Abstract
In this research, to analyze the concentration of heavy metal ions in water, we tried to find the measuring time at which the faradaic electric currents flowing by the pure oxidation-reduction reaction from the pushing up mercury electrode of the stripping scan square wave voltammetry(SV+SWV) methods system becomes larger than the capacitance electric current. In order to do this, a method for analyzing signals using FPGA has been designed and we conducted 120 experiments using it. As a result, when the frequency of the square wave is 40Hz, The valid potential-current signal was measured from 96.6667% to 96.7155% of the end of the pulse of the forward and reverse, and the optimal signal was measured at 96.6667%. In addition, the experiment was carried out 40 times by changing the pulse height of the square wave from 10Mv to 40Mv. As a result, at a size smaller than 40Mv, there is little change in the magnitude of the potential-current, and an invalid signal was generated when it is out of this size.