Evaluation of the Reproducibility of Radiation Output from Diagnostic X-ray Equipment(Standards Based on IEC 60601-2-54)

진단용 X선 장치에서 방사선출력의 재현성 평가(IEC 60601-2-54 표준규격을 기반으로)

  • 한범희 (서남대학교 방사선학과) ;
  • 정홍량 (한서대학교 보건의료학과) ;
  • 임청환 (한서대학교 보건의료학과) ;
  • 김종일 (전북대학교 방사선과학기술학과) ;
  • 이상호 (서남대학교 방사선학과) ;
  • 한상현 (서남대학교 방사선학과) ;
  • 홍동희 (서남대학교 방사선학과) ;
  • 김창규 (김천대학교 방사선학과) ;
  • 유인규 (한서대학교 보건의료학과) ;
  • 모은희 (전북대학교 방사선과학기술학과)
  • Received : 2014.01.18
  • Accepted : 2014.02.20
  • Published : 2014.02.28

Abstract

For five diagnostic X-ray generators (DR), four units turned out to be appropriate in tests on the reproducibility of radiation output suggested in the IEC 60601-2-54 standard, but in one unit of the X-ray equipment, an item measured in a combination of 50% of the highest tube voltage of the diagnostic X-ray equipment, the test setting of Group C with authorized output doses between $1{\mu}Gy$ and $5{\mu}Gy$ of mAs turned out to be inappropriate. As a result, the radiation dose to the IEC 60601-2-54 standard for quantification standards proposed by the radiation output from diagnostic X-ray imaging device reproducibility of performance management should be aware that an important evaluation factor.

총 5대를 대상으로 한 진단용X선발생장치(DR)는 IEC 60601-2-54 표준규격에서 제시한 방사선출력의 재현성 시험에서 4대가 적합으로 나타났으며, 다만 1대의 X선 장치에서 C Group의 시험 설정인 진단용X선장치의 최고 관전압의 50%와 관전류 시간 곱(mAs)이 $1{\mu}Gy$에서 $5{\mu}Gy$ 사이의 인가출력량의 조합으로 측정한 항목에서 부적합으로 나타났다. 이에 따라 방사선량을 정량화하기 위한 표준으로 IEC 60601-2-54 표준규격에서 제시한 방사선출력의 재현성이 진단용 X선 영상장치 성능관리에 중요한 평가 요소임을 인식해야 한다.

Keywords

References

  1. IEC 61223-1 Evaluation and routine testing in medical imaging departments-Part 1 : General aspects, 2005. 09. 30
  2. Kim Jung-Min, "Diagnostic X-ray equipment to keep the performance management plan", Korea Food & Drug Administration, Radiation Health Newsletter, Vol. 19, No. 1, pp. 15, 2012.
  3. IEC 61223-2-11 Evaluation and routine testing in medical imaging departments-Part 2-11:Constancy tests-Equipment for general direct radiography, 09. 30, 2005.
  4. Yoon Jung-Seop , Kim Choon-Sik, Ko Shin-Gwan, "Medical X-ray equipment on the management of the maintenance and performance review", Journal of Radiological Science and Technology, Vol. 15, No. 1, pp. 25-26, 1992.
  5. IEC 60601-2-54 Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy, 01. 02, 2012.
  6. Kim Hyun-Ji, Kim Jung-Min, Choi In-Seok, "Features and Trends of IEC Particular Standards for Medical Equipment Related to Diagnostic X-ray Based on IEC 60601-1:2005 Ed. 3.0", Journal of Radiological Science and Technology, Vol.36, No. 1, pp. 1-10, 2013.
  7. Whang Choon-Sung, "A Study on the Output Characteristics of X-ray Generator", Dept. of Electrical Engineering, Graduate School of Industry, Chosun University, pp. 33-35, 1999
  8. Na Kil-Ju, "A Study on the Reproducibility of Exposure Doses and Half Value Layer by Shield for Medical X-ray Generator", Department of Electrical Engineering Graduate School of Mokpo National University, pp. 48-50, 2010.
  9. Kwon Duk-Moon, " Investigation of Diagnostic X-ray Equipments Characteristics and Improvement of Collimator Using Laser", Department of Sensor and System Engineering Graduate School Yeungnam University, pp. 57-59, 2002.
  10. Kang Byung-Sam, Lee Kang-min, Shim Woo-Yong, "Analyze for the Quality Control of General X-ray Systems in Capital region", Journal of Radiological Science and Technology, Vol. 35, No. 2, pp. 93-101, 2012.
  11. Dong Kyung-Rae, Lee Seun-Joo, Kweon Dae-Cheol, Goo Eun-Hoe, "Actual Condition of Quality Control of X-ray Imaging System in Primary Care Institution: focused on Gwangju Metropolitan City", The Journal of the Korean Association for Radiation Protection, Vol. 35, No. 1, pp. 34-42, 2010.