Abstract
Purpose The nuclear medicine examination, there is a difficulty to carry out the inspection of both on the day of residual isotope due to the half-life. In this study, by studying the mutual influence and $^{18}F$-FDG of $^{99m}TcO_4{^-}$, I would like to explain the matters to be considered in the case of performing the same day. Materials and Methods With the NEMA-1994 Phantom, and experiments were performed 3 times. Create a 1: 4 Background ratio HOT and the $^{99m}TcO_4{^-}$ The first experiment: After underwent SPECT in INFINIA (GE Healthcare, MI, USA), and were injected with $^{18}F$-FDG 37 MBq in the Background area, 13 once for 60 minutes under the same conditions was time Scan. Create a 1: 4 Background ratio HOT and the $^{18}F$-FDG second is: The Scan in PET/CT Discovery 600 (GE Healthcare, MI, USA), and 148 MBq after injection $^{99m}TcO_4{^-}$ the Background area, once for 60 minutes, 6 under the same conditions was time Scan. Create a 1: 4 Background ratio HOT and the $^{18}F$-FDG experiments las, increments of 296 MBq and 148 MBq the 1 Bed Scan after $^{99m}TcO_4{^-}$, was 1 Bed Scan under the same conditions. Non BKG area and HOT, I was measured comparing the Total Counts and SNR or CNR. Results Showed a significant difference in the ratio CNR of enforcement during SPECT $^{18}F$-FDG is, (p>0.05). The $^{99m}TcO_4{^-}$ was no significant difference between the SNR ratio of PET / CT at the time of the effective date (p<0.05). I got the results $^{99m}TcO_4{^-}$ that reduce the Total Counts of PET / CT scan. Conclusion If you make a PET / CT scan, may affect the test using the $^{99m}TcO_4{^-}$ up to 12 hours, when it is performed before the $^{99m}TcO_4{^-}$, does not affect the SNR and SUV, PET / CT scan I reduced the detection efficiency. The inspection of day, we'd like to recommend a way to complement the detection efficiency to increase the inspection time of PET / CT in move forward the inspection using the $^{99m}TcO_4{^-}$.
핵의학 검사는 반감기로 인한 동위원소의 잔류로 당일에 두 가지 검사를 시행하는데 어려움이 있다. 본 연구는 $^{18}F$-FDG와 $^{99m}TcO_4{^-}$의 상호 영향을 연구함으로써 당일 검사를 진행할 경우 고려될 사항에 대해 알아보고자 한다. NEMA-1994 Phantom을 이용하여 세 번의 실험을 진행했다. 첫 실험은 $^{99m}TcO_4{^-}$을 HOT과 BKG 비 4:1로 만들어 GE사의 INFINIA 장비로 SPECT를 시행하였고, BKG 영역에 $^{18}F$-FDG 37 MBq을 주입 후 동일 조건으로 60분에 1회씩 13회 Scan 했다. 두 번째는 $^{18}F$-FDG를 HOT과 BKG 비 4:1 로 만들어 GE사의 PET/CT Discovery 600 에서 Scan 하였고, BKG 영역에 $^{99m}TcO_4{^-}$을 148 MBq 주입 후 동일 조건으로 60분에 1회씩 6회 Scan 했다. 마지막 실험은 $^{18}F$-FDG를 HOT과 BKG 비 4 : 1로 만들고 1 Bed Scan 후 $^{99m}TcO_4{^-}$을 148 MBq 및 296 MBq씩 증가 시키며 같은 조건으로 1 Bed Scan했다. HOT과 BKG 영역의 비, CNR 혹은 SNR 그리고 총 획득 계수를 측정 후 비교 했다. $^{18}F$-FDG는 SPECT 시행 시 비율 및 CNR에 유의한 차이를 보였다(p>0.05). $^{99m}TcO_4{^-}$는 PET/CT 시행 시 비율 및 SNR이 유의한 차이를 보이지 않았다(p<0.05). $^{99m}TcO_4{^-}$이 PET/CT 검사의 Total Counts를 감소시킨다는 결과를 획득했다. PET/CT 검사를 한 경우 12시간 까지도 $^{99m}TcO_4{^-}$을 이용한 검사에 영향을 미칠 수 있으며, $^{99m}TcO_4{^-}$를 먼저 시행한 경우에는 PET/CT 검사에 SUV 및 SNR에 영향은 없지만 검출 효율을 감소시켰다. 당일 검사 시에는 $^{99m}TcO_4{^-}$을 이용한 검사를 먼저 진행하고 PET/CT의 검사 시간을 늘려 검출 효율을 보완하는 방법을 권장하고자 한다.