A Study on the Use of Active Protocol Using the Change of Pitch and Rotation Time in PET/CT
PET/CT에서 Pitch와 Rotation Time의 변화를 이용한 능동적인 프로토콜 사용에 대한 연구
- Jang, Eui Sun (Department of Nuclear Medicine, Samsung Medical Center) ;
- Kwak, In Suk (Department of Nuclear Medicine, Samsung Medical Center) ;
- Park, Sun Myung (Department of Nuclear Medicine, Samsung Medical Center) ;
- Choi, Choon Ki (Department of Nuclear Medicine, Samsung Medical Center) ;
- Lee, Hyuk (Department of Nuclear Medicine, Samsung Medical Center) ;
- Kim, Soo Young (Department of Nuclear Medicine, Samsung Medical Center) ;
- Choi, Sung Wook (Department of Nuclear Medicine, Samsung Medical Center)
- 장의순 (삼성서울병원 핵의학과) ;
- 곽인석 (삼성서울병원 핵의학과) ;
- 박선명 (삼성서울병원 핵의학과) ;
- 최춘기 (삼성서울병원 핵의학과) ;
- 이혁 (삼성서울병원 핵의학과) ;
- 김수영 (삼성서울병원 핵의학과) ;
- 최성욱 (삼성서울병원 핵의학과)
- Received : 2013.10.08
- Accepted : 2013.11.02
- Published : 2013.11.23
Abstract
Purpose: The Change of CT exposure condition have a effect on image quality and patient exposure dose. In this study, we evaluated effect CT image quality and SUV when CT parameters (Pitch, Rotation time) were changed. Materials and Methods: Discovery Ste (GE, USA) was used as a PET/CT scanner. Using GE QA Phantom and AAPM CT Performance Phantom for evaluate Noise of CT image. Images are acquired by using 24 combinations that four stages pitch (0.562, 0.938, 1.375, 1.75:1) and six stages X-ray tube rotation time (0.5s-1.0s). PET images are acquired using 1994 NEMA PET Phantom (
PET/CT검사에서 CT촬영조건의 변화는 영상의 화질 및 환자가 받는 피폭선량에 영향을 미친다. 본 연구는 CT 매개 변수 중 Pitch와 X-선관 회전시간 변화에 따른 선량대비 CT 영상의 질과 이로 인해 PET상에서 SUV에 미치는 영향을 비교 평가하고자 하였다. Discovery STe PET/CT 장비를 이용하여 영상을 획득하였다. QA Phantom과 AAPM Phantom을 이용한 CT 영상 획득 시 Pitch는 0.562, 0.938, 1.375, 1.75:1로 4단계, X-선관 회전시간은 0.5에서 1.0까지 0.1초씩 증가시켜 6단계로 나누어 총 24개 조합을 적용한 영상을 각각 획득하였다. PET 영상은