Abstract
This convergence study analyzed the effectiveness of digital radiography system of copper(Cu) filter in the added filtration for the removal of lower energy radiation through dose and image evaluation. We were analyzed from April to June 2015 result of the examination. Cu filter was applied to each non, 0.1, 0.2, 0.3 mm according to change of kV and mAs and doses were evaluated. Image quality was evaluated by PSNR, MAE, MSE, CNR, SNR and qualitative analysis was performed by seven items for resolution and contrast from chest x-ray criteria of national cancer checkup. The absorbed doses with Cu were lowered by 16-88 % than non-filter but the gaps decreased as kV increased. PSNR were over 30 dB and all significant and CNR and SNR were superior with non-filter but in the qualitative analysis, there were different statistical significant according to each item. The score of 0.1 mm filter was high at pulmonary blood vessel observation and in the 0.3 mm Cu, there were no statistical signigicant except high density and full of air portion. Cu filter can improve image quality with lower radiation dose using better radiation quality and correction power at digital radiography system.
본 연구는 저에너지 방사선을 제거하기 위해 사용되는 구리 여과판의 디지털 방사선 장비에서 유용성을 분석하고자 선량, 화질평가의 융복합 연구를 실시하였다. 실험기간은 2015년 4월부터 6월까지 이었다. 관전압과 관전류 변화에 따른 구리 여과판 사용 전, 0.1, 0.2, 0.3 mm 의 선량을 평가하였다. 화질평가는 PSNR, MAE, MSE, CNR, SNR 정성적 평가로는 국가암검진 흉부 평가표의 해상, 대조도 평가 7문항을 이용하였다. 흡수선량은 16-88 % 정도 여과판 사용 전에 비해 측정치 낮아졌고, 관전압이 높아짐에 따라 격차가 작아짐을 알 수 있었다. PSNR은 30 dB이상으로 모두 의미 있는 수치였고, CNR, SNR은 여과판 미사용 시 우수했으나 정성적 평가에서는 항목마다 통계적 유의성이 달랐다. 0.1 mm 여과판에서는 폐혈관 관찰 부위에서 측정치가 높고, 0.3 mm도 밀도가 높고 공기가 많은 부위를 제외하고 통계적으로 유의하지 않았다. 구리 여과판은 좋은 선질과 디지털촬영 장비의 장점인 보정 능력 이용하여 더 적은 선량으로 화질을 향상 시킬 수 있다.