초록

Objective : To improve pedicle screw placement accuracy with minimal radiation and low cost, we developed specially designed K-wire with a marker. To evaluate the accuracy of thoracolumbar pedicle screws placed using the novel guide-pin and portable X-rays. Methods : Observational cohort study with computerized tomography (CT) analysis of in vivo and in vitro pedicle screw placement. Postoperative CT scans of 183 titanium pedicle screws (85 lumbar and 98 thoracic from T1 to L5) placed into 2 cadavers and 18 patients were assessed. A specially designed guide-pin with a marker was inserted into the pedicle to identify the correct starting point (2 mm lateral to the center of the pedicle) and aiming point (center of the pedicle isthmus) in posteroanterior and lateral X-rays. After radiographically confirming the exact starting and aiming points desired, a gearshift was inserted into the pedicle from the starting point into the vertebral body through the center of pedicle isthmus. Results : Ninety-nine percent (181/183) of screws were contained within the pedicle (total 183 pedicle screws : 98 thoracic pedicle screws and 85 lumbar screws). Only two of 183 (1.0%) thoracic pedicle screws demonstrated breach (1 lateral in a patient and 1 medial in a cadaver specimen). None of the pedicle breaches were associated with neurologic or other clinical sequelae. Conclusion : A simple, specially designed guide-pin with portable X-rays can provide correct starting and aiming points and allows for accurate pedicle screw placement without preoperative CT scan and intraoperative fluoroscopic assistance.

키워드

참고문헌

  1. Amiot LP, Lang K, Putzier M, Zippel H, Labelle H : Comparative results between conventional and computer-assisted pedicle screw installation in the thoracic, lumbar, and sacral spine. Spine (Phila Pa 1976) 25 : 606-614, 2000 https://doi.org/10.1097/00007632-200003010-00012
  2. Belmont PJ Jr, Klemme WR, Dhawan A, Polly DW Jr : In vivo accuracy of thoracic pedicle screws. Spine (Phila Pa 1976) 26 : 2340-2346, 2001 https://doi.org/10.1097/00007632-200111010-00010
  3. Brown CA, Lenke LG, Bridwell KH, Geideman WM, Hasan SA, Blanke K : Complications of pediatric thoracolumbar and lumbar pedicle screws. Spine (Phila Pa 1976) 23 : 1566-1571, 1998 https://doi.org/10.1097/00007632-199807150-00012
  4. Cotrel Y, Dubousset J, Guillaumat M : New universal instrumentation in spinal surgery. Clin Orthop Relat Res 227 : 10-23, 1988
  5. Esses SI, Sachs BL, Dreyzin V : Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members. Spine (Phila Pa 1976) 18 : 2231-2238; discussion 2238-2239, 1993 https://doi.org/10.1097/00007632-199311000-00015
  6. Farber GL, Place HM, Mazur RA, Jones DE, Damiano TR : Accuracy of pedicle screw placement in lumbar fusions by plain radiographs and computed tomography. Spine (Phila Pa 1976) 20 : 1494-1499, 1995 https://doi.org/10.1097/00007632-199507000-00010
  7. Ferrick MR, Kowalski JM, Simmons ED Jr : Reliability of roentgenogram evaluation of pedicle screw position. Spine (Phila Pa 1976) 22 : 1249-1252; discussion 1253, 1997 https://doi.org/10.1097/00007632-199706010-00016
  8. Hyun SJ, Kim YJ, Cheh G, Yoon SH, Rhim SC : Free hand pedicle screw placement in the thoracic spine without any radiographic guidance : technical note, a cadaveric study. J Korean Neurosurg Soc 51 : 66-70, 2012 https://doi.org/10.3340/jkns.2012.51.1.66
  9. Kim KD, Johnson JP, Babbitz JD : Image-guided thoracic pedicle screw placement : a technical study in cadavers and preliminary clinical experience. Neurosurg Focus 10 : E2, 2001
  10. Kim YJ, Lenke LG, Bridwell KH, Cho YS, Riew KD : Free hand pedicle screw placement in the thoracic spine : is it safe? Spine (Phila Pa 1976) 29 : 333-342; discussion 342, 2004 https://doi.org/10.1097/01.BRS.0000109983.12113.9B
  11. Kim YJ, Lenke LG, Cheh G, Riew KD : Evaluation of pedicle screw placement in the deformed spine using intraoperative plain radiographs : a comparison with computerized tomography. Spine (Phila Pa 1976) 30 : 2084-2088, 2005 https://doi.org/10.1097/01.brs.0000178818.92105.ec
  12. Kuntz C 4th, Maher PC, Levine NB, Kurokawa R : Prospective evaluation of thoracic pedicle screw placement using fluoroscopic imaging. J Spinal Disord Tech 17 : 206-214, 2004 https://doi.org/10.1097/00024720-200406000-00007
  13. Laine T, Makitalo K, Schlenzka D, Tallroth K, Poussa M, Alho A : Accuracy of pedicle screw insertion : a prospective CT study in 30 low back patients. Eur Spine J 6 : 402-405, 1997 https://doi.org/10.1007/BF01834068
  14. Lee CH, Hyun SJ, Kim YJ, Kim KJ, Jahng TA, Kim HJ : Accuracy of free hand pedicle screw installation in the thoracic and lumbar spine by a young surgeon : an analysis of the first consecutive 306 screws using computed tomography. Asian Spine J 8 : 237-243, 2014 https://doi.org/10.4184/asj.2014.8.3.237
  15. Li G, Lv G, Passias P, Kozanek M, Metkar US, Liu Z, et al. : Complications associated with thoracic pedicle screws in spinal deformity. Eur Spine J 19 : 1576-1584, 2010 https://doi.org/10.1007/s00586-010-1316-y
  16. Liljenqvist UR, Halm HF, Link TM : Pedicle screw instrumentation of the thoracic spine in idiopathic scoliosis. Spine (Phila Pa 1976) 22 : 2239-2245, 1997 https://doi.org/10.1097/00007632-199710010-00008
  17. Liljenqvist UR, Link TM, Halm HF : Morphometric analysis of thoracic and lumbar vertebrae in idiopathic scoliosis. Spine (Phila Pa 1976) 25 : 1247-1253, 2000 https://doi.org/10.1097/00007632-200005150-00008
  18. Lonstein JE, Denis F, Perra JH, Pinto MR, Smith MD, Winter RB : Complications associated with pedicle screws. J Bone Joint Surg Am 81 : 1519-1528, 1999 https://doi.org/10.2106/00004623-199911000-00003
  19. Rajasekaran S, Vidyadhara S, Ramesh P, Shetty AP : Randomized clinical study to compare the accuracy of navigated and non-navigated thoracic pedicle screws in deformity correction surgeries. Spine (Phila Pa 1976) 32 : E56-E64, 2007 https://doi.org/10.1097/01.brs.0000252094.64857.ab
  20. Sarlak AY, Tosun B, Atmaca H, Sarisoy HT, Buluc L : Evaluation of thoracic pedicle screw placement in adolescent idiopathic scoliosis. Eur Spine J 18 : 1892-1897, 2009 https://doi.org/10.1007/s00586-009-1065-y
  21. Suk SI, Kim WJ, Lee SM, Kim JH, Chung ER : Thoracic pedicle screw fixation in spinal deformities : are they really safe? Spine (Phila Pa 1976) 26 : 2049-2057, 2001 https://doi.org/10.1097/00007632-200109150-00022
  22. Weinstein JN, Spratt KF, Spengler D, Brick C, Reid S : Spinal pedicle fixation : reliability and validity of roentgenogram-based assessment and surgical factors on successful screw placement. Spine (Phila Pa 1976) 13 : 1012-1018, 1988 https://doi.org/10.1097/00007632-198809000-00008
  23. West JL 3rd, Ogilvie JW, Bradford DS : Complications of the variable screw plate pedicle screw fixation. Spine (Phila Pa 1976) 16 : 576-579, 1991 https://doi.org/10.1097/00007632-199105000-00016
  24. Yoo JU, Ghanayem A, Petersilge C, Lewin J : Accuracy of using computed tomography to identify pedicle screw placement in cadaveric human lumbar spine. Spine (Phila Pa 1976) 22 : 2668-2671, 1997 https://doi.org/10.1097/00007632-199711150-00016

피인용 문헌

  1. A medium invasiveness multi-level patient’s specific template for pedicle screw placement in the scoliosis surgery vol.16, pp.None, 2015, https://doi.org/10.1186/s12938-017-0421-0
  2. A method to design and manufacture low-cost patient-specific templates for spinal surgery: evaluation of multiple additive manufacturing methods vol.2, pp.4, 2015, https://doi.org/10.2217/3dp-2018-0019
  3. Accuracy and deviation analysis of robot‐assisted spinal implants: A retrospective overview of 105 cases and preliminary comparison to open freehand surgery in lumbar spondylolisthesis vol.17, pp.4, 2015, https://doi.org/10.1002/rcs.2273