DOI QR코드

DOI QR Code

Clinical Image Evaluation of Film Mammograms in Korea: Comparison with the ACR Standard

  • Gwak, Yeon Joo (Department of Radiology, Kyungpook National University Medical Center) ;
  • Kim, Hye Jung (Department of Radiology, Kyungpook National University Medical Center) ;
  • Kwak, Jin Young (Department of Radiology and Research Institute of Radiological Science, Yonsei University College of Medicine) ;
  • Son, Eun Ju (Department of Radiology, Gangnam Severance Hospital, Yonsei University College of Medicine) ;
  • Ko, Kyung Hee (Department of Radiology, Bundang CHA Hospital, CHA University College of Medicine) ;
  • Lee, Jin Hwa (Department of Radiology, Dong-A University College of Medicine) ;
  • Lim, Hyo Soon (Department of Radiology, Chonnam National University Medical School) ;
  • Lee, You Jin (Department of Radiology, Kyungpook National University Hospital) ;
  • Park, Ji Won (Department of Radiology, Kyungpook National University Medical Center) ;
  • Shin, Kyung Min (Department of Radiology, Kyungpook National University Medical Center) ;
  • Jang, Yun-Jin (Department of Radiology, Kyungpook National University Hospital)
  • Received : 2012.05.18
  • Accepted : 2013.03.06
  • Published : 2013.09.01

Abstract

Objective: The goal of this study is to compare the overall quality of film mammograms taken according to the Korean standards with the American College of Radiology (ACR) standard for clinical image evaluation and to identify means of improving mammography quality in Korea. Materials and Methods: Four hundred and sixty eight sets of film mammograms were evaluated with respect to the Korean and ACR standards for clinical image evaluation. The pass and failure rates of mammograms were compared by medical facility types. Average scores in each category of the two standards were evaluated. Receiver operating characteristic curve analysis was used to identify an optimal Korean standard pass mark by taking the ACR standard as the reference standard. Results: 93.6% (438/468) of mammograms passed the Korean standard, whereas only 80.1% (375/468) passed the ACR standard (p < 0.001). Non-radiologic private clinics had the lowest pass rate (88.1%: Korean standard, 71.8%: ACR standard) and the lowest total score (76.0) by the Korean standard. Average scores of positioning were lowest (19.3/29 by the Korean standard and 3.7/5 by the ACR standard). A cutoff score of 77.0 for the Korean standard was found to correspond to a pass level when the ACR standard was applied. Conclusion: We suggest that tighter regulations, such as, raising the Korean pass mark, subtracting more for severe deficiencies, or considering a very low scores in even a single category as failure, are needed to improve the quality of mammography in Korea.

Keywords

References

  1. Korean National Cancer Center. Report on the statistics for national cancer incidence 2006-2007. Korean National Cancer Center, 2009
  2. Ko SS; Korean Breast Cancer Society. Chronological changing patterns of clinical characteristics of Korean breast cancer patients during 10 years (1996-2006) using nationwide breast cancer registration on-line program: biannual update. J Surg Oncol 2008;98:318-323 https://doi.org/10.1002/jso.21110
  3. Lee CH. Screening mammography: proven benefit, continued controversy. Radiol Clin North Am 2002;40:395-407 https://doi.org/10.1016/S0033-8389(01)00015-X
  4. Kim SH, Kim MH, Oh KK. Analysis and comparison of breast density according to age on mammogram between Korean and Western women. J Korean Radiol Soc 2000;42:1009-1014
  5. Bassett LW. The regulation of mammography. Semin Ultrasound CT MR 1996;17:415-423 https://doi.org/10.1016/S0887-2171(96)90028-1
  6. Suleiman OH, Spelic DC, McCrohan JL, Symonds GR, Houn F. Mammography in the 1990s: the United States and Canada. Radiology 1999;210:345-351 https://doi.org/10.1148/radiology.210.2.r99fe45345
  7. Destouet JM, Bassett LW, Yaffe MJ, Butler PF, Wilcox PA. The ACR's Mammography Accreditation Program: ten years of experience since MQSA. J Am Coll Radiol 2005;2:585-594 https://doi.org/10.1016/j.jacr.2004.12.005
  8. McLelland R, Hendrick RE, Zinninger MD, Wilcox PA. The American College of Radiology Mammography Accreditation Program. AJR Am J Roentgenol 1991;157:473-479 https://doi.org/10.2214/ajr.157.3.1872231
  9. Korean Society of Breast Imaging. A manual of the mammography quality control. Seoul: Sung Mun Gak, 2001
  10. Korean Society of Radiology. Guidance on installation and management of special medical equipments. Available from: http://www.radiology.or.kr. Accessed April 30, 2012
  11. Bassett LW, Farria DM, Bansal S, Farquhar MA, Wilcox PA, Feig SA. Reasons for failure of a mammography unit at clinical image review in the American College of Radiology Mammography Accreditation Program. Radiology 2000;215:698-702 https://doi.org/10.1148/radiology.215.3.r00jn32698
  12. Korean Institute for Accreditation of Medical Image. Available from: http://www.ikiami.or.kr/info/KMI418QD.aspx. Accessed April 30, 2012
  13. Korea Ministry of Health and Welfare. National health insurance system. Available from: http://english.mw.go.kr.Accessed April 30, 2012
  14. Li Y, Poulos A, McLean D, Rickard M. A review of methods of clinical image quality evaluation in mammography. Eur J Radiol 2010;74:e122-e131 https://doi.org/10.1016/j.ejrad.2009.04.069
  15. Lee SH, Choe YH, Chung SY, Kim MH, Kim EK, Oh KK, et al. Establishment of quality assessment standard for mammographic equipments: evaluation of phantom and clinical images. J Korean Radiol Soc 2005;53:117-127
  16. Moon WK, Kim TJ, Cha JH, Cho KS, Choi EW, Lee YJ, et al. Clinical image evaluation of mammograms: a national survey. J Korean Radiol Soc 2003;49:507-511
  17. National Institute of Food and Drug Safety Evaluation. Statistics of the diagnostic radiation generator devices installation in the last 5 years. 2010. Available from: http://www.nifds.go.kr. Accessed April 30, 2012

Cited by

  1. Clinical Image Quality in Daily Practice of Breast Cancer Mammography Screening vol.65, pp.3, 2013, https://doi.org/10.1016/j.carj.2014.02.001
  2. Mammography Quality in Turkey: Auditors’ Report on a Nationwide Survey vol.14, pp.1, 2017, https://doi.org/10.5812/iranjradiol.32936