DOI QR코드

DOI QR Code

INTERMITTENT, NON CYCLIC SEVERE MECHANICAL AORTIC VALVE REGURGITATION

  • Choi, Jong Hyun (Division of Cardiology, Pusan National University Hospital) ;
  • Choi, Jung Hyun (Division of Cardiology, Pusan National University Hospital) ;
  • Song, Seunghwan (Division of Cardiology, Pusan National University Hospital) ;
  • Lee, Myung-Yong (Division of Cardiology, Dankook University College of Medicine)
  • Received : 2013.06.23
  • Accepted : 2013.11.12
  • Published : 2013.12.27

Abstract

Mechanical aortic prosthesis dysfunction can result from thrombosis or pannus formation. We describe an unusual case of intermittent, non cyclic mechanical aortic prosthesis dysfunction due to pannus formation with thrombus in the absence of systolic restriction of disk excursion, that presented with intermittent severe aortic regurgitation.

Keywords

Cited by

  1. Clinical Outcomes of Repeat Aortic Valve Replacement for Subaortic Pannus in Mechanical Aortic Valve vol.82, pp.10, 2013, https://doi.org/10.1253/circj.cj-18-0352
  2. Effects of pannus formation on the flow around a bileaflet mechanical heart valve vol.15, pp.6, 2020, https://doi.org/10.1371/journal.pone.0234341