Development of Geometrical Quality Control Real-time Analysis Program using an Electronic Portal Imaging
전자포탈영상을 이용한 기하학적 정도관리 실시간 분석 프로그램의 개발
- Lee, Sang-Rok (Department of Radiation Oncology, Korea University Anam Hospital) ;
- Jung, Kyung-Yong (Department of Radiation Oncology, Korea University Anam Hospital) ;
- Jang, Min-Sun (Department of Radiation Oncology, Korea University Anam Hospital) ;
- Lee, Byung-Gu (Department of Radiation Oncology, Korea University Anam Hospital) ;
- Kwon, Young-Ho (Department of Radiation Oncology, Korea University Anam Hospital)
- 이상록 (고려대학교 안암병원 방사선종양학과) ;
- 정경용 (고려대학교 안암병원 방사선종양학과) ;
- 장민선 (고려대학교 안암병원 방사선종양학과) ;
- 이병구 (고려대학교 안암병원 방사선종양학과) ;
- 권영호 (고려대학교 안암병원 방사선종양학과)
- Received : 2012.03.28
- Accepted : 2012.08.10
- Published : 2012.09.30
Abstract
Purpose: To develop a geometrical quality control real-time analysis program using an electronic portal imaging to replace film evaluation method. Materials and Methods: A geometrical quality control item was established with the Eclipse treatment planning system (Version 8.1, Varian, USA) after the Electronic Portal Imaging Device (EPID) took care of the problems occurring from the fixed substructure of the linear accelerator (CL-iX, Varian, USA). Electronic portal image (single exposure before plan) was created at the treatment room's 4DTC (Version 10.2, Varian, USA) and a beam was irradiated in accordance with each item. The gaining the entire electronic portal imaging at the Off-line review and was evaluated by a self-developed geometrical quality control real-time analysis program. As for evaluation methods, the intra-fraction error was analyzed by executing 5 times in a row under identical conditions and procedures on the same day, and in order to confirm the infer-fraction error, it was executed for 10 days under identical conditions of all procedures and was compared with the film evaluation method using an Iso-align
목 적: 필름평가방법을 대신하여 전자포탈영상을 이용한 기하학적 정도관리 실시간 분석 프로그램을 개발하고자 한다. 대상 및 방법: 전자포탈영상장치(Electronic Portal Image Device, EPID)가 선형가속기(CL-iX, Varian, USA)의 하부에 고정된 구조 때문에 발생되는 문제를 해결한 후 Eclipse treatment planning system (Version 8.1, Varian, USA)으로 기하학적 정도관리 항목을 생성하였다. 치료실의 4DTC (Version 10.2, Varian, USA)에서 전자포탈영상(single exposure before plan)을 생성하고 항목별로 빔을 조사하였다. Off-line review에서 전자포탈영상을 획득하고 이를 자체 개발한 기하학적 정도관리 실시간 분석 프로그램으로 평가하였다. 평가 방법으로 치료 중 오차(Intra-fraction error)는 모든 절차를 같은 날에 동일 조건으로 5회 연속 실시하여 분석하였고 치료 간 오차(Inter-fraction error)의 확인을 위해 모든 절차를 동일 조건에서 10일간 실시한 후 Iso-align