Abstract
The C-band compact linear accelerator (linac) is being developed at Dongnam Institute of radiological & Medical Sciences (DIRAMS) for medical and industrial applications. This paper was focused on the output measurement of the electron beam generated from the prototype electron linac. The dose rate was measured in unit of cGy/min per unit pulse frequency according to the IAEA TRS-398 protocol. Exradin-A10 Markus type plane parallel chamber used for the measurement was calibrated in terms of dose to water at the reference depth in water. The beam quality index ($R_{50}$) was determined by the radiochromic film with a solid water phantom approximately due to low energy electrons. As a result, the determined electron beam output was $17.0cGy/(min{\cdot}Hz$. The results were used to monitor the accelerator performance during the development procedure.
의료용 및 산업용으로 활용하기 위한 C-밴드형 콤팩트 선형가속기의 개발이 동남권원자력의학원에서 진행되고 있다. 본 논문에서는 선형가속기 시작품에서 발생한 전자빔의 출력 측정 결과를 보고하고자 한다. 출력 측정은 물흡수선량에 대해 교정된 Exradin-A10 마커스형 평행평판형 전리함을 사용하여 물속 기준 깊이에서 IAEA TRS-398 프로토콜에 따라 흡수선량율을 결정하는 과정으로 진행되었다. 전자선 에너지가 낮은 점으로 인하여 선질지표($R_{50}$)은 필름 측정법을 써서 근사적으로 결정하였다. 결과로서 단위 펄스 진동수당의 선형가속기 전자빔의 출력은 $17.0cGy/(min{\cdot}Hz$로 나타났다. 본 연구의 결과는 개발 중인 전자가속기의 성능 평가 자료로 활용될 것이다.