실리콘겔 인공유방의 기계적 및 물리화학적 특성에 대한 연구

Study for Mechanical and Physicochemcial Properties of Silicone Gel Filled Mammary Implants

  • 백홍 (한국화학융합시험연구원 의료기기본부 의료용품팀) ;
  • 장동혁 (한국화학융합시험연구원 의료기기본부 의료용품팀) ;
  • 송정민 (한국화학융합시험연구원 의료기기본부 의료용품팀) ;
  • 이승영 (한국화학융합시험연구원 의료기기본부 의료용품팀) ;
  • 서무엽 (한국화학융합시험연구원 의료기기본부 의료용품팀) ;
  • 박길종 (한국화학융합시험연구원 의료기기본부 의료용품팀) ;
  • 맹은호 (한국화학융합시험연구원 의료기기본부 의료용품팀)
  • Baek, H. (Medical Device team, Medical Device Assessment Headquarter, Korea Testing & Research institute) ;
  • Jang, D.H. (Medical Device team, Medical Device Assessment Headquarter, Korea Testing & Research institute) ;
  • Song, J.M. (Medical Device team, Medical Device Assessment Headquarter, Korea Testing & Research institute) ;
  • Lee, S.Y. (Medical Device team, Medical Device Assessment Headquarter, Korea Testing & Research institute) ;
  • Seo, M.Y. (Medical Device team, Medical Device Assessment Headquarter, Korea Testing & Research institute) ;
  • Park, G.J. (Medical Device team, Medical Device Assessment Headquarter, Korea Testing & Research institute) ;
  • Maeng, E.H. (Medical Device team, Medical Device Assessment Headquarter, Korea Testing & Research institute)
  • 투고 : 2012.05.10
  • 심사 : 2012.06.21
  • 발행 : 2012.06.30

초록

The purpose of this study is to develop the guideline of the physicochemical and mechanical properties evaluation for silicone gel filled breast implants. First of all, the use and development status for silicone gel filled breast implants were investigated, and then, standard and criteria about performance evaluation established by the international organizations such as ASTM, FDA guidance and ISO were examined. To evaluate the mechanical properties, data research and testing for breaking strength, elongation, tensile set, joint intensity, silicone gel cohesion, weight loss from heating, static rupture resistance, impact resistance test, fatigue test, and gel bleed were performed. On the other hand, to evaluate the physicochemical properties, volatile matter, extent of cross linking, heavy metals, and extractable were analyzed. In this study, results for general function, mechanical properties and physicochemical properties were examined and reviewed for the accordance with international standard, and objective and standardized guideline was provided.

키워드

참고문헌

  1. ASTM D412, Standard Test Methods for Vulcanized rubber and Thermoplastic Elastomers - Tension. 2006.
  2. ASTM D624, Standard Test Methods for Tear strength of conventional vulcanized rubber and Thermoplastic Elastomers. 2007.
  3. ASTM F703, Standard Specification for Implantable Breast prostheses. 2007.
  4. ASTM F2051, Standard Specification for Implantable Saline filled Breast prostheses. 2007.
  5. FDA, Guidance for Industry and FDA staff : Saline, Silicone Gel, and Alternative Breast Implants. 2006.
  6. ISO 34-1, Rubber, vulcanized or thermoplastic - Determination of tear strength - Part I : Trouser, angel and cresecent test pieces. 2004.
  7. ISO 37, Rubber, vulcanized or thermoplastic - Determination of tensile stress-strain properties. 2005.
  8. ISO 14607, Non-active surgical implants - Mammary implants - Particular requirements. 2007.
  9. ISO 14949, Implants for surgery - Two-part addition-cure silicone elastomers. 2001.
  10. J. Sanchez-Guerrero, P.H. Schur, J.S. Sergent and M.H. Liang, "Silicone breast implants and rheumatic disease. Clinical, immunologic, and epidemiologic studies," Arthritis Rheum., vol. 37, pp. 158-168, 1994. https://doi.org/10.1002/art.1780370203
  11. D.A. Kessler, "The basis of the FDA's decision on breast implants," N Engl J Med., vol. 326, pp. 1713-1715, 1992. https://doi.org/10.1056/NEJM199206183262525
  12. M. Angell, "Breast implants - protection or paternalism?," N Engl J Med., vol. 326, pp. 1695-1696, 1992. https://doi.org/10.1056/NEJM199206183262510
  13. C.J. Burns, T.J. Laing and B.W. Gillespie, "The epidemiology of scleroderma among women: assessment of risk from exposure to silicone and silica.," J Rheumatol., vol. 23, pp. 1904-1911, 1996.
  14. I.M. Lee, N.R. Cook, N.A. Shadick, E. Pereira and J.E. Buring, "Prospective cohort study of breast implants and the risk of connective-tissue diseases," Int. J Epidemiol., vol. 40, pp. 230-238, 2011. https://doi.org/10.1093/ije/dyq164
  15. J.E. Mark, "Polymer Data Handbook," Oxford University Press, pp. 411-435, 1999.
  16. S. Necchi, D. Molina, S. Turri, F. Rossetto, and G. Pennati. "Failure of silicone gel breast implants: Is the mechanical weakening due to shell swelling a significant cause of prostheses rupture?," J Mech Behav Biomed Mater., vol. 4, pp. 2002-2008, 2011. https://doi.org/10.1016/j.jmbbm.2011.06.019
  17. M.W. Lieberman, E.D. Lykissa, R. Barrios, C.N. Ou, G. Kala, and S.V. Kala, "Cyclosiloxanes produce fatal liver and lung damage in mice," Environ Health Perspect., vol. 107, pp. 161-165, 1999. https://doi.org/10.1289/ehp.99107161
  18. S.V. Kala, E.D. Lykissa, M.W. Neely, and M.W. Lieberman., "Low molecular weight silicones are widely distributed after a single subcutaneous injection in mice," Am J Pathol., vol. 152, pp. 645-649, 1998.