DOI QR코드

DOI QR Code

Influence of Iodinated Magnetic Resonance Contrast Media and Isotope 99mTc on Changes of Computed Tomography Number

  • Kim, Sang-Beom (Department of Neurology, Kyung Hee University Hospital at Gangdong) ;
  • Lee, Jin-Hyeok (Department of Radiology, Korea Veterans Hospital) ;
  • Ahn, Jae-Ouk (Department of Medical IT Engineering, Soonchunhyang University) ;
  • Cho, Jae-Hwan (Department of International Radiological Science, Hallym University of Graduate Studies)
  • 투고 : 2015.06.09
  • 심사 : 2015.06.25
  • 발행 : 2015.09.30

초록

The purpose of the study was to identify how isotope and magnetic resonance imaging (MRI) contrast media impact on noise to computed tomography (CT) examination. For the study, divide the phantoms to two groups: 1) saline, saline + different kinds of contrast agent without $^{99m}Tc$ administration; 2) $^{99m}Tc$ administration: saline, saline + different kinds of contrast agent with $^{99m}Tc$ administration. CT contrast agent was used for Iopamidol$^{(R)}$ and Dotarem. And MRI contrast agent was used for Primovist$^{(R)}$ and Gadovist$^{(R)}$. To obtain an image, we used CT scanner. With an obtained image, we set the $1cm^2$ region of interest in the middle of bottle to measure the noise and CT number. As a result, there was no difference in CT number before and after inserting $^{99m}Tc$ into all contrast media including Normal Saline. However, when it comes to Noise, there was a difference before and after inserting $^{99m}Tc$ into every contrast media except MRI contrast media such as Primovist$^{(R)}$ and Gadovist$^{(R)}$.

키워드

참고문헌

  1. S. M. Lee, S. J. Kim, H. Guo, and Y. K. Kang, Biomaterials Research, 12, 23 (2008).
  2. J. H. Lee, H. K. Lee, J. H. Cho, and M. J. Cheon, J. Magn. 19, 372 (2014). https://doi.org/10.4283/JMAG.2014.19.4.372
  3. J. H. Cho, J. H. Lee, C. S. Park, S. Y. Lee, J. Lee, D. H. Moon, and H. K. Lee, J. Magn. 19, 248 (2014). https://doi.org/10.4283/JMAG.2014.19.3.248
  4. Y. C. Heo, H. K. Lee, C. S. Park, and J. H. Cho, J. Magn. 20, 40 (2015). https://doi.org/10.4283/JMAG.2015.20.1.040
  5. K. S. Kwon, and J. S. Chung, J. Korean Biol. Nurs. Sci. 6, 57 (2004).
  6. T. Ozulker, F. Ozulker, and T. Ozpacaci, Hell. J. Nucl. Med. 12, 287 (2009).
  7. I. Podreka, E. Suess, G. Goldenberg, M. Steiner, T. Brucke, C. Muller, W. Lang, R. D. Neirinckx, and L. Deecke, J. Necl. Med. 28, 1657 (1987).
  8. M. Dondi, R. Franchi, M. Levorato, A. Zuccala, R. Gaggi, M. Mirelli, A. Stella, F. Marchetta, F. Losinno, and N. Monetti, J. Necl. Med. 30, 615 (1989).
  9. P. Rask, K. Karp, B. Edlund, P. Eriksson, T. Mooe, and U. Wiklund, J. Necl. Med. 35, 983 (1994).
  10. J. G. Bong, H. J. Kim, S. G. So, J. D. Lee, and S. I. Gwon, J. Biomed. Eng. Res. 20, 107 (1999).
  11. Y. H. Oh, Medical Policy Forum. 10, 46 (2013).
  12. J. C. Jang and Y. S. Jang, J. Korean Soc. Radiol. 30, 113 (1994). https://doi.org/10.3348/jkrs.1994.30.1.113
  13. L. W. Goldman, J. Nucl. Med. Technol. 35, 115 (2007). https://doi.org/10.2967/jnmt.107.042978
  14. H. Johnston, M. Hilts, and A. Jirasek, Med. Phys. 42, 1666 (2015). https://doi.org/10.1118/1.4914419
  15. S. Radzi, G. Cowin, M. Robinson, J. Pratap, A. Volp, M. A. Schuetz, and B. Schmutz, Quant Imaging Med. Surg. 4, 163 (2014).
  16. S. C. Han, Y. E. Chung, Y. H. Lee, K. K. Park, M. J. Kim, and K. W. Kim, AJR Am. J. Roentgenol. 203, 788 (2014). https://doi.org/10.2214/AJR.13.10980