심근 관류 스펙트에서 Half-Time Scan과 새로운 재구성법이 적용된 정상군 데이터를 기반으로 한 정량적 분석 결과의 차이 비교

The Effects of Discrepancy in Reconstruction Algorithm between Patient Data and Normal Database in AutoQuant Evaluation: Focusing on Half-Time Scan Algorithm in Myocardial SPECT

  • 이형진 (서울대학교병원 핵의학과) ;
  • 도용호 (서울대학교병원 핵의학과) ;
  • 조성욱 (서울대학교병원 핵의학과) ;
  • 김진의 (서울대학교병원 핵의학과)
  • Lee, Hyung-Jin (Department of Nuclear Medicine, Seoul National University Hospital) ;
  • Do, Yong-Ho (Department of Nuclear Medicine, Seoul National University Hospital) ;
  • Cho, Seong-Wook (Department of Nuclear Medicine, Seoul National University Hospital) ;
  • Kim, Jin-Eui (Department of Nuclear Medicine, Seoul National University Hospital)
  • 투고 : 2014.03.28
  • 심사 : 2014.04.21
  • 발행 : 2014.05.23

초록

최근 장비회사에서 제공하고 있는 half-time 기법과 recon algorithm은 검사시간의 단축과 영상의 질은 향상되지만, 핵의학 검사의 특성 상 정량적 분석 시 자동으로 이뤄지는 검사의 경우 기존의 결과값과 어떻게 달라지는지에 대하여 언급된 바는 없었다. 심근 관류 SPECT는 과거 다른 영상법과 마찬가지로 판정의 주관성이 문제로 제기되어 왔으나, 근래에는 다양한 자동 정량화 분석법이 상용화되어 보급되고 있어 이러한 문제를 극복하였다. 그러나 $^{13}N$을 이용한 PET의 검사와는 달리 자동 정량화 프로그램을 통해 얻어지는 관류정량값의 경우 실제 관류량을 측정하는 것이 아니고 최대 섭취화소에 대한 상대적 섭취값을 표시하는 간접적 또는 반정량적 측정치임을 고려해야 한다. 이 연구에서는 정량적 게이트 심근 관류 SPECT에서 정상군 데이터를 재편집하였을 때 또한 그 데이터가 어떠한 성질을 가지고 있느냐에 따른 결과값의 변화를 조사하였다. 정량 분석 시 사용되는 Cedars사의 AutoQuant 프로그램은 정상군 데이터를 다시 편집할 수 있는 기능이 있어서 각 병원의 검사법에 맞게 정상군 데이터를 삽입하여 사용이 가능하다. SRS, SSS, Extent, TPD와 같은 결과 지표값이 본원에서 재편집한 정상군 데이터를 이용하여 보았더니 약 30% 낮은 값을 보였다. 재편집한 정상군 데이터의 성질은 한국인이면서 $^{201}Tl$, $^{99m}Tc-MIBI$를 사용하는 dual isotope 방법이었고, half-time method인 Philips사의 Astonish를 적용한 정상군 데이터였다. 본 조사를 통하여 임상적으로 어떠한 결과를 내리는 것은 성급한 일이지만, 정상군 데이터를 자체적으로 확보하고 이를 자동 정량화방법에 이용한다면 게이트 심근관류 SPECT의 임상적 유용성이 더욱 커질 것으로 기대해본다.

Purpose: The new reconstruction algorithms (NRA) provided by vendor aim to shorten the acquisition scan time. Whereas depending on the installed version AutoQuant program used for myocardial SPECT quantitative analysis did not contain the normal data that NRA is applied. Thus, the purpose of this paper is to compare the results according to AutoQuant versions in myocardial SPECT applied NRA and half-time scan (HT). Materials and Methods: Rest Tl and stress MIBI data of total 80 (40 men, 40 women) patients were gathered. Data were applied HT acquisition and ASTONISH (Philips) software which is NRA. Modified autoquant of SNUH and old version of AutoQuant (full-time scan) provided by company were compared. Comparison groups were classified as coronary artery disease (CAD), 24 hrs delay and almost normal patients who have a simple pain patient. Perfusion distribution aspect, summed stress score (SSS), summed rest score (SRS), extent and total perfusion deficit (TPD) of each 25 patient who have above diseases were compared and evaluated. Results: The case of CAD, when using re-edited AutoQuant (HT) SSS and SRS showed about 30% reduction (P<0.0001), Extent showed about 38% reduction and TPD showed about 30% reduction in the tendency (P<0.0001). In the score of the perfusion, especially on the part of infero-medium, infero-apical, lateral-medium and lateral-apical regions were the biggest change. The case of the 24 hrs delay patient SRS (P=0.042), Extent (P=0.018) and TPD (P=0.0024) showed about 13-18% reduction. And the case of simple pain patient, comparison of 4 results showed about 5-7% reduction. Conclusion: This study was started based on expectation that results could be affected by normal patient data. Normal patient data is possible to change by race and gender. It was proved that combination of new reconstruction algorithm for reducing scan time and analysis program according to scan protocol with NRA could also be affected to results. Clinical usefulness of gated myocardial SPECT is possibly increased if each hospital properly collects normal patient data for their scan acquisition protocol.