DOI QR코드

DOI QR Code

Measurement of Linear Energy Spectra for 135 MeV/u Carbon Beams in HIMAC Using Prototype TEPC

프로토 타입 조직등가비례계수기의 중입자가속기연구소의 135 MeV/u 탄소 이온에 대한 선형에너지 스펙트럼 측정

  • Received : 2014.03.31
  • Accepted : 2014.05.29
  • Published : 2014.05.31

Abstract

TEPC (Tissue Equivalent Proportional Counter) was usually used for high LET radiation dosimetry. We developed a prototype TEPC for micro-dosimetry in the range of $0.2{\sim}300 keV/{\mu}m$. And, the simulated site diameter of the TEPC is $2{\mu}m$, of similar size to a cell nucleus. For purposes of characterization the response for high LET radiation of the TEPC has been investigated under 135MeV/u Carbon ions in HIMAC (Heavy Ion Medical Accelerator). We determined the gas multiplication factor and measured the lineal energy spectrum [yd(y)] of 135 MeV/u Carbon ions. The value of the gas multiplication factor was 315 at 700 V bias voltage. As a result of the experiment, we could more understand the performance of the TEPC for high LET (Linear Energy Transfer) radiation. And the procedure of high LET radiation dosimetry using TEPC is established.

Keywords

References

  1. T. Sakaguchi, T. Doke, N. Hasebe, T. Hayashi, T. Kashiwagi, J. Kikuchi, S. Kono, S. Nagaoka, T. Nakano, T. Takagi, K. Takahashi, and S. Takahashi, "LET distribution measurement with a new real-time radiation monitoring device-III onboard the space shuttle STS-84", Nucl. Instr. Meth., Vol. A 437, pp. 75-87, 1999.
  2. J. E. Keith, G. D. Badhwar, and D. J. Lindstrom, "Neutron spectrum and dose-equivalent in shuttle flights during solar maximum", Nucl. Tracks Radiat. Mes., Vol. 20, pp. 41-47, 1992. https://doi.org/10.1016/1359-0189(92)90083-8
  3. Operational Radiation Safety Program for Astronauts in Low-Earth Orbit: A Basic Frame Work, NCRP, Report No. 142, 2002.
  4. European Communities, "Cosmic radiation exposure of aircraft crew", Rad. Prot., Vol. 140, 2004.
  5. H. H. Rossi and W. Rosenzweig, "A device for the measurement of dose as a function of specific ionisation", Radiology, Vol. 64, 1955.
  6. P. Beck, M. Latoch, S. Rollet, and G. Stehno, "TEPC reference measurements at aircraft altitudes during a solar storm", Adv. in Space Res., Vol. 36, pp. 1627-1633, 2005. https://doi.org/10.1016/j.asr.2005.05.035
  7. D. Zhou, E. Semones, M. Weyland, and S. Johnson, "Radiation measured with TEPC and CR-39 PNTDs in low earth orbit", Adv. in Space Res., Vol. 40, pp. 1571-1574, 2007. https://doi.org/10.1016/j.asr.2006.12.006
  8. U. W. Nam, C. H. Lim, J. J. Lee, J. H. Pyo, B. K. Moon, D. H. Lee, Y. S. Park, H. O. Kim, M. K. Moon, and S. H. Kim, "Development and characterization of tissue equivalent proportional counter for radiation monitoring in international space station", J. Astron. Space Sci. Vol. 30, pp. 107-112, 2013. https://doi.org/10.5140/JASS.2013.30.2.107
  9. http://www.amptek.com/pdf/a250.pdf/ (retrieved on Mar. 28, 2014).
  10. Electrical Grounding Architecture for Unmanned SpaceCcraft, NASA-HDBK-4001, 1998.
  11. D. J. Thomas and M. Burke, "W value measurements for protons in tissue-equivalent gas and its constituent gases", Phys. Med. Biol., Vol. 30, pp. 1201-1213, 1985. https://doi.org/10.1088/0031-9155/30/11/003
  12. S. Tsuda, T. Sato, F. Takahashi, D. Satoh, A. Endo, S. Sasaki, Y. Namito, H. Iwase, S. Ban, and M. Takada, "Measurement of microdosimetric spectra with a wall-less tissueequivalent proportional counter for a 290 MeV/u $^{12}C$ beam", Phys. Med. Biol., Vol. 55, pp. 5089-5101, 2010. https://doi.org/10.1088/0031-9155/55/17/013
  13. ICRU 1983, Microdosimetry, ICRU report 36 (Bethesda, MD:ICRU)

Cited by

  1. Development and Characterization of Multi-Segmented Tissue Equivalent Proportional Counter for Microdosimetry vol.24, pp.2, 2015, https://doi.org/10.5369/JSST.2015.24.2.101
  2. Calibration of TEPC for CubeSat Experiment to Measure Space Radiation vol.32, pp.2, 2015, https://doi.org/10.5140/JASS.2015.32.2.145