초록

This paper presents a new method to determine sensitivity coefficients of temperature and humidity of torque transducers by using a natural and seasonal variation of ambient conditions at the laboratory. We had measured the sensitivities of the torque transducers over almost one year using the KRISS 2 kN m torque standard machine. The sensitivity data acquired at various ambient conditions were processed using our measurement model to extract the sensitivity coefficients of temperature and humidity simultaneously with high precision. A comparison with a previous method using an environmental control chamber was carried out to test the feasibility of using our new method. Two results agreed within the uncertainty. We revealed that the torque measuring errors could be 8 times higher than the measurement and calibration capability of KRISS torque standard machine if the sensitivity changes due to the temperature and humidity are not properly corrected during a calibration.

키워드

참고문헌

  1. Park, Y. K., Kim, M. S., Kim, J. H., and Kang, D. I., "Torque Standards and Evaluation," J. Korean Soc. Precis. Eng., Vol. 21, No. 9, pp. 26-31, 2004.
  2. Kim, M. S., Park, Y. K., Kim, J. H., and Kang, D. I., "2 kNm Deadweight Torque Standard Machine in KRISS," Proc. of Autumn Conference, pp. 656-659, 2004.
  3. Sanponpute, T. and Arksonnarong, N., "Temperature and Humidity Dependence on Stability of Torque Measuring Devices," Proc. of IMEKO 22nd TC3 conference, Paper No. 22, 2014.
  4. Roske, D., "Final Report on the Torque Key Comparison CCM.T-K1," Metrologia, Vol. 46, Paper No. 07002, 2009.
  5. Bruge, A., "Simplified Measurements of the Humidity Coefficient of Torque Transducers in Calibration Laboratories," Acta IMEKO, Vol. 3, No. 2, pp. 32-38, 2014.
  6. BIPM, IEC, IFCC, ILAC, ISO, IUPAC, IUPAP and OIML, "Supplement 1 to the 'Guide to the Expression of Uncertainty in Measurement'- Propagation of Distributions using a Monte Carlo Method," Joint Committee for Guides in Metrology, JCGM, Vol. 101, 2008.