Abstract

The purpose of this study were to determine the intra-rater and inter-rater reliability of shoulder passive range of motion measurement using the "Clinometer + bubble level", a smartphone application and to compare with the intra-rater and inter-rater reliability of measurement using a goniometer. Twenty six patients with stroke were recruited for this study. Two raters measured the passive range of motion of four types of shoulder movements (forward flexion; FF, abduction; ABD, external rotation at $90^{\circ}$ abduction; ER90 and internal rotation at $90^{\circ}$ abduction; IR90) using a goniometer and a smartphone to determine within-day inter-rater reliability. A retest session was performed thirty minutes later to determine within-day intra-rater reliability. The reliability was assessed using intraclass correlation coefficients (ICC) and the standard error of measurement (SEM). The ICC (2,1) for the inter-rater reliabilities of the goniometer and smartphone were good in FF and ABD [ICC (2,1)=.75~.87] and excellent in ER90 [ICC (2,1)=.90~.95]. The intra-rater reliabilities for the goniometer and smartphone were good or more than good, with an ICC (3,1) value >.75, the exception was IR90 measured by rater 2 on the smartphone. These results suggest that smartphone could be used as an alternative method tool for measurement of passive shoulder range of motion in patients with stroke.

Keywords

References

  1. Andrews AW, Bohannon RW. Decreased shoulder range of motion on paretic side after stroke. Phys Ther. 1989;69(9):768-772. https://doi.org/10.1093/ptj/69.9.768
  2. Boon AJ, Smith J. Manual scapular stabilization: Its effect on shoulder rotational range of motion. Arch Phys Med Rehabil. 2000;81(7):978-983. https://doi.org/10.1053/apmr.2000.5617
  3. Cadet ER. Evaluation of glenohumeral instability. Orthop Clin North Am. 2010;41(3):287-295. https://doi.org/10.1016/j.ocl.2010.02.005
  4. Cadogan A, Laslett M, Hing W, et al. Reliability of a new hand-held dynamometer in measuring shoulder range of motion and strength. Man Ther. 2011;16(1):97-101. https://doi.org/10.1016/j.math.2010.05.005
  5. Clarkson HM. Musculoskeletal Assessment: Joint range of motion and manual muscle strength. 2nd ed. Philadelphia, PA, Lippincott Williams & Wilkins, 2000:108-120.
  6. Conroy BE, DeJong G, Horn SD. Hospital-based stroke rehabilitation in the united states. Top Stroke Rehabil. 2009;16(1):34-43. https://doi.org/10.1310/tsr1601-34
  7. de Jong LD, Dijkstra PU, Stewart RE, et al. Repeated measurements of arm joint passive range of motion after stroke: Interobserver reliability and sources of variation. Phys Ther. 2012;92(8):1027-1035. https://doi.org/10.2522/ptj.20110280
  8. de Jong LD, Hoonhorst MH, Stuive I, et al. Arm motor control as predictor for hypertonia after stroke: A prospective cohort study. Arch Phys Med Rehabil. 2011;92(9):1411-1417. https://doi.org/10.1016/j.apmr.2011.03.026
  9. de Jong LD, Nieuwboer A, Aufdemkampe G. The hemiplegic arm: Interrater reliability and concurrent validity of passive range of motion measurements. Disabil Rehabil. 2007;29(18):1442-1448. https://doi.org/10.1080/09638280601056145
  10. de Winter AF, Heemskerk MA, Terwee CB, et al. Inter-observer reproducibility of measurements of range of motion in patients with shoulder pain using a digital inclinometer. BMC Musculoskelet Disord. 2004;5:18. https://doi.org/10.1186/1471-2474-5-18
  11. Dijkerman HC, Ietswaart M, Johnston M, et al. Does motor imagery training improve hand function in chronic stroke patients? A pilot study. Clin Rehabil. 2004;18(5):538-549. https://doi.org/10.1191/0269215504cr769oa
  12. Ellenbecker TS, Roetert EP, Piorkowski PA, et al. Glenohumeral joint internal and external rotation range of motion in elite junior tennis players. J Orthop Sports Phys Ther. 1996;24(6):336-341. https://doi.org/10.2519/jospt.1996.24.6.336
  13. Ferriero G, Sartorio F, Foti C, et al. Reliability of a new application for smartphones (DrGoniometer) for elbow angle measurement. PM R. 2011;3(12): 1153-1154. https://doi.org/10.1016/j.pmrj.2011.05.014
  14. Ferriero G, Vercelli S, Sartorio F, et al. Reliability of a smartphone-based goniometer for knee joint goniometry. Int J Rehabil Res. 2013;36:(2):146-151. https://doi.org/10.1097/MRR.0b013e32835b8269
  15. Gajdosik RL, Bohannon RW. Clinical measurement of range of motion. Review of goniometry emphasizing reliability and validity. Phys Ther. 1987;67 (12):1867-1872. https://doi.org/10.1093/ptj/67.12.1867
  16. Green S, Buchbinder R, Forbes A, et al. A standardized protocol for measurement of range of movement of the shoulder using the plurimeter-v inclinometer and assessment of its intrarater and interrater reliability. Arthritis Care Res. 1998;11(1):43-52. https://doi.org/10.1002/art.1790110108
  17. Han YS, Choi JK, Hwang BH, et al. A study on elderly for improvement of usability on smartphone. J Soc e-Bus Stud. 2012;17(1):39-52.
  18. Hardwick DD, Lang CE. Scapular and humeral movement patterns of people with stroke during range of motion exercises. J Neurol Phys Ther. 2011;35(1):18-25. https://doi.org/10.1097/NPT.0b013e318208efa1
  19. Hayes K, Walton JR, Szomor ZR, et al. Reliability of five methods for assessing shoulder range of motion. Aust J Physiother. 2001;47(4):289-294. https://doi.org/10.1016/S0004-9514(14)60274-9
  20. Hebert LJ, Moffet H, McFadyen BJ, et al. Scapular behavior in shoulder impingement syndrome. Arch Phys Med Rehabil. 2002;83(1):60-69. https://doi.org/10.1053/apmr.2002.27471
  21. Hoving JL, Buchbinder R, Green S, et al. How reliably do rheumatologists measure shoulder movement? Ann Rheum Dis. 2002;61(7):612-616. https://doi.org/10.1136/ard.61.7.612
  22. Howard G, Howard VJ, Katholi C, et al. Decline in us stroke mortality: An analysis of temporal patterns by sex, race, and geographic region. Stroke. 2001;32(10):2213-2220. https://doi.org/10.1161/hs1001.096047
  23. Jacquot F, Charpentier A, Khelifi S, et al. Measuring the cobb angle with the iphone in kyphoses: A reliability study. Int Orthop. 2012;36:(8):1655-1660. https://doi.org/10.1007/s00264-012-1579-5
  24. Juul-Kristensen B, Lund H, Hansen K, et al. Test-retest reliability of joint position and kinesthetic sense in the elbow of healthy subjects. Physiother Theory Pract. 2008;24(1):65-72. https://doi.org/10.1080/09593980701378173
  25. Jette DU, Latham NK, Smout RJ, et al. Physical therapy interventions for patients with stroke in inpatient rehabilitation facilities. Phys Ther. 2005;85(3):238-248.
  26. Kisner C, Colby LA. Therapeutic Exercise: Foundations and techniques. 4th ed. Philadelphia, PA, F.A. Davis Co., 2002:34-38.
  27. Kolber MJ, Saltzman SB, Beekhuizen KS, et al. Reliability and minimal detectable change of inclinometric shoulder mobility measurements. Physiother Theory Pract. 2009;25(8):572-581. https://doi.org/10.3109/09593980802667995
  28. Kolber MJ, Vega F, Widmayer K, et al. The reliability and minimal detectable change of shoulder mobility measurements using a digital inclinometer. Physiother Theory Pract. 2011; 27(2):176-184. https://doi.org/10.3109/09593985.2010.481011
  29. Korea Communications Commission. Wired and wireless subscribers statistics (2012, Dec) [Internet]. Seoul, 2013 Feb 7 [Cited 2013 May 21]. Available from: http://old.kcc.go.kr/user.do?mode=view&page=P02060400&dc=K02060400&boardId=1030&cp=1&boardSeq=35848.
  30. Lewis JS, Valentine RE. Intraobserver reliability of angular and linear measurements of scapular position in subjects with and without symptoms. Arch Phys Med Rehabil. 2008;89(9):1795-1802. https://doi.org/10.1016/j.apmr.2008.01.028
  31. Logan LR. Rehabilitation techniques to maximize spasticity management. Top Stroke Rehabil. 2011;18(3):203-211. https://doi.org/10.1310/tsr1803-203
  32. McConnell J, Donnelly C, Hamner S, et al. Passive and dynamic shoulder rotation range in uninjured and previously injured overhead throwing athletes and the effect of shoulder taping. PM R. 2012;4(2):111-116. https://doi.org/10.1016/j.pmrj.2011.11.010
  33. Mottram SL. Dynamic stability of the scapula. Man Ther. 1997;2(3):123-131. https://doi.org/10.1054/math.1997.0292
  34. Muehlbauer T, Roth R, Mueller S, et al. Intra and intersession reliability of balance measures during one-leg standing in young adults. J Strength Cond Res. 2011;25(8):2228-2234. https://doi.org/10.1519/JSC.0b013e3181fb393b
  35. Muir SW, Corea CL, Beaupre L. Evaluating change in clinical status: Reliability and measures of agreement for the assessment of glenohumeral range of motion. N Am J Sports Phys Ther. 2010;5(3):98-110.
  36. Mullaney MJ, McHugh MP, Johnson CP, et al. Reliability of shoulder range of motion comparing a goniometer to a digital level. Physiother Theory Pract. 2010;26(5):327-333. https://doi.org/10.3109/09593980903094230
  37. Norkin CC, White DJ. Measurement of Joint Motion: A guide to goniometry. 4th ed. Philadelphia, PA, F.A. Davis Co., 2009:21.
  38. Ockendon M, Gilbert RE. Validation of a novel smartphone accelerometer-based knee goniometer. J Knee Surg. 2012;25(4):341-345. https://doi.org/10.1055/s-0031-1299669
  39. Ottenbacher KJ, Jannell S. The results of clinical trials in stroke rehabilitation research. Arch Neurol. 1993;50(1):37-44. https://doi.org/10.1001/archneur.1993.00540010033014
  40. Park JH, Kwon YC. Standardization of Korean version of the mini-mental state examination (MMSE-K) for use in the elderly. Part II. Diagnostic validity. J Korean Neuropsychiatr Assoc. 1989;28(3):508-513.
  41. Park KN, Ha SM, Chung SD, et al. Intra-rater test-retest reliability of Ely's test using smartphone in patients with lumbar extension-rotation syndrome. Phys Ther Kor. 2012;19(4):46-54. https://doi.org/10.12674/ptk.2012.19.4.046
  42. Portney LG, Watkins MP. Foundations of Clinical Research: Applications to practice. 3rd ed. Upper Saddle River, NJ, Prentice Hall Inc., 2009:594-595.
  43. Radomski MV, Trombly CA. Occupational Therapy for Physical Dysfunction. 6th ed. Baltimore, MD, Williams & Wilkins, 2008:94.
  44. Rajaratnam BS, Venketasubramanian N, Kumar PV, et al. Predictability of simple clinical tests to identify shoulder pain after stroke. Arch Phys Med Rehabil. 2007;88(8):1016-1021. https://doi.org/10.1016/j.apmr.2007.05.001
  45. Riddle DL, Rothstein JM, Lamb RL. Goniometric reliability in a clinical setting: Shoulder measurements. Phys Ther. 1987;67(5):668-673. https://doi.org/10.1093/ptj/67.5.668
  46. Sabari JS, Maltzev I, Lubarsky D, et al. Goniometric assessment of shoulder range of motion: Comparison of testing in supine and sitting positions. Arch Phys Med Rehabil. 1998;79(6): 647-651. https://doi.org/10.1016/S0003-9993(98)90038-7
  47. Sackley C, Brittle N, Patel S, et al. The prevalence of joint contractures, pressure sores, painful shoulder, other pain, falls, and depression in the year after a severely disabling stroke. Stroke. 2008;39(12):3329-3334. https://doi.org/10.1161/STROKEAHA.108.518563
  48. Shrout PE, Fleiss JL. Intraclass correlations: Uses in assessing rater reliability. Psychol Bull. 1979;86 (2):420-428. https://doi.org/10.1037/0033-2909.86.2.420
  49. Shin SH, Ro du H, Lee OS, et al. Within-day reliability of shoulder range of motion measurement with a smartphone. Man Ther. 2012;17(4):298-304. https://doi.org/10.1016/j.math.2012.02.010
  50. Statistics Korea. Causes of Death Statistics in 2010 [Internet]. Daejeon, 2011 Sep 8 [Cited 2013 Aug 2]. Available from: http://kostat.go.kr/portal/korea/kor_nw/2/6/2/index.board?bmode=read&aSeq=250282.
  51. Tveita EK, Ekeberg OM, Juel NG, et al. Range of shoulder motion in patients with adhesive capsulitis; Intra-tester reproducibility is acceptable for group comparisons. BMC Musculoskelet Disord. 2008;9:49. https://doi.org/10.1186/1471-2474-9-49
  52. Valentine RE, Lewis JS. Intraobserver reliability of 4 physiologic movements of the shoulder in subjects with and without symptoms. Arch Phys Med Rehabil. 2006;87(9):1242-1249. https://doi.org/10.1016/j.apmr.2006.05.008

Cited by

  1. Effect of backward walking training using an underwater treadmill on muscle strength, proprioception and gait ability in persons with stroke vol.6, pp.3, 2017, https://doi.org/10.14474/ptrs.2017.6.3.120
  2. 지속적인 전방머리자세 시 머리-목 관절 위치 감각의 변화 vol.5, pp.2, 2017, https://doi.org/10.15268/ksim.2017.5.2.011
  3. 어깨뼈 다이나믹 테이핑이 어깨 수술 환자의 통증과 기능장애 수준, 상지 자세와 관절가동범위에 미치는 영향 vol.13, pp.4, 2018, https://doi.org/10.13066/kspm.2018.13.4.149
  4. 진동자극 환경에서의 요부 안정화 운동이 만성요통환자의 균형능력 및 기능장애수준에 미치는 영향 vol.26, pp.1, 2019, https://doi.org/10.12674/ptk.2019.26.1.019
  5. Study on the Reliability of Smartphone-Based Measurements of Leg Segment and Joint Angles During Gait vol.36, pp.3, 2019, https://doi.org/10.7736/kspe.2019.36.3.293
  6. 진동을 동반한 슬링 운동이 어깨 손상 환자의 관절가동범위, 근력, 통증, 기능장애 수준에 미치는 영향 vol.26, pp.3, 2014, https://doi.org/10.12674/ptk.2019.26.3.011
  7. Validity and reliability of a smartphone inclinometer app for measuring passive upper limb range of motion in a stroke population* vol.42, pp.22, 2014, https://doi.org/10.1080/09638288.2019.1585972