DOI QR코드

DOI QR Code

Time-Varying Hemodynamic Characteristics Simulation using Computerized Mock Circulatory Loop System with Servo Flow Regulator

  • Moon, Youngjin (Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center) ;
  • Son, Kuk Hui (Department of Thoracic and Cardiovascular Surgery, Gachon University Gil Medical Center, Gachon University) ;
  • Choi, Jaesoon (Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center)
  • Received : 2015.09.30
  • Accepted : 2015.12.22
  • Published : 2015.12.31

Abstract

A mock circulatory loop system has been developed to construct a simulator for trainees in cardiopulmonary bypass systems or to simulate a test environment for cardiac-assist devices. This paper proposes a computerized mock circulatory loop system whose node is modularized by using a servo control flow regulator to simulate dynamic change of the hemodynamic status. To observe the effect of time-varying resistance, one with hemodynamic properties, the proposed system replicates the planned cross-sectional areas of the outlet of a ventricular assist device in terms of voltage input of a servo valve. The experiment is performed (1) for steady-input commands of selected area sizes and (2) for dynamic commands such as monotonous increase and decrease, and oscillatory functions of the voltage input, and a computer program based on LabVIEW (National Instruments, Austin, USA) processes every measured data and control command to the servo valve. The results show that the pressure and flow at the target points with respect to time-varying resistance match intuitive estimation: the pressure at the outlet and the pressure drop between both sides of the valve increased and the flow at the outlet decreased for increased resistance.

Keywords

References

  1. Wieting, D. W., Akers, W., Feola, M., and Kennedy, J. H., "Analysis of a variable volume intra-aortic balloon pump in a mock circulatory system," J. Biomech., vol. 4, pp. 37-44, 1971. https://doi.org/10.1016/0021-9290(71)90014-5
  2. Donovan, F., "Design of a hydraulic analog of the circulatory system for evaluating artificial hearts," Artif. Cell. Blood Sub., vol. 3, pp. 439-449, 1975.
  3. Guyton, A. C., and Hall, J. E., Textbook of medical physiology, 10 ed. Saunders, Philadelphia, ch. vol. 17, pp. 175-182, 2001.
  4. Ferrari, G., De Lazzari, C., Kozarski, M., Clemente, F., Gorc-zynska, K., Mimmo, R., Monnanni, E., Tosti, G., and Guaragno, M., "A hybrid mock circulatory system: testing a prototype under physiologic and pathological conditions," ASAIO j., vol. 48, pp. 487-494, 2002. https://doi.org/10.1097/00002480-200209000-00009
  5. Timms, D., Hayne, M., McNeil, K., and Galbraith, A., "A complete mock circulation loop for the evaluation of left, right, and biventricular assist devices," Artif. organs, vol. 29, pp. 564-572, 2005. https://doi.org/10.1111/j.1525-1594.2005.29094.x
  6. Timms, D. L., Gregory, S. D., Greatrex, N. A., Pearcy, M. J., Fraser, J. F., and Stein- seifer, U., "A compact mock circulation loop for the in vitro testing of cardiovascular devices," Artif. Organs, vol. 35, pp. 384-391, 2010.
  7. Legendre, D., Fonseca, J., Andrade, A., Biscegli, J. F., Manrique, R., Guerrino, D., Prakasan, A. K., Ortiz, J. P., and Lucchi, J. C., "Mock circulatory system for the evaluation of left ventricular assist devices, endoluminal prostheses, and vascular diseases," Artif. Organs, vol. 32, pp. 461-467, 2008. https://doi.org/10.1111/j.1525-1594.2008.00569.x
  8. Wu, Y., Allaire, P. E., Tao, G., and Olsen, D., "Modeling, estimation, and control of human circulatory system with a left ventricular assist device," IEEE Trans. Contr. Syst. Tech., vol. 15, pp. 754-767, 2007. https://doi.org/10.1109/TCST.2006.890288
  9. Conlon, M. J., Russell, D. L., and Mussivand, T., "Development of a mathematical model of the human circulatory system," Ann. Biomed. Eng., vol. 34, pp. 1400-1413, 2006. https://doi.org/10.1007/s10439-006-9164-y
  10. Geertsema, A. A., Rakhorst, G., Mihaylov, D., Blanksma, P. K., and Verkerke, G. J., "Development of a numerical simulation model of the cardiovascular system," Artif. organs, vol. 21, pp. 1297-1301, 1997.
  11. Baloa, L., Boston, J., and Antaki, J., "Elastance-based control of a mock circulatory system," Ann. Biomed Eng., vol. 29, pp. 244-251, 2001. https://doi.org/10.1114/1.1355275
  12. Liu, Y., Allaire, P., Wood, H., and Olsen, D., "Design and initial testing of a mock human circulatory loop for left ventricular assist device performance testing," Artif. organs, vol. 29, pp. 341-345, 2005. https://doi.org/10.1111/j.1525-1594.2005.29058.x
  13. Hwang, C., Jeong, G. S., Jung, M. W., Kang, J. S., Lee, J. J., Park, Y. D., Lee, K. B., and Sun, K., "Development of wearable electropneumatic biventricular assist device," ASAIO J., vol. 51, pp. 38A, 2005.
  14. Merritt, H. E., Hydraulic control systems, Wiley New York, 1967.