Abstract

Noncontact electrocardiogram (ECG) measurement using capacitive-coupled technique is a very reliable long-term noninvasive health-care remote monitoring system. It can be used continuously without interrupting the daily activities of the user and is one of the most promising developments in health-care technology. However, ECG signal is a very small electric signal. A robust system is needed to separate the clean ECG signal from noise in the measurement environment. Noise may come from many sources around the system, for example, bad contact between the sensor and body, common-mode electrical noise, movement artifacts, and triboelectric effect. Thus, in this paper, the extended Kalman filter (EKF) is applied to denoise a real-time ECG signal in capacitive-coupled sensors. The ECG signal becomes highly stable and noise-free by combining the common analog signal processing and the digital EKF in the processing step. Furthermore, to achieve ubiquitous monitoring, android-based application is developed to process the heart rate in a realtime ECG measurement.

Keywords

References

  1. A. J. Lopez and P. C. Richardson, "Capacitive electrocardiographic and bioelectric electrodes", IEEE Trans. Biomed. Eng., vol. 16, pp. 99, Jan. 1969.
  2. Y. G. Lim, K. K. Kim, and K. S. Park, "ECG measurement on a chair without conductive contact", IEEE Trans. Biomed. Eng., vol. 53, pp. 956-959, May 2006. https://doi.org/10.1109/TBME.2006.872823
  3. A. Ueno, Y. Akabane, T. Kato, H. Hoshino, S. Kataoka, Y. Ishiyama "Capacitive sensing of electrocardiographic potential through cloth from the dorsal surface of the body in a supine position: A preliminary study", IEEE Trans. Biomed. Eng., vol. 54, no. 4, pp. 759-766, 2007. https://doi.org/10.1109/TBME.2006.889201
  4. S. Leonhardt and A. Aleksandrowicz, "Non-contact ECG monitoring for automotive application", in Proc 5th Int. BSN Conf., pp.183-185, Hong Kong, China, 2008.
  5. S. M. Lee, K. S. Sim, K. K. Kim, Y. G. Lim, and K. S. Park, "Thin and flexible active electrodes with shield for capacitive electrocardiogram measurement", Med Biol. Eng. Comput., vol. 48, pp. 447-457, 2010. https://doi.org/10.1007/s11517-010-0597-y
  6. J. H. Kim, S. C. Lee, B. G. Lee, and W. Y. Chung, "Mobile healthcare system based on bluetooth medical device", J. Sensor Sci. & Tech., vol. 21, no. 4, pp. 241-248, 2012. https://doi.org/10.5369/JSST.2012.21.4.241
  7. M. A. D. Raya and L. G. Sison, "Adaptive noise cancellation of motion artifact in stress ECG signals using accelerometer", Proc. 2nd Joint EMBS/BMES Conference. 2002.
  8. A. Gautam, Y. D. Lee, and W. Y. Chung, "ECG signal denoising with signal averaging and filtering algorithm", Third International Conference on Convergence and Hybrid Information Technology, 2008.
  9. S. M. Lee, K. K. Kim, and K. S. Park, "Wavelet approach to artifact noise removal from capacitive coupled electrocardiograph", 30th Annual International IEEE EMBS Con- ference, Canada, pp. 2944-4947, 2008.
  10. P. E. McSharry, G. D. Clifford, L. Tarassenko, and L. A. Smith, "A dynamical model for generating synthetic electrocardiogram signals", IEEE Transaction on Biomedical Engineering, vol. 50, no. 3, pp. 289-294, 2003. https://doi.org/10.1109/TBME.2003.808805
  11. R. Sameni, M. B. Shamsollahi, C. Jutten, and M. B. Zadeh, "Filtering noisy ECG signals using the extended Kalman filter based on a modified dynamic ECG model", IEEE Computes in Cardiology, vol. 32, pp. 1017-1020, 2005.
  12. M. Niknazar, B. Rivet, and C. Jutten, "Fetal ECG extraction by extended state Kalman filtering based on single-channel recordings", IEEE Transactions on Biomedical Engineering, vol. 60, no. 5, pp 1345-1352, 2013. https://doi.org/10.1109/TBME.2012.2234456
  13. J. Pan and W. J. Tompkins "A real-time QRS detection algorithm", IEEE Trans Biomed. Eng, vol. 32, no. 3, pp. 230-236, 1985.

Cited by

  1. Wearable Noncontact Armband for Mobile ECG Monitoring System vol.10, pp.6, 2016, https://doi.org/10.1109/TBCAS.2016.2519523
  2. A Robust Algorithm for Real-Time Peak Detection of Photoplethysmograms Using a Personal Computer Mouse vol.15, pp.8, 2015, https://doi.org/10.1109/JSEN.2015.2424979
  3. Reduction of Motion Artifacts and Improvement of R Peak Detecting Accuracy Using Adjacent Non-Intrusive ECG Sensors vol.16, pp.5, 2016, https://doi.org/10.3390/s16050715
  4. Denoising ECG Signals by Using Extended Kalman Filter to Train Multi-Layer Perceptron Neural Network vol.52, pp.6, 2018, https://doi.org/10.3103/S0146411618060044