대한방사선치료학회지 (The Journal of Korean Society for Radiation Therapy)
- 제24권2호
- /
- Pages.95-106
- /
- 2012
- /
- 1598-8449(pISSN)
비소세포폐암 III, IV기 환자에 있어서 Shell-Type Pseudo Target을 이용한 세기 조절 방사선치료계획기법의 유용성 평가
Assessment of the Usefulness of an IMRT Plan Using a Shell-Type Pseudo Target with Patients in Stage III or IV of NSCLC
- 이상봉 (경상대학교병원 방사선종양학과) ;
- 박기주 (경상대학교병원 방사선종양학과) ;
- 박두찬 (경상대학교병원 방사선종양학과) ;
- 김만우 (경상대학교병원 방사선종양학과) ;
- 김준곤 (경상대학교병원 방사선종양학과) ;
- 노성환 (경상대학교병원 방사선종양학과)
- Lee, Sang-Bong (Department of Radiation Oncology, Gyeongsang National University Hospital) ;
- Park, Ki-Ju (Department of Radiation Oncology, Gyeongsang National University Hospital) ;
- Park, Du-Chan (Department of Radiation Oncology, Gyeongsang National University Hospital) ;
- Kim, Man-Wo (Department of Radiation Oncology, Gyeongsang National University Hospital) ;
- Kim, Jun-Gon (Department of Radiation Oncology, Gyeongsang National University Hospital) ;
- Noh, Sung-Hwan (Department of Radiation Oncology, Gyeongsang National University Hospital)
- 투고 : 2012.03.28
- 심사 : 2012.08.17
- 발행 : 2012.09.30
초록
목 적: 비소세포폐암(Non-Small Cell Lung Cancer, NSCLC) III, IV기 환자의 방사선 치료 시 Shell-Type Pseudo Target의 유무에 따른 IMRT 치료계획의 유용성을 알아보고자 한다. 대상 및 방법: NSCLC III, IV기 환자 22명을 대상으로 Shell-Type Pseudo Target이 있을 때와(SPT) Shell-Type Pseudo Target이 없을 때(WSPT) 각각의 세기변조방사선치료(Intensity-Modulated Radiation Therapy, IMRT)계획을 세운 후, 선량-체적 히스토그람(Dose-Volume Histogram, DVH) 분석을 통해 종양 표적 내의 균일성 지수(Homogeneity Index, HI), 종양 표적에 대한 선량조형 지수(Conformity Index, CI), 척수 최대선량(Spinal Cord Maximum Dose), 식도에 대한
Purpose: The objective of this study was to investigate the usefulness of an IMRT treatment plan according to whether there was a shell-type pseudo target during radiation therapy for patients in Stage III or IV of non-small cell lung cancer (NSCLC). Materials and Methods: After setting an IMRT (Intensity-Modulated Radiation Therapy, IMRT) plan for when there was a shell-type pseudo target (SPT) and when there was none (WSPT) with 22 patients in Stage III or IV of NSCLC, the investigator analyzed dose-volume histograms (DVHs) and made assessment with dosimetric comparisons such as homogeneity index (HI) inside the tumor target, conformity index (CI) of the tumor target, spinal cord maximum dose, Esophagus