DOI QR코드

DOI QR Code

The Effect of Water Droplets on the Nano Particle Size Distribution using the SMPS System

SMPS 시스템에서 용매(물)가 나노입도측정결과에 미치는 영향

  • 황보선애 (한국표준과학연구원 신기능재료표준센터) ;
  • 추민철 (한국표준과학연구원 신기능재료표준센터)
  • Received : 2013.03.12
  • Accepted : 2013.04.15
  • Published : 2013.04.28

Abstract

In this paper we have studied the effect of water droplet size on nano-particle size distribution using SMPS(Scanning Mobility Particle Sizer)system. It can be seen that the unknown peak at >100 nm was caused by water droplets which did not dry completely when DI water was used as a solvent in the SMPS system. Therefore, it is important to dry water droplets generated from atomizer in the SMPS system when measuring the particle size distribution using less than 100 nm nano-particles in diameter. From this study, It can be concluded that the napion was a useful material as dryer ones and using EAG(Electro Aerosol Generator) as a particle generator was the most effective in reducing the effect of water droplets.

Keywords

References

  1. M. S. Kim, K. H. Choi, Y. H. Kim and J. H. Yi: Clean Technology, 38 (2007) 159 (Korean).
  2. B.-K. Choi, K.-H. Kim, D.-S. So and I.-J. Yu: Prospectives of Industrial Chemistry, 10 (2007) 48.
  3. Y. Kim and J. Yi: News & Information for Chemical Engineers, 23 (2005) 338.
  4. A. D. Maynard: Ann. Occup. Hyg., 51 (2007) 1.
  5. A. D. Maynard: A research strategy for addressing risk, PEN 03. Project on Emerging Nanotechnologies (2006).
  6. Maynard, A. D., Aitken, R. J., Butz, T., Colvin, V., Donaldson, K., Oberdrster, G., Philbert, M. A., Ryan, J., Seaton, A., Stone, V., Tinkle, S. S., Tran, L., Walker, N. J., and Warheit, D. B.: Nature, 444 (2006) 267. https://doi.org/10.1038/444267a
  7. Maynard, A. D. and Kuempel, E. D.: J. Nanopart. Res., 7 (2005) 587. https://doi.org/10.1007/s11051-005-6770-9
  8. Kwangsik Park: J. Environ Toxicol, 20 (2005) 259 (Korean).
  9. Keum-Shil Cho, Jin-Sup Shin, and Jung-Hyun Kim: Polumer Science and technology, 15 (2004) 198 (Korean).