DOI QR코드

DOI QR Code

Assessment for the Temperature according to the Electrode Diameter of Radio Frequency Hyperthermia Using Agar Phantom

고주파 온열치료기의 전극의 직경에 따른 한천 팬톰의 온도분포 분석

  • Lee, Yong Hee (Department of Radiation Oncology, Dongsan Medical Center, Keimyung University School of Medicine) ;
  • Oh, Young Kee (Department of Radiation Oncology, Dongsan Medical Center, Keimyung University School of Medicine) ;
  • Kim, Hwa Yeong (Department of Radiation Oncology, Dongsan Medical Center, Keimyung University School of Medicine) ;
  • Jeon, Kyung Soo (Department of Radiation Oncology, Dongsan Medical Center, Keimyung University School of Medicine) ;
  • Choi, Eun Cheol (Department of Radiation Oncology, Dongsan Medical Center, Keimyung University School of Medicine) ;
  • Park, Seung Gyu (Department of Radiation Oncology, Dongsan Medical Center, Keimyung University School of Medicine) ;
  • Kim, Ok Bae (Department of Radiation Oncology, Dongsan Medical Center, Keimyung University School of Medicine) ;
  • Kim, Jin Hee (Department of Radiation Oncology, Dongsan Medical Center, Keimyung University School of Medicine)
  • 이용희 (계명대학교 동산의료원 방사선종양학과) ;
  • 오영기 (계명대학교 동산의료원 방사선종양학과) ;
  • 김화영 (계명대학교 동산의료원 방사선종양학과) ;
  • 전경수 (계명대학교 동산의료원 방사선종양학과) ;
  • 최은철 (계명대학교 동산의료원 방사선종양학과) ;
  • 박승규 (계명대학교 동산의료원 방사선종양학과) ;
  • 김옥배 (계명대학교 동산의료원 방사선종양학과) ;
  • 김진희 (계명대학교 동산의료원 방사선종양학과)
  • Received : 2013.12.06
  • Accepted : 2014.03.10
  • Published : 2014.03.31

Abstract

Hyperthermia is effective treatment modality when it combine with the radiotherapy treatment. It is important to verify the temperature distribution of (patient's) body for the safety and effective treatment during raising the temperature. In this study, we raised the temperature in agar phantom using radio frequency (RF) Hyperthermia and protocol that manufacture recommend. Temperature distribution measured 5 section of 5 cm thickness with agar phantom. When the temperature was raised according to the increase energy. Temperature distribution was elevated at similar domain regardless of energy. The temperature tend to be increased at up side then bottom side and also increase when A large electrode was used than small one.

온열 치료는 방사선치료와 병행했을 때 종양에 대한 효과적인 치료방법이다. 환자의 안전과 효과적인 치료를 위해서는 가온 시 인체의 온도의 분포를 확인하는 것이 중요하다. 본 연구는 사용중인 고주파온열치료기와 제조사가 권하고 있는 protocol를 이용하여 펜톰의 온도를 높여주었다. 온도측정에 사용된 펜톰은 한천(agar)으로 5 cm두께로 단면을 측정할 수 있게 제작되었다. 가온 시 온도의 분포는 에너지가 클수록 온도가 높게 올라갔으며, 온도 분포는 에너지에 상관없이 비슷한 영역이 가온되었다. 아래쪽보다 위쪽으로 온도가 높게 올라가는 경향을 보이고 있으며, 큰 전극보다는 작은 전극을 사용했을 때 온도가 높게 올라가고 치우치는 온도분포를 보인다.

Keywords

References

  1. de Bruijne M, Van der Zee J, Ameziane A, Van Rhoon GC: Schlag PM: Quality control of superficial hyperthermia by treatment evaluation: Int J Hyperthermia. 27(3):199-213(2011) https://doi.org/10.3109/02656736.2010.525226
  2. Overgaard J, Gonzalez Gonzalez D, Hulshof MCCM, et al: Trial of hyperthermia as adjuvant to radiotherapy for recurrentor metastatic malignant melanoma. Lancet 345:540-543(1995) https://doi.org/10.1016/S0140-6736(95)90463-8
  3. Vernon CC, Hand JW, Field SB, et al: Radiotherapy with or without hyperthermia in the treatment of superficial localized breast cancer - results from five randomised controlled trials. Int J Radiation Oncology Biol Phys 35:731-744(1996) https://doi.org/10.1016/0360-3016(96)00154-X
  4. Jones EL, Oleson JR, Prosnitz LR, et al: Randomized trial of hyperthermia and radiation for superficial tumors. J Clin Oncol 23:3079-3085(2005) https://doi.org/10.1200/JCO.2005.05.520
  5. 추성실, 김귀언, 노준규, et al: 암 치료를 위한 RF 온열장치의 개발과 가온 특성. 대한암학회지 18(2):183-193(1986)
  6. P Wust, B Hildebrandt, G Sreenivasa, et al: Hyperthermia in combined treatment of cancer. Lancet Oncol. 3(8):487- 97(2002) https://doi.org/10.1016/S1470-2045(02)00818-5
  7. Rietveld PJM, Van Putten WLJ, Van der Zee J, Van Rhoon GC: Comparison of the clinical effectiveness of the 433 Mhz lucite cone applicator with that of a conventional waveguide applicator in applications of superficial hyperthermia. Int J Radiat Oncol Biol Phys 43:681-687(1999) https://doi.org/10.1016/S0360-3016(98)00443-X
  8. Lee WM, Gelvich EA, Van der Baan P, Mazokhin VN, Van Rhoon GC: Assessment of the performance characteristics of a prototype 12-element capacitive contact flexible microstrip applicator (CFMA-12) for superficial hyperthermia. Int J Hyperthermia 20:607-624(2004) https://doi.org/10.1080/02656730410001695915
  9. Diederich CJ, Stauffer PR: Pre-clinical evaluation of a microwave planar array applicator for superficial hyperthermia. Int J Hyperthermia 9:227-246(1993) https://doi.org/10.3109/02656739309022537
  10. Lee ER, Wilsey TR, Tarczy-Hornoch P, et al: Body conformable915MHz microstrip array applicators for large surface areahyperthermia. IEEE Trans Biomed Eng 39:470-483(1992) https://doi.org/10.1109/10.135541
  11. Samulski TV, Fessenden P, Lee ER, Kapp DS, Tanabe E, McEuen A: Spiral microstrip hyperthermia applicators: Technical design and clinical performance. Int J Radiat Oncol Biol Phys 18:233-42(1990) https://doi.org/10.1016/0360-3016(90)90288-U
  12. Rossetto F, Stauffer PR: Theoretical characterization of dual concentric conductor microwave applicators for hyperthermia at 433 MHz. Int J Hyperthermia 17:258-270(2001) https://doi.org/10.1080/02656730010029342
  13. Jacobsen S, Stauffer PR, Neuman DG: Dual-mode antenna design for microwave heating and noninvasive thermometry of superficial tissue disease. IEEE Trans Biomed Eng 47:1500-1509(2000) https://doi.org/10.1109/10.880102
  14. Juang T, Stauffer PR, Neuman DG, Schlorff JL: Multilayer conformal applicator for microwave heating and brachytherapy treatment of superficial tissue disease. Int J Hyperthermia 22:527-544(2006) https://doi.org/10.1080/02656730600931815
  15. Andreas Jordan, Regina Scholz, Klaus Maier-Hau!, et al: Presentation of a new magnetic field therapy system for the treatment of human solid tumors with magnetic fluid hyperthermia. Journal of Magnetism and Magnetic Materials 225: 118-126(2001) https://doi.org/10.1016/S0304-8853(00)01239-7