Abstract

This study included 198 patients who underwent prospectively electrocardiogram-triggered high-pitch spiral mode(FLASH mode) in cardiac CT for health check: 123 patients who underwent manual exposure control(Group A) and 75 patients who underwent automatic exposure control(Group B). Patients were classified according to BMI grades: Grade 1, Grade 2, Grade 3, Grade 4. Radiation dose, objective and subjective image quality between two groups were compared. In Group B, tube voltage were significantly decreased in all BMI grades. Both $CTDI_{vol}$ and effective dose were significantly reduced in the BMI Grade 1, 3, and 4 whereas they were slightly reduced in the Grade 2(p>0.05), SNR was significantly decreased in the Grade 1 and increased in the Grade 3(p<0.05), but there was no significant difference in the Grade 2, 4 between the two groups(p>0.05). CNR was significantly decreased in the Grade 1(p<0.05), and there was no significant difference in the Grade 2, 3, 4 between the two groups(p>0.05). The subjective image quality showed no significant difference in all BMI Grades between the two groups(p>0.05). Automatic exposure control can lead to a significant reduction of radiation exposure dose without degradation of subjective image quality.

건강검진 목적으로 심장 CT검사를 시행한 198명의 대상자 중 수동노출조절 방법(n=123)의 대조군과 자동노출조절 방법(n=75)의 실험군으로 BMI에 따라 Grade 1(BMI < 23kg/$m^2$), Grade 2(23 $$\leq_-$$ BMI < 25kg/$m^2$), Grade 3(25 $$\leq_-$$ BMI < 30kg/$m^2$), Grade 4(30 $$\leq_-$$ BMI kg/$m^2$)로 분류하고 영상의 질을 객관적, 주관적 평가로 분석하였고 방사선량을 비교하였다. 자동노출조절 방법을 사용하였을 때 모든 Grade에서 관전압이 유의하게 감소하였고, $CTDI_{vol}$, effective dose 값은 Grade 2에서는 통계적으로 유의하지 않게 감소하였으나 Grade 1, 3, 4에서는 유의하게 감소하였다(p<0.05). SNR은 Grade 1에서 유의하게 감소하였고 Grade 3에서는 유의하게 증가하였지만(p<0.05) Grade 2와 4에서는 유의한 차이가 없었다(p>0.05). CNR은 Grade 1에서 유의하게 감소하였고(p<0.05), Grade 2, 3, 4에서는 유의한 차이는 없었다(p>0.05). 영상의 주관적인 평가에서도 두 그룹간의 모든 Grade에서 유의한 차이가 없었다(p>0.05). 결론적으로 자동노출조절 방법은 수동노출조절 방법에 비해 영상의 주관적 평가는 변하지 않으면서도 피폭선량을 줄일 수 있는 유용한 방법으로 나타났다.

Keywords

References

  1. N. R. Mollet, F. Cademartiri, C. A. van Mieghem, G. Runza, E. P. McFadden, and T. Banks, High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography, Circulation, Vol.112, pp.2318-2323, 2005. https://doi.org/10.1161/CIRCULATIONAHA.105.533471
  2. S. Leschka, H. Alkadhi, A. Plass, L. Desbiolles, J. Grunenfelder, and B. Marincek, Accuracy of MSCT coronary angiography with 64-slice technology: first experience, Eur Heart J, Vol.26, pp.1482-1487, 2005. https://doi.org/10.1093/eurheartj/ehi261
  3. L. Husmann, T. Schepis, H. Scheffel, O. Gaemperli, S. Leschka, and I. Valenta, Comparison of diagnostic accuracy of 64-slice computed tomography coronary angiography in patients with low, intermediate, and high cardiovascular risk, Acad Radiol, Vol.15, pp.452-461, 2008. https://doi.org/10.1016/j.acra.2007.12.008
  4. J. Hausleiter, T. Meyer, M. Hadamitzky, E. Huber, M. Zankl, and S. Martinoff, Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates, Circulation, Vol.113, pp.1305-1310, 2006. https://doi.org/10.1161/CIRCULATIONAHA.105.602490
  5. S. Leschka, P. Stolzmann, F. T. Schmid, H. Scheffel, B. Stinn, and B. Marincek, Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose, Eur Radiol, Vol.18, pp.1809-1817, 2008. https://doi.org/10.1007/s00330-008-0966-1
  6. Z. Szucs-Farkas, L. Kurmann, T. Strautz, M. A. Patak, P. Vock, and S. T. Schindera, Patient exposure and image quality of low-dose pulmonary computed tomography angiography: comparison of 100- and 80-kVp protocols, Invest Radiol, Vol.43, pp.871-876, 2008. https://doi.org/10.1097/RLI.0b013e3181875e86
  7. B. Bischoff, F. Hein, T. Meyer, M. Hadamitzky, S. Martinoff, A. Schomig, and J. Hausleiter, Impact of a reduced tube voltage on CT angiography and radiation dose: results of the PROTECTION I study, JACC Cardiovasc Imaging, Vol.2, pp.940-946, 2009. https://doi.org/10.1016/j.jcmg.2009.02.015
  8. H. T. Abada, C. Larchez, B. Daoud, A. Sigal-Cinqualbre, and J. F. Paul, MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose, AJR Am J Roentgenol, Vol.186, pp.387-390, 2006. https://doi.org/10.2214/AJR.05.0216
  9. J. Hausleiter, S. Martinoff, M. Hadamitzky, E. Martuscelli, I. Pschierer, G. M. Feuchtner, P. Catalan-Sanz, B. Czermak, T. S. Meyer, F. Hein, B. Bischoff, M. Kuse, A. Schomig, and S. Achenbach, Image quality and radiation exposure with a low tube voltage protocol for coronary CT angiography results of the PROTECTION II Trial, JACC Cardiovasc Imaging, Vol.3, pp.1113-1123, 2010. https://doi.org/10.1016/j.jcmg.2010.08.016
  10. T. R. Johnson, K. Nikolaou, and S. Busch, Diagnostic accuracy of dual source computed tomography in the diagnosis of coronary artery disease, Invest Radiol, Vol.42, pp.684-691, 2007. https://doi.org/10.1097/RLI.0b013e31806907d0
  11. F. Tatsugami, L. Husmann, B. A. Herzog, N. Burkhard, I. Valenta, and O. Gaemperli, Evaluation of a body mass index-adapted protocol for low-dose 64-MDCT coronary angiography with prospective ECG triggering, AJR Am J Roentgenol, Vol.192, pp.635-638, 2009. https://doi.org/10.2214/AJR.08.1390
  12. P. Stolzmann, S. Leschka, H. Scheffel, T. Krauss, L. Desbiolles, and A. Plass, Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose, Radiology, Vol.249, pp.71-80, 2008. https://doi.org/10.1148/radiol.2483072032
  13. H. Alkadhi, P. Stolzmann, H. Scheffel, L. Desbiolles, S. Baumuller, and A. Plass, Radiation dose of cardiac dual-source CT: the effect of tailoring the protocol to patient specific parameters, Eur J Radiol, Vol.68, pp.385-391, 2008. https://doi.org/10.1016/j.ejrad.2008.08.015
  14. 김윤경, 김유경. 체질량 지수 및 관전압 변화에 따른 CT 관상동맥 조영술의 영상의 질 및 방사선 피폭량 비교, 대한영상의학회지, 제62권, 제1호, pp.29-35, 2010. https://doi.org/10.3348/jksr.2010.62.1.29
  15. M. K. Kalra, M. M. Maher, T. L. Toth, B. Schmldt, B. L. Westerman, H. T. Morgan, and S. Salnl, Techniques and Applications of Automatic Tube Current Modulation for CT1, Radiology, Vol.233, pp.649-657, 2004. https://doi.org/10.1148/radiol.2333031150
  16. T. H. Mulkens, P. Bellinck, and M. Baeyaert, Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation, Radiology, Vol.237, pp.213-223, 2005. https://doi.org/10.1148/radiol.2363041220
  17. L. Yu, H. Li, J. G. Fletcher, C. H. McCulloug, Automatic selection of tube potential for radiation dose reduction in CT: a general strategy, Med Phys, Vol.37, No.1, pp.234-243, 2010. https://doi.org/10.1118/1.3264614
  18. Y. J. Park, Y. J. Kim, J. W. Lee, H. Y. Kim, Y. J. Hong, H. J. LEE, Automatic Tube Potential Selection With Tube Current Modulation (APSCM) in coronary CT angiography: Comparison of image quality and radiation dose with conventional body mass index-based protocol, Journal of Cardiovascular Computed Tomography, Vol.6, pp.184-190, 2012. https://doi.org/10.1016/j.jcct.2012.04.002
  19. 서재현, 김영덕, 정창기, 이종호, 유명송, Dual Source CT에서 관상동맥 검사시 환자의 피폭선량 감소에 대한 연구, 대한전산화단증기술학회지, Vol.14, No.1, pp.219-225, 2012.
  20. T. Pflederer, L. Rudofsky, D. Ropers, S. Bachmann, M. Marwan, and W. G. Daniel, Image quality in a low radiation exposure protocol for retrospectively ECG-gated coronary CT angiography, AJR Am J Roentgenol, Vol.192, pp.1045-1050, 2009. https://doi.org/10.2214/AJR.08.1025
  21. A. B. Sigal-Cinqualbre, R. Hennequin, H. T. Abada, X. Chen, and J. F. Paul, Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose, Radiology, Vol.231, pp.169-174, 2004. https://doi.org/10.1148/radiol.2311030191
  22. Y. Nakayama, K. Awai, Y. Funama, M. Hatemura, M. Imuta, and T. Nakaura, Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise, Radiology, Vol.237, pp.945-951, 2005. https://doi.org/10.1148/radiol.2373041655