DOI QR코드

DOI QR Code

치수혈류 측정을 위한 광 프로브 개발

Development of an Optical Probe for Measuring Blood Flow in Dental Pulp

  • 장경환 (연세대학교 생체공학협동과정) ;
  • 최준열 (연세대학교 의과학과) ;
  • 구정모 (연세대학교 생체공학협동과정) ;
  • 권민경 (연세대학교 의과학과) ;
  • 김덕원 (연세대학교 의과대학 의학공학교실)
  • 투고 : 2012.02.01
  • 심사 : 2012.07.10
  • 발행 : 2012.08.01

초록

To diagnose dental pulp vitality, electric pulp tester has been widely used, which is a method to test condition of nerve. However, especially in the case of patients with trauma, nerve desensitization could temporarily occur even though nerve might be recovered by blood flow within the pulp later, which implies that blood flow in dental pulp is also an important factor for diagnosing vitality. This paper described the development of a probe that relatively measured blood flow in dental pulp using photoplethysmography (PPG). The probe emits four different wavelength light sources including three visible and an infrared light. We tested which light source detect sensitively the blood flow in dental pulp. As a result, green light had the largest peak to peak voltage and the power spectrum among different wavelengths.

키워드

참고문헌

  1. H.J. Fulling and J.O. Anderson, "Influence of maturation status and tooth type of permanent teeth upon electrometric and thermal pulp testing," European Journal of Oral Sciences, vol. 84, no. 5 pp. 286-290, 1976. https://doi.org/10.1111/j.1600-0722.1976.tb00491.x
  2. I.G. Chanbers, "The role and methods of pulp testing in oral diagnosis; a review," International Endodontic Journal, vol. 15, no. 1, pp.1-15 1982. https://doi.org/10.1111/j.1365-2591.1982.tb01331.x
  3. J.Y. Lee, K. Yanpiset, A. Sigurdsson and W.F. Vann Jr, "Laser Doppler flowmetry for monitoring traumatized teeth," Dental Traumatology, vol. 17, no. 5, pp. 231-235, 2001.
  4. D.S. Ramsay, J. Artun and S.S. Martinen, "Reliability of pulpal blood flow measurements utilizing laser Doppler flowmetry," Journal of Dental Research, vol. 70, no. 11, pp. 1427-1430, 1991. https://doi.org/10.1177/00220345910700110601
  5. J.M. Schmitt, R.L. Webber and E.C. Walker, "Optical determination of dental pulp vitality," IEEE Trans Biomed Eng, vol. 38, no. 4, pp. 346-351, 1991. https://doi.org/10.1109/10.133229
  6. J.M. Schnettler and J.A. Walliace, "Pulse oximetry as a diagnostic tool of pulpal vitality," Journal of Endodontics, vol. 17, no. 10, pp. 488-490, 1991. https://doi.org/10.1016/S0099-2399(06)81795-4
  7. I. Sunada, "New Method for Measuring the Length of the Root canal," Journal of Dental Research, vol. 41, no. 2, pp.375-387, 1962. https://doi.org/10.1177/00220345620410020801
  8. Kim Hee Jin, Dental morphology, Seoul: Hyunmoon, p.16, 2007.
  9. J.G. Webster, Design of Pulse Oximeters, 2rd Edition, Philadelphia: Institute of Physics Publishing, pp.1-244, 1997.
  10. V. Sarkela, K.K. Harri, K. Oikarinen and E. Herrala, "Dental pulp vitality measurement based on muliwavelength photoplethysmography," SPIE, vol. 2631, pp. 72-83, 1995.
  11. T.M. Odor, T.R. Pitt Ford and F. McDonald, "Effect of probe design and bandwidth on laser doppler readings from vital and root-filled teeth," Medical Engineering and Physics, vol. 18, no. 5, pp. 359-364, 1996. https://doi.org/10.1016/1350-4533(95)00061-5
  12. T. Uzunov, P. Uzunova, I. Angelov and A. Gisbrecht, "Comparative investigation of the penetration of different wavelength visible LED radiation into dental tissue," SPIE, vol. 7027, 2008.
  13. Y. Mendelson and B.D. Ochs, "Noninvasive pulse oximetry utilizing skin reflectance photoplethy-smography," IEEE Transactions on Biomedical Engineering, vol. 35, no. 10, pp. 798-805, 1988. https://doi.org/10.1109/10.7286
  14. Rovert R.A, "Method and apparatus for detecting the presence of caries in teeth using visible luminescence," United States Patent, no. 31815, 1985.