Abstract
Lately, the number of interventional radiology is increased by the extension of procedure in medical radiation, and radiation exposure may be appeared differently by interventional radiologists, it is caused increase of radiation dose for radiation worker, patient, and radiologists. This study has done a comparative analysis characteristic of radiation exposure for five radiologists who executed interventional cardiology for 303 patients in S university hospital of Gyeong-Buk from Nov. 1, 2011 to Jan. 31, 2011. The average exposure time of five radiologists was 697.95sec. The average of cumulative DAP(exp) for patients was $52,730mGycm^2$ and the average of total DAP for patients was $104,875.14mGycm^2$. The average of frames for image was 855.52 frames in acquired images, and the average of frames for images was 802.2 frames in exposure images. They were statistically significant differences (p<0.05). Exposure time, cumulative DAP(fluro), cumulative DAP(exp), total DAP, acquired image, and exposure image were high correlation except cumulative DAP(exp), and acquired runs in x-ray exposure characteristics of machine. Exposure time was a great influence on radiologist. It signified that the more exposure time lead to the more radiation dose for radiologist. Radiation dose is related to ability, experience, difficulty, and precision of procedures in interventional procedure. The number of angiography and exposure time is difficult to control by radiologists. Therefore, it is in need of reasonable system which was evaluated the real dose of medical teams in interventional proceedings. We think that self education and training are required to reduce radiation dose for radiologists and radiation workers.
최근 방사선 의료에서 중재적 시술의 확대로 인하여 시술 건수가 증가하고 있으며, 시술자에 따라 방사선 노출이 다르게 나타 날 수 있으며 시술자와 환자가 받는 방사선 피폭의 증가를 가져온다. 본 연구는 2011년 11월 01일부터 2012년 01월 31일 까지 경북지역 S대학 병원에 내원한 303명의 심장내과 환자를 대상으로 중재적 방사선시술을 시행한 시술자 5명에 따른 방사선 노출특성을 비교분석하였다. 투시시간의 경우 5명의 시술의의 평균 투시시간은 697.95초, cumulative DAP(exp)의 경우 평균 환자 면적선량은 $52,730mGycm^2$, total DAP의 경우 평균 환자면적선량은 $104,875.14mGycm^2$, acquired image의 경우 평균 영상은 855.52frame, exposure image의 평균 영상은 802.2frame로 나타나 통계적으로 유의한 차이를 보였다(p<0.05). 장비의 X선 노출 특성 즉, 투시시간, cumulative DAP(fluro), cumulative DAP(exp), total DAP, acquired image, exposure image는 서로 높은 상관관계를 보이나 cumulative DAP(exp)와 acquired runs는 상관관계를 보이지 않았다. 시술자와 가장 큰 비중을 차지하는 것은 투시시간으로 나타나 투시시간이 길어질수록 방사선 피폭이 증가함을 의미한다. 시술시 피폭선량은 시술자에 따라 시술능력과 경험, 시술의 난이도 및 정밀 시술여부와 관련 있으며, 혈관 조영의 횟수와 투시시간이 정해진 것이 아니어서 인위적인 조절이 어려운 부분이라 할 수 있다. 따라서, 중재방사선 절차를 수행하는 의료진의 실질 피폭을 합리적으로 평가하는 시스템이 필요하며, 불필요한 피폭을 줄이기 위해 시술자를 비롯한 방사선 작업종사자에 대한 자체교육과 훈련이 필요할 것으로 사료된다.