Abstract

This study aimed to propose the necessity and system establishment of noise reduction facility via evaluating noise level exposed by the radiographer due to MRI scan. Noise measurements were carried out using at S general hospital in Daejeon using 1.5 Tesla MRI (7 exams) and 3.0 Tesla MRI (16 exams), while using SC-804 noise meter. The measurement distance was from the soundproof door of the MRI room to the radiographer which measured 100cm, and the measurement height, the height to the radiographer's ears when working, 100cm. The noise measured for each exam was an average of three measures per exam which observed the noise occurring in each sequence recorded every 20 seconds. As the results, the maximum of noise exposed by the radiographer is 73.3 dB(A), which is MRCP by the 3.0 Tesla device, and also the maximum of average noise is 66.9(3.1) dB(A), which is Myelogram by the 3.0 Tesla device. Average noise by each device is 61.9(4.1) dB(A) by the 3.0 Tesla device and 52.0(3.1) dB(A) by the 1.5 Tesla device, which comes to the result that the 3.0 Tesla MRI device is about 10 dB(A) degree higher(p <0.001). The noise level exposed by the radiographer does not affect auditory acuity, but the level is able to incur a non auditory effect. The reflect sound can be removed by installing curtains in the rear wall of MRI control room in order to reduce the noises, but, first of all, An institutional system is needed in order to prevent noise.

본 연구는 MRI 검사로 인하여 방사선사가 노출되는 소음의 양을 평가하여 소음저감 시설의 필요성과 제도 마련을 제안하고자 하였다. 소음측정은 대전광역시 S 종합병원의 1.5 Tesla MRI 장비(7개 검사)와 3.0 Tesla MRI 장비(16개 검사)를 대상으로 하였고, 소음측정기는 SC-804를 사용하였다. 소음측정 거리는 MRI 검사실 방음문에서 검사자의 업무 위치까지 100cm 이며, 측정 높이는 업무 시 검사자의 귀 높이 100cm 이다. 검사별 소음측정은 각 검사의 시퀀스(Sequence)마다 발생되는 소음 수치를 관측하여 20초마다 기록하였고 검사별 3회씩 측정하여 평균값을 제시하였다. 연구결과 방사선사가 노출되는 소음의 최댓값은 73.3 dB(A)로 3.0 Tesla 장비에서 시행한 MRCP 검사, 검사별 평균소음의 최댓값은 66.9(3.1) dB(A)로 역시 3.0 Tesla 장비에서 시행한 Myelogram 검사이다. 장비별 평균소음은 3.0 Tesla 장비가 61.9(4.1) dB(A), 1.5 Tesla 장비가 52.0(3.1) dB(A)로 3.0 Tesla MRI 장비가 약 10 dB(A) 정도 높았다(p<0.001). 방사선사가 노출되는 소음의 양은 청력에 영향을 미치는 수준은 아니지만 비청력적영향이 발생할 수 있는 수준이다. 소음을 저감하기 위해 MRI 조정실 후면에 커튼을 설치하여 반사음을 제거할 수 있지만, 제도 마련이 선행되어야 할 것이다.

Keywords

References

  1. 박영혜, "자기공명영상(MRI) 검사 전.후 불안민감성 제수준에 따른 폐쇄공포, 소음민감성 및 활력징후 비교," 성인간호학회지, 제20권, 제6호, pp.950-959, 2008.
  2. 김종봉, 자기공명영상 검사시 환자의 불안수준에 영향을 미치는 요인, 전남대학교, 석사학위논문, 2005.
  3. 조준철, 자기공명영상(MRI)과 체질량지수(BMI)와의 상관관계에 관한 연구, 을지대학교, 석사학위논문, 2009.
  4. J. C. Melendez and E. McCrank, "Anxiety-related reactions associated with magnetic resonance imaging examinations," JAMA, Vol.270, No.6, pp.745-747, 1993. https://doi.org/10.1001/jama.1993.03510060091039
  5. K. J. Murphy and J. A. Brunberg, "Adult claustrophobia, anxiety and sedation in MRI," Magnetic resonance imaging, Vol.15, No.1, pp.51-54, 1997. https://doi.org/10.1016/S0730-725X(96)00351-7
  6. M. Killion, E. DeVilbiss, and J. Stewar, "An earplug with uniform 15-dB attenuation," Hearing Journal, Vol.41, No.5, pp.14-17, 1988.
  7. 김영수, 이성수, 안규동, 이병국, "자기공명영상검사 환자의 폐쇄공포증에 관한 연구," 순천향산업의학, 제7권, 제1호, pp.13-24, 2001.
  8. T. Peuvrel, Faculty of Engineering and Applied Science Institute of Sound and Vibration Research, University of Southampton, Ph.D(Degree of Master) Thesis, 2002.
  9. 법제처, 산업안전보건기준에 관한 규칙, 제4장 소음 및 진동에 의한 건강장해의 예방, 제1절 통칙 제512조, 2016.
  10. A. Smith, "Noise, performance efficiency and safety," J. of Contents Association, Vol.62, No.1, pp.1-5, 1990.
  11. 송현옥, 임청환, "자기공명영상검사 시 발생하는 소음이 환자와 방사선사에 미치는 영향에 관한 연구," 한국콘텐츠학회 춘계종합학술대회논문지, pp.153-154, 2015.
  12. 정성택, 경사 자계 펄싱으로 인한 뇌 기능 영상에서의 소음 영향 연구, 한국과학기술원, 박사학위논문, 1999.
  13. 환경용어연구회, 환경공학용어사전(소음의 생리기능에 대한 영향), 성안당, 1996.
  14. 박형숙, "소음.진동에 의한 가축피해 사례분석," 환경영향평가, 제17권, 제6호, pp.381-391, 2008.
  15. 류일선, "소음.진동으로 인한 가축피해실태 및 사례," 대한수의사회지, 제38권, 제5호, pp.469-477, 2002.
  16. 김정만, "소음이 건강에 미치는 영향 : 산업현장을 중심으로," 대한환경공학회지, 제29권, 제2호, pp.131-137, 2007.
  17. F. Hennel, F. Girard, and T. Loenneker, "Silent-MRI with soft gradient pulses," Magn Reson Med, Vol.42, No.1, pp.6-10, 1999. https://doi.org/10.1002/(SICI)1522-2594(199907)42:1<6::AID-MRM2>3.0.CO;2-D
  18. 김준석, 일개 석유화학 공장의 소음 노출평가 및 개선에 따른 소음저감 효율 비교, 대구가톨릭대학교 대학원, 석사학위논문, 2008.
  19. 구현우, "소아에서 고자장 자기공명영상," 대한의사협회지, 제53권, 제12호, pp.1093-1102, 2010. https://doi.org/10.5124/jkma.2010.53.12.1093
  20. 윤용학, 윤희정, 이성국, 김경희, 권기홍, "MRI 검사 시 음악요법을 동반한 의사소통이 환자의 불안에 미치는 영향," 한국산학기술학회논문지, 제17권, 제3호, pp.93-102, 2016. https://doi.org/10.5762/KAIS.2016.17.3.93
  21. 홍준의, 최후남, 고현덕, 김태일, 살아있는 과학교과서, 휴머니스트, 2011.
  22. 식품의약품안전처 의교기기안전국, 의료기기 기준규격 별표 2 (식품의약품안전처 고시 제2014-155호), 2014.
  23. 보건복지부, 특수의료장비의 설치 및 운영에 관한 규칙(보건복지부령 제339호), 2016.