Optimal Exposure Conditions according to Detector Type in Chest Digital Radiography

디지털흉부X선촬영에서 검출기 방식에 따른 최적의 노출조건

  • Lee, Won-Jeong (Department of Radiology, Daejeon Health Sciences College)
  • 이원정 (대전보건대학교 방사선학과)
  • Received : 2015.08.25
  • Accepted : 2015.12.24
  • Published : 2015.12.31

Abstract

The aim of this study was to set up the optimal exposure condition according to detector type considering image quality (IQ) with radiation dose in chest digital radiography. We used three detector type such as flat-panel detector (FP) and computed radiography (CR), and charge-coupled device (CCD). Entrance surface dose (ESD) was measured at each exposure condition combined tube voltage with tube current using dosimeter, after attaching on human phantom, it was repeated 3 times. Phantom images were evaluated independently by three chest radiologists after blinding image informations. Standard exposure condition using each institution was 117 kVp-AEC at FP and 117 kVp-8 mAs at CR, and 117 kVp-8 mAs at CCD. Statistical analysis was performed by One way ANOVA (Dunnett T3 test) using SPSS ver. 19.0. In FP, IQ scores were not significant difference between 102 kVp-4 mAs and 117 kVp-AEC (28.4 vs. 31.1, p=1.000), even though ESD was decreased up to 50% ($62.3{\mu}Gy$ vs. $125.1{\mu}Gy$). In CR, ESD was greatly decreased from 117 kVp-8 mAs to 90 kVp-8 mAs without significant difference of IQ score (p=1.000, 24.6 vs. 19.5). In CCD, IQ score of 117 kVp-8 mAs was similar with 109 kVp-8 mAs (29.6 vs. 29.0), with decreasing from $320.8{\mu}Gy$ to $284.7{\mu}Gy$ (about 11%). We conclude that optimal x-ray exposure condition for chest digital radiography is 102 kVp-4 mAs in FP and 90 kVp-8 mAs in CR, and 109 kVp-8 mAs in CCD.

Keywords

References

  1. Choi, JI. Na, DG. Kim, HH. Shin, YM. Ahn, KJ. Lee, JY.(2004),"Quality control of medical imaging."J Korean Radiol Soc, 50:317-331 https://doi.org/10.3348/jkrs.2004.50.5.317
  2. Strotzer, M. Volk, M. Reiser, M. Lenhart. M. Manke, C. Gmeinwieser, J. Holzknecht, N. Link, J. Feuerbach, S.(2000), "Chest radiography with a large-area detector based on cesium-iodide/amorphous-silicon technology: image quality and dose requirement in comparison with an asymmetric screen-film system."J Thorac Imaging, 15(3):157-161 https://doi.org/10.1097/00005382-200007000-00001
  3. Park, BR. Sung, DW.(2010),"A comparative study of image quality and radiation dose with changes in tube voltage and current for a digital chest radiography."J Korean Radiol Soc, 62:131-137 https://doi.org/10.3348/jksr.2010.62.2.131
  4. Chung, MJ. Lee, KY. Lee, BY. Kim, HJ. Lim, HK.(2010),"Diagnostic reference level of patient dose during a plain chest radiography examination in Korea. J Korean Radiol Soc, 62:523-528 https://doi.org/10.3348/jksr.2010.62.6.523
  5. Seibert, JA.(2008),"Digital radiography: image quality and radiation dose. Health Phys, 95(5):586-598 https://doi.org/10.1097/01.HP.0000326338.14198.a2
  6. Strotzer, M. Volk, M. Frund, R. Hamer, O. Zorger, N. Feuerbach, S.(2002),"Routine chest radiography using a flat-panel detector: image quality at standard detector dose and 33% dose reduction. AJR Am J Roentgenol, 178(1):169-171 https://doi.org/10.2214/ajr.178.1.1780169
  7. Uffmann, M. Schaefer-Prokop, C.(2009), "Digital radiography: The balance between image quality and required radiation dose. Eur J Radilo, 72:202-208 https://doi.org/10.1016/j.ejrad.2009.05.060
  8. Launders, JH. Cowen, AR. Bury, RF. Hawkridge, P.(2001),"Towards image quality, beam energy and effective dose optimisation in digital thoracic radiography. Eur Radiol, 11:870-875 https://doi.org/10.1007/s003300000525
  9. Lee, WJ. Park, JS. Kim, SJ. Ko, KS. Chu, SD. Park, SY. et al.(2010),"The first report on evaluation the thoracic radiology of the medical institution for pneumoconiosis in Korea. J Korean Radiol Soc, 63(5):431-438 https://doi.org/10.3348/jksr.2010.63.5.431
  10. Ganten, M. Radeleff, B. Kampschulte, A. Daniels, MD. Kauffmann, GW. Hansmann, J.(2003),"Comparing Image Quality of Flat-Panel Chest Radiography with Storage Phosphor Radiography and Film-Screen Radiography. AJR Am J Roentgenol, 181:171-176 https://doi.org/10.2214/ajr.181.1.1810171
  11. Uffmann, M. Neitzel, U. Prokop, M. Kabalan, N. Weber, M. Herold, CJ. Schaefer-Prokop, C.(2005), "Flat-panel-detector chest radiography: Effect of tube voltage on image quality. Radiology, 235:642-650 https://doi.org/10.1148/radiol.2352031730
  12. van Soldt, RT. Zweers, D. van den Berg, L. Geleijns, J. Jansen, JT. Zoetelief, J.(2003), "Survey of posteroanterior chest radiography in The Netherlands: patient dose and image quality. Br J Radiol, 76: 398-405 https://doi.org/10.1259/bjr/76222078
  13. Lee, WJ. Kim, SJ. Park, JS. Seon, JR.(2012),"Relationship of image quality and radiation dose for chest radiography in the medical institution for pneumoconiosis: a comparison to the Korean Diagnostic Reference Level. J Korean Radiol Soc, 66:255-262 https://doi.org/10.3348/jksr.2012.66.3.255
  14. Hennings, SP. Garmer, M. Jaeger, HJ. et al.(2001),"Digital chest radiography with a large-area flat-panel silicon X-ray detector: clinical comparison with conventional radiography. Eur Radiol, 11:1688-1696 https://doi.org/10.1007/s003300000795
  15. Kroft, LJ. Veldkamp, WJ. Mertens, BJ. Boot, MV. Geleijns, J.(2005),"Comparison of eight different digital chest radiography systems: variation in detection of simulated chest disease. AJR Am J Roentgenol, 185(2):339-346 https://doi.org/10.2214/ajr.185.2.01850339
  16. Redlich, U. Hoeschen, C. Doehring, W.(2005),"Assessment and optimisation of the image quality of chest-radiography systems. Radiat Prot Dosimetry, 114:264-268 https://doi.org/10.1093/rpd/nch559
  17. Honey, ID. Mackenzie, A. Evans, DS.(2005),"Investigation of optimum energies for chest imaging using film-screen and computed radiography. Br J Radiol, 78:422-427 https://doi.org/10.1259/bjr/32912696
  18. Launders, JH. Cowen, AR. Bury, RF. Hawkridge, P.(1998),"A case study into the effect of radiographic factors on image quality and dose for a selenium based digital chest radiography system. Radiation Protection Dosimetry, 80:279-282 https://doi.org/10.1093/oxfordjournals.rpd.a032523
  19. Muhogora, W. Padovani, R. Msaki, P.(2011),"Initial quality performance results using a phantom to simulate chest computed radiography. J Med Phys, 36(1):22-28 https://doi.org/10.4103/0971-6203.75468