초록

A 3D ultrasound image is desired in many medical examinations. However, the implementation of a 2D array, which is needed for a 3D image, is challenging with respect to fabrication, interconnection and cabling. A 2D sparse array, which needs fewer elements than a dense array, is a realistic way to achieve 3D images. Because the number of ways the elements can be placed in an array is extremely large, a method for optimizing the array configuration is needed. Previous research placed the target point far from the transducer array, making it impossible to optimize the array in the operating range. In our study, we focused on optimizing a 2D sparse array transducer for 3D imaging by using a simulated annealing method. We compared the far-field optimization method with the near-field optimization method by analyzing a point-spread function (PSF). The resolution of the optimized sparse array is comparable to that of the dense array.

키워드

참고문헌

  1. P. K. Weber and L. Peter "1D and 2D Sparse Array Optimization," Instrumentation Science & Technology, Vol. 27, pp. 235-246 (1999) https://doi.org/10.1080/10739149908085857
  2. A. Austeng and S. Holm "Spare 2D arrays for real-time 3D ultrasound," IEEE Trasactions On Ultrasonics, Ferroelectrics, and Frequency Control, Vol. 49 pp. 1073-1086 (2002) https://doi.org/10.1109/TUFFC.2002.1026019
  3. J. W. Choe, O. Oralkan and P. T. Khuri-Yakub, "Design optimization for a 2-D sparse transducer array for 3-D ultrasound imaging," Proceeding of the IEEE Ultrason. Symp, pp. 1928-1931 (2010)
  4. B. Diarra, H. Liebgott, P. Tortoli and C. Cachard, "Sparse array techniques for 2D array ultrasound imaging," Proceeding of the Acoustics 2012 Nantes Conference, pp. 1591-1596 (2012)
  5. M. Karaman, I. O. Wygant, O. Oralkan and B. T. Khuri-Yakub, "Minimally redundant 2-D array designs for 3-D medical ultrasound imaging," IEEE Transactions on Medical Imaging, Vol. 28, pp. 1051-1061 (2009) https://doi.org/10.1109/TMI.2008.2010936
  6. A. Austeng and S. Holm, "Sparse 2-D arrays for 3-D phased array imaging-design method," IEEE Trasactions on Ultrasonics, Ferroelectrics, and Frequency Control, Vol. 49, pp. 2073-2085 (2002)
  7. Y. Jia, M. Xu, M. Ding and M. Yuchi, "3D ultrasound coherence imaging based on 2D array design," Proceeding of the Medical Imaging 2013: Ultrasonic Imaging, Tomography, and Therapy (2013)