, and G. Kotek, "Laboratory prototype for experimental validation of MR-guided radiofrequency head and neck hyperthermia," Phys Med Biol, vol. 59, no. 9, pp. 2139-2154, May 2014. https://doi.org/10.1088/0031-9155/59/9/2139
  13. S. Kahng, K. Kahng, I. Yang, and T. Park, "A Metamaterial-Based Handset Antenna with the SAR Reduction", J. Electr. Eng. Technol., vol. 9, no. 2, pp. 622-627, Mar. 2014. https://doi.org/10.5370/JEET.2014.9.2.622
  14. H. D. Trefna, J. Vrba, and M. Persson, "Timereversal focusing in microwave hyperthermia for deep-seated tumors," Physics in Medicine and Biology, vol. 55, no. 8, pp. 2167-2185, Apr. 2010. https://doi.org/10.1088/0031-9155/55/8/004
  15. O. Fiser, J. Vrba, "Oriented Array of Waveguide Applicators for Local Thermotherapy", in 8th European Conference on Antennas and Propagation, 2014. EuCAP 2014, 2014, pp. 2169-2172, 2014
  16. O. Fiser, I. Merunka, L. Vojackova, J. Vrba, Utilization of Waveguide Applicators Combination for Electromagntic Field Focusing, In 24th International Conference Radioelektronika 2014. Bratislava: Slovak University of Technology, 2014.
  17. O. Fiser, I. Merunka, L. Vojackova, J. Vrba, „Space Distribution of SAR and Temperature in Human Model with Tumor Using Waveguide Applicator Array, In European Microwave Week 2014 Conference Proceedings. London: Horizon House Publications, 2014, p. 484-487, 2014.
  18. P. Hasgall, E. Neufeld, M. Gosselin, A. Klingenbock, and N. Kuster. (2011) IT'IS database for thermal and electromagnetic parameters of biological tissues. [Online]. Available: www.itis.ethz.ch/database
  19. B. Vrbova, P. Novakova, M. Halfarova, J. Vrba, "Planning of head and neck tumours treatment", in 29th Annual Meeting of the European Society for Hypertermic Oncology, 2014. ESHO 2014, 2014, pp. 32-33
  20. SPEAG AG, SEMCAD X Reference Guide. [Online]. Available: http://download.speag.com/releases/semcad/1485/documentation/SEMCAD-X-Manual.pdf
  21. C. Boutelier, L. Bougues, and J. Timbal, "Experimental study of convective heat transfer coefficient for the human body in water," J Appl Physiol Respir Environ Exerc Physiol, vol. 42, no. 1, pp. 93-100, Jan. 1977.
  22. R. Ortega-Palacios, L. Leija, A. Vera, and M. F. J. Cepeda, "Measurement of breast-tumor phantom dielectric properties for microwave breast cancer treatment evaluation," 2010, pp. 216-219.
  23. R. Zajicek, L. Oppl, J. Vrba, "Broadband measurement of complex permittivity using reflection method and coaxial probes", Radioengineering, vol. 17, pp 14-19, 2008.