가상 Frisch-그리드를 이용한 CdZnTe 감마선 소자 제작

Fabrication of Virtual Frisch-Grid CdZnTe ${\gamma}$-Ray Detector

  • 박찬선 (고려대학교 방사선학과) ;
  • 김필수 (고려대학교 방사선학과) ;
  • 조평곤 (대구가톨릭대학교 방사선학과) ;
  • 김정민 (고려대학교 방사선학과) ;
  • 최종학 (고려대학교 방사선학과) ;
  • 김기현 (고려대학교 방사선학과)
  • Park, Chansun (Department of Radiologic Science, Korea University) ;
  • Kim, Pilsu (Department of Radiologic Science, Korea University) ;
  • Cho, PyongKon (Department of Radiological Science, Catholic University of Daegu) ;
  • Choi, Jonghak (Department of Radiologic Science, Korea University) ;
  • Kim, Jungmin (Department of Radiologic Science, Korea University) ;
  • Kim, KiHyun (Department of Radiologic Science, Korea University)
  • 투고 : 2014.10.14
  • 심사 : 2014.12.08
  • 발행 : 2014.12.31

초록

Traveling heater method(THM) 방법을 이용하여 성장시킨 CdZnTe(CZT) 단결정 방사선 소자에 대한 고에너지(high energy) 감마선 에너지 분해능(energy resolution)을 평가하고자 $6{\times}6{\times}12mm^3$ 크기의 CZT 소자를 제작하였다. 두꺼운 방사선 소자의 경우, 전자에 비해 상대적으로 이동속도가 느린 정공(hole)으로 인해 발생하는 hole-tailing 효과가 심화되어 고에너지 영역의 에너지 분해능이 저하되는 현상이 발생한다. 전자(electron)와 정공(hole)의 두 개의 전하 운반자(charge carrier) 중에서 하나의 전하 운반자를 선택적으로 수집하여 에너지 분해능을 높이는 것이 가능하다. 가상 Frisch-그리드(virtual Frisch-grid) 소자는 소자 내부의 가중 퍼텐셜(weighting potential)을 조절하여 전자에 의한 유도전류(induced current)만을 선택적으로 이용하는 방법으로써 제작 과정과 적용이 용이하다. 본 연구에서는 THM 방법으로 성장한 큰 부피의 CZT 방사선 소자의 특성과 가상 Frisch-그리드의 효용성을 평가하였다. 가상 Frisch-그리드의 적절한 위치와 너비는 Maxwell ver.14(ANSYS, 미국)를 이용하여 모의실험으로 정하였다. $^{137}Cs$ 동위원소를 이용한 펄스 높이 스펙트럼(pulse height spectrum) 측정에서 가상 Frisch-그리드를 적용했을 때 662 keV 피크에 대해 2.2%의 에너지 분해능을 확인할 수 있었다.

Large volume of $6{\times}6{\times}12mm^3$ CdZnTe ${\gamma}$-ray detector was fabricated with CdZnTe single crystals grown by Traveling Heater Method (THM) to evaluate the energy resolution of 662 keV in $^{137}Cs$. Hole tailing effect which originated from the large mobility difference in electron and hole degrade energy resolution of radiation detector and its effects become more severe for a large volume detectors. Generally, single carrier collection technique is very useful method to remove/minimize hole tailing effect and thereby improvement in energy resolution. Virtual Frisch-grid technique is also one of single charge collection method through weighting potential engineering and it is very simple and easily applicable one. In this paper, we characterized CZT detector grown by THM and evaluated the effectiveness of virtual Frisch-grid technique for a high energy gamma-ray detector. The proper position and width of virtual Frisch-grid was determined from electric field simulation using ANSYS Maxwell ver. 14.0. Energy resolution of 2.2% was achieved for the 662 keV ${\gamma}$-peak of $^{137}Cs$ with virtual Frisch-grid CdZnTe detector.

키워드

참고문헌

  1. D. S. McGregor, Z. He, H. A. Seifert, D. K. Wehe, and R. A. Rojeski: Single Charge Carrier Type Sensing with a Parallel Strip Pseudo-Frisch-Grid CdZnTe Semiconductor Radiation Detector, Appl. Phys. Letts., 72(7), 792-795, 1998 https://doi.org/10.1063/1.120895
  2. A. Kargar, A. M. Jones, W. J. McNeil, M. J. Harrison, and D. S. McGregor: CdZnTe Frisch collar detectors for ${\gamma}$-ray spectroscopy, Nucl. Instr. and Meth. Phys. Res. A, 558, 497-503, 2006 https://doi.org/10.1016/j.nima.2005.11.211
  3. M. van Pamelen and C. Budtz-Jorgensen: CdZnTe Drift Detector with Correction for Hole Trapping, Nucl. Instr. and Meth. A, 411, 197-200, 1998 https://doi.org/10.1016/S0168-9002(98)00283-6
  4. A. Owens and A. G. Kozorezov: Single Carrier Sensing in Compound Semiconductor Detectors, Nucl. Instr. and Meth. A, 563, 31-36, 2006 https://doi.org/10.1016/j.nima.2006.01.101
  5. Z. He: Review of the Shockley-Ramo theorem and its application in semiconductor gamma-ray detectors, Nucl. Instr. Meth. A, 463, 250-267, 2001 https://doi.org/10.1016/S0168-9002(01)00223-6
  6. D. S. McGregor, R. A. Rojeski, Z. He, D. K. Wehe, M. Driver, and M. Blakely: Geometrically Weighted Semiconductor Frisch Grid Radiation Spectrometers, Nucl. Instr. and Meth. A, 422, 164-168, 1999. https://doi.org/10.1016/S0168-9002(98)00948-6
  7. A. E. Bolotnikov, N. M. Abdul-Jabbarb, S. Babalola et al.: Optimization of virtual Frischgrid CdZnTe detector designs for imaging and spectroscopy of gamma rays, SPIE, 6706-02, 2007
  8. K. H. Kim, A. E. Bolotnikov, G. S. Camarda et al.: New Approaches for Making Large-Volume and Uniform CdZnTe and CdMnTe Detectors: IEEE Trans. Nucl. Sci. 59(4), 1510, 2012 https://doi.org/10.1109/TNS.2012.2202917