DOI QR코드

DOI QR Code

Comparison of the Accuracy of Noninvasive Hemoglobin Sensor (NBM-200) and Portable Hemoglobinometer (HemoCue) with an Automated Hematology Analyzer (LH500) in Blood Donor Screening

  • Kim, Moon Jung (Department of Biomedical Laboratory Science, Eulji University College of Health Science) ;
  • Park, Quehn (Research Institute of Medicine and Pharmacy, Chung-Ang University) ;
  • Kim, Myung Hee (Quality Assurance Division, Hanmaum Blood Center) ;
  • Shin, Jeong Won (Department of Laboratory Medicine, Soonchunhyang University College of Medicine) ;
  • Kim, Hyun Ok (Department of Laboratory Medicine, Yonsei University College of Medicine)
  • Published : 2013.07.01

Abstract

Background: The Hb levels of prospective blood donors are usually determined using a finger prick test. A new noninvasive Hb device has the advantage of not causing any sampling pain. The purpose of this study was to evaluate the accuracy of the noninvasive Hb sensor and to compare its measurements with those of a currently used portable hemoglobinometer. Methods: Hb was measured using a noninvasive Hb sensor (NBM-200; OrSense, Israel), a portable hemoglobinometer (HemoCue; HemoCue AB, Sweden), and an automated hematology analyzer (LH500; Beckman Coulter, USA). The correlations between Hb measurements taken by the NBM-200 and HemoCue with those by an automated hematology analyzer were assessed using intraclass correlation coefficients (ICCs). Hb measurements were compared among 3 different Hb level groups. Results: The mean Hb values of 506 blood donors were 14.1 g/dL by the NBM-200, 14.0 g/dL by the LH500, and 14.3 g/dL by the HemoCue. The correlation between the LH500 and the NBM-200 was substantial (ICC=0.69), while that between the LH500 and the HemoCue agreed almost perfectly (ICC=0.86). Conclusions: The possibility to judge to be eligible for donors who are ineligible to donate was substantial when using NBM-200. Even though the NBM-200 has the apparent advantage of noninvasiveness, its use in pre-screening should be given meticulous attention. Since pre-donation testing is crucial to protecting donors' health, complete evaluation of the instrument should be performed prior to use.

Keywords

References

  1. Cable RG. Hemoglobin determination in blood donors. Transfus Med Rev 1995;9:131-44. https://doi.org/10.1016/S0887-7963(05)80052-5
  2. Boulton FE, Nightingale MJ, Reynolds W. Improved strategy for screening prospective blood donors for anaemia. Transfus Med 1994;4:221-5. https://doi.org/10.1111/j.1365-3148.1994.tb00275.x
  3. James V, Jones KF, Turner EM, Sokol RJ. Statistical analysis of inappropriate results from current Hb screening methods for blood donors. Transfusion 2003;43:400-4. https://doi.org/10.1046/j.1537-2995.2003.00316.x
  4. Ross DG, Gilfillan AC, Houston DE, Heaton WA. Evaluation of hemoglobin screening methods in prospective blood donors. Vox Sang 1986;50:78-80. https://doi.org/10.1111/j.1423-0410.1986.tb04850.x
  5. Noiri E, Kobayashi N, Takamura Y, Iijima T, Takagi T, Doi K, et al. Pulse total-hemoglobinometer provides accurate noninvasive monitoring. Crit Care Med 2005;33:2831-5. https://doi.org/10.1097/01.CCM.0000190430.96750.51
  6. Neville RG. Evaluation of portable haemoglobinometer in general practice. Br Med J (Clin Res Ed) 1987;294:1263-5. https://doi.org/10.1136/bmj.294.6582.1263
  7. Lotfi R, Wernet D, Starke U, Northoff H, Cassens U. A noninvasive strategy for screening prospective blood donors for anemia. Transfusion 2005;45:1585-92. https://doi.org/10.1111/j.1537-2995.2005.00574.x
  8. Gayat E, Aulagnier J, Matthieu E, Boisson M, Fischler M. Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies. PLoS One 2012;7:e30065. https://doi.org/10.1371/journal.pone.0030065
  9. Landis JR and Koch. The measurement of observer agreement for categorical data. Biometrics 1977;33:159-74. https://doi.org/10.2307/2529310
  10. Dhingra-Kumar N and Kela K. $CuSO_4$ gravimetric method for Hb screening of prospective donors--should it be discarded? Transfus Med 1997;7:245-7. https://doi.org/10.1046/j.1365-3148.1997.d01-34.x
  11. Sawant RB, Bharucha ZS, Rajadhyaksha SB. Evaluation of hemoglobin of blood donors deferred by the copper sulphate method for hemoglobin estimation. Transfus Apher Sci 2007;36:143-8. https://doi.org/10.1016/j.transci.2006.11.001
  12. Bhaskaram P, Balakrishna N, Radhakrishna KV, Krishnaswamy K. Validation of hemoglobin estimation using Hemocue. Indian J Pediatr 2003;70:25-8. https://doi.org/10.1007/BF02722739
  13. Neufeld L, Garcia-Guerra A, Sanchez-Francia D, Newton-Sanchez O, Ramirez-Villalobos MD, Rivera-Dommarco J. Hemoglobin measured by Hemocue and a reference method in venous and capillary blood: a validation study. Salud Publica Mex 2002;44:219-27. https://doi.org/10.1590/S0036-36342002000300005
  14. Prakash S, Kapil U, Singh G, Dwivedi SN, Tandon M. Utility of HemoCue in estimation of hemoglobin against standard blood cell counter method. J Assoc Physicians India 1999;47:995-7.
  15. Gordeuk VR, Brittenham GM, Bravo J, Hughes MA, Keating LJ. Prevention of iron deficiency with carbonyl iron in female blood donors. Transfusion 1990;30:239-45. https://doi.org/10.1046/j.1537-2995.1990.30390194345.x
  16. Cogswell ME, Parvanta I, Yip R, Brittenham GM. Innaccuracy of capillary hemoglobin testing for the detection of anemia and iron deficiency during pregnancy. Am J Epidemiol 1998;147(S5):206.
  17. Bahadur S, Jain S, Jain M. Estimation of hemoglobin in blood donors: A comparative study using hemocue and cell counter. Transfus Apher Sci 2010;43:155-7. https://doi.org/10.1016/j.transci.2010.07.010
  18. Conway AM, Hinchliffe RF, Earland J, Anderson LM. Measurement of haemoglobin using single drops of skin puncture blood: is precision acceptable? J Clin Pathol 1998;51:248-50. https://doi.org/10.1136/jcp.51.3.248
  19. Mills AF and Meadows N. Screening for anaemia: evaluation of a haemoglobinometer. Arch Dis Child 1989;64:1468-71. https://doi.org/10.1136/adc.64.10.1468

Cited by

  1. Accuracy of Continuous Noninvasive Hemoglobin Monitoring : A Systematic Review and Meta-Analysis vol.119, pp.2, 2013, https://doi.org/10.1213/ane.0000000000000272
  2. Evaluation of two‐step haemoglobin screening with HemoCue for blood donor qualification in mobile collection sites vol.107, pp.4, 2014, https://doi.org/10.1111/vox.12165
  3. Continuous noninvasive hemoglobin monitoring: ready for prime time? vol.21, pp.3, 2013, https://doi.org/10.1097/mcc.0000000000000197
  4. Systematic Review and Meta-Analysis of Method Comparison Studies of Masimo Pulse Co-Oximeters (Radical-7™ or Pronto-7™) and HemoCue® Absorption Spectrometers (B-H vol.43, pp.3, 2013, https://doi.org/10.1177/0310057x1504300310
  5. Comparison of three noninvasive methods for hemoglobin screening of blood donors vol.55, pp.2, 2013, https://doi.org/10.1111/trf.12819
  6. Performance analysis of newly developed point‐of‐care hemoglobinometer (TrueHb) against an automated hematology analyzer (Sysmex XT 1800i) in terms of precision in hemoglobin measurement vol.37, pp.4, 2015, https://doi.org/10.1111/ijlh.12314
  7. Point‐of‐care haemoglobin measurement – state of the art or a bleeding nuisance? vol.70, pp.11, 2013, https://doi.org/10.1111/anae.13231
  8. Evaluation of Two Noninvasive Hemoglobin Testing Devices vol.26, pp.3, 2013, https://doi.org/10.17945/kjbt.2015.26.3.273
  9. Assessment of a non-invasive haemoglobin sensor NBM 200 among pregnant women in rural India vol.2, pp.2, 2013, https://doi.org/10.1136/bmjinnov-2015-000085
  10. Hemoglobin assessment: precision and practicability evaluated in the Netherlands—the HAPPEN study vol.56, pp.8, 2016, https://doi.org/10.1111/trf.13546
  11. When visual inspection of the palpebral conjunctivae falls short … vol.56, pp.8, 2013, https://doi.org/10.1111/trf.13717
  12. Comparison of the NBM 200 non-invasive haemoglobin sensor with Sahli's haemometer among adolescent girls in rural India vol.2, pp.4, 2013, https://doi.org/10.1136/bmjinnov-2016-000139
  13. Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions vol.8, pp.None, 2013, https://doi.org/10.2147/jbm.s103788
  14. Capillary blood for point-of-care testing vol.54, pp.5, 2013, https://doi.org/10.1080/10408363.2017.1343796
  15. Non‐invasive evaluation of blood oxygen saturation and hematocrit from T1 and T2 relaxation times: In‐vitro validation in fetal blood vol.78, pp.6, 2013, https://doi.org/10.1002/mrm.26599
  16. Post marketing study of hemodynamic and hematological noninvasive readings in a blood bank vol.6, pp.None, 2018, https://doi.org/10.1177/2050312118796065
  17. Comparison of noninvasive and invasive point‐of‐care testing methods with reference method for hemoglobin measurement vol.32, pp.3, 2018, https://doi.org/10.1002/jcla.22309
  18. In Vitro Determination of Concentration-Dependent Optical Properties of Hemoglobin vol.4, pp.1, 2018, https://doi.org/10.1515/cdbme-2018-0162
  19. Evaluation of new non-invasive & conventional invasive methods of haemoglobin estimation in blood donors vol.149, pp.6, 2019, https://doi.org/10.4103/ijmr.ijmr_301_17
  20. Invasive and non-invasive point-of-care testing and point-of-care monitoring of the hemoglobin concentration in human blood - how accurate are the data? vol.64, pp.5, 2019, https://doi.org/10.1515/bmt-2018-0066
  21. Invasive and non-invasive point-of-care testing and point-of-care monitoring of the hemoglobin concentration in human blood - how accurate are the data? vol.64, pp.5, 2019, https://doi.org/10.1515/bmt-2018-0066
  22. Comparison of hemoglobin concentrations measured by HemoCue and a hematology analyzer in Indian children and adolescents 1‐19 years of age vol.42, pp.4, 2013, https://doi.org/10.1111/ijlh.13209
  23. Noninvasive Hemoglobin Level Prediction in a Mobile Phone Environment: State of the Art Review and Recommendations vol.9, pp.4, 2013, https://doi.org/10.2196/16806
  24. Comparison between blood hemoglobin concentration determined by point-of-care device and complete blood count in adult patients with dengue vol.15, pp.8, 2013, https://doi.org/10.1371/journal.pntd.0009692
  25. Reagent-free hemoglobin estimation on a spinning disc vol.168, pp.None, 2013, https://doi.org/10.1016/j.microc.2021.106463