Radiation Dose Reduction of Chest CT with Iterative Reconstruction in Image Space - Part I: Studies on Image Quality Using Dual Source CT

  • Hwang, Hye Jeon (Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine) ;
  • Seo, Joon Beom (Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine) ;
  • Lee, Jin Seong (Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine) ;
  • Song, Jae-Woo (Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine) ;
  • Kim, Song Soo (Department of Radiology, Chungnam National University Hospital, Chungnam National University School of Medicine) ;
  • Lee, Hyun Joo (Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine) ;
  • Lim, Chae Hun (Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine)
  • Published : 2012.12.01

Abstract

Objective: To determine whether the image quality (IQ) is improved with iterative reconstruction in image space (IRIS), and whether IRIS can be used for radiation reduction in chest CT. Materials and Methods: Standard dose chest CT (SDCT) in 50 patients and low dose chest CT (LDCT) in another 50 patients were performed, using a dual-source CT, with 120 kVp and same reference mAs (50 mAs for SDCT and 25 mAs for LDCT) employed to both tubes by modifying a dual-energy scan mode. Full-dose data were obtained by combining the data from both tubes and half-dose data were separated from a single tube. These were reconstructed by using a filtered back projection (FBP) and IRIS: full-dose FBP (F-FBP); full-dose IRIS (F-IRIS); half-dose FBP (H-FBP) and half-dose IRIS (H-IRIS). Objective noise was measured. The subjective IQ was evaluated by radiologists for the followings: noise, contrast and sharpness of mediastinum and lung. Results: Objective noise was significantly lower in H-IRIS than in F-FBP (p < 0.01). In both SDCT and LDCT, the IQ scores were highest in F-IRIS, followed by F-FBP, H-IRIS and H-FBP, except those for sharpness of mediastinum, which tended to be higher in FBP. When comparing CT images between the same dose and different reconstruction (F-IRIS/F-FBP and H-IRIS/H-FBP) algorithms, scores tended to be higher in IRIS than in FBP, being more distinct in half-dose images. However, despite the use of IRIS, the scores were lower in H-IRIS than in F-FBP. Conclusion: IRIS generally helps improve the IQ, being more distinct at the reduced radiation. However, reduced radiation by half results in IQ decrease even when using IRIS in chest CT.

Keywords

References

  1. Kalra MK, Maher MM, Toth TL, Schmidt B, Westerman BL, Morgan HT, et al. Techniques and applications of automatic tube current modulation for CT. Radiology 2004;233:649-657 https://doi.org/10.1148/radiol.2333031150
  2. Lell MM, May M, Deak P, Alibek S, Kuefner M, Kuettner A, et al. High-pitch spiral computed tomography: effect on image quality and radiation dose in pediatric chest computed tomography. Invest Radiol 2011;46:116-123 https://doi.org/10.1097/RLI.0b013e3181f33b1d
  3. McNitt-Gray MF. AAPM/RSNA Physics Tutorial for Residents: topics in CT. Radiation dose in CT. Radiographics 2002;22:1541-1553 https://doi.org/10.1148/rg.226025128
  4. Mulkens TH, Bellinck P, Baeyaert M, Ghysen D, Van Dijck X, Mussen E, et al. Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 2005;237:213-223 https://doi.org/10.1148/radiol.2363041220
  5. Prasad SR, Wittram C, Shepard JA, McLoud T, Rhea J. Standard-dose and 50%-reduced-dose chest CT: comparing the effect on image quality. AJR Am J Roentgenol 2002;179:461-465 https://doi.org/10.2214/ajr.179.2.1790461
  6. Rothenberg LN, Pentlow KS. Radiation dose in CT. Radiographics 1992;12:1225-1243 https://doi.org/10.1148/radiographics.12.6.1439023
  7. Schenzle JC, Sommer WH, Neumaier K, Michalski G, Lechel U, Nikolaou K, et al. Dual energy CT of the chest: how about the dose? Invest Radiol 2010;45:347-353
  8. Wildberger JE, Mahnken AH, Schmitz-Rode T, Flohr T, Stargardt A, Haage P, et al. Individually adapted examination protocols for reduction of radiation exposure in chest CT. Invest Radiol 2001;36:604-611 https://doi.org/10.1097/00004424-200110000-00006
  9. Yu L, Li H, Fletcher JG, McCollough CH. Automatic selection of tube potential for radiation dose reduction in CT: a general strategy. Med Phys 2010;37:234-243 https://doi.org/10.1118/1.3264614
  10. Dinkel HP, Sonnenschein M, Hoppe H, Vock P. Low-dose multislice CT of the thorax in follow-up of malignant lymphoma and extrapulmonary primary tumors. Eur Radiol 2003;13:1241-1249
  11. Mayo JR, Hartman TE, Lee KS, Primack SL, Vedal S, Muller NL. CT of the chest: minimal tube current required for good image quality with the least radiation dose. AJR Am J Roentgenol 1995;164:603-607 https://doi.org/10.2214/ajr.164.3.7863879
  12. Mayo JR, Kim KI, MacDonald SL, Johkoh T, Kavanagh P, Coxson HO, et al. Reduced radiation dose helical chest CT: effect on reader evaluation of structures and lung findings. Radiology 2004;232:749-756 https://doi.org/10.1148/radiol.2323030899
  13. Takahashi M, Maguire WM, Ashtari M, Khan A, Papp Z, Alberico R, et al. Low-dose spiral computed tomography of the thorax: comparison with the standard-dose technique. Invest Radiol 1998;33:68-73 https://doi.org/10.1097/00004424-199802000-00002
  14. Yamada T, Ono S, Tsuboi M, Saito H, Sato A, Matsuhashi T, et al. Low-dose CT of the thorax in cancer follow-up. Eur J Radiol 2004;51:169-174 https://doi.org/10.1016/j.ejrad.2003.09.017
  15. Gregor J, Benson T. Computational analysis and improvement of SIRT. IEEE Trans Med Imaging 2008;27:918-924 https://doi.org/10.1109/TMI.2008.923696
  16. Lasio GM, Whiting BR, Williamson JF. Statistical reconstruction for x-ray computed tomography using energy-integrating detectors. Phys Med Biol 2007;52:2247-2266 https://doi.org/10.1088/0031-9155/52/8/014
  17. Xu J, Tsui BM. Electronic noise modeling in statistical iterative reconstruction. IEEE Trans Image Process 2009;18:1228-1238 https://doi.org/10.1109/TIP.2009.2017139
  18. AAPM. AAPM Report No. 96: the Measurement, Reporting, and Management of Radiation Dose in CT. College Park: American Association of Physicists in Medicine, 2008
  19. Ravenel JG, Scalzetti EM, Huda W, Garrisi W. Radiation exposure and image quality in chest CT examinations. AJR Am J Roentgenol 2001;177:279-284 https://doi.org/10.2214/ajr.177.2.1770279
  20. Pontana F, Duhamel A, Pagniez J, Flohr T, Faivre JB, Hachulla AL, et al. Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients. Eur Radiol 2011;21:636-643 https://doi.org/10.1007/s00330-010-1991-4
  21. Pontana F, Pagniez J, Flohr T, Faivre JB, Duhamel A, Remy J, et al. Chest computed tomography using iterative reconstruction vs filtered back projection (Part 1): evaluation of image noise reduction in 32 patients. Eur Radiol 2011;21:627-635 https://doi.org/10.1007/s00330-010-1990-5
  22. Prakash P, Kalra MK, Digumarthy SR, Hsieh J, Pien H, Singh S, et al. Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience. J Comput Assist Tomogr 2010;34:40-45 https://doi.org/10.1097/RCT.0b013e3181b26c67
  23. Hwang HJ, Seo JB, Lee JS, Song JW, Kim SS, Lee HJ, et al. Radiation dose reduction of chest CT with iterative reconstruction in image space - Part II: assessment of radiologists' preferences using dual source CT. Korean J Radiol 2012;13:720-727 https://doi.org/10.3348/kjr.2012.13.6.720
  24. Kyriakou Y, Kalender WA. Intensity distribution and impact of scatter for dual-source CT. Phys Med Biol 2007;52:6969-6989 https://doi.org/10.1088/0031-9155/52/23/014

Cited by

  1. Radiation Dose Reduction of Chest CT with Iterative Reconstruction in Image Space - Part II: Assessment of Radiologists' Preferences Using Dual Source CT vol.13, pp.6, 2012, https://doi.org/10.3348/kjr.2012.13.6.720
  2. Radiation Dose Reduction via Sinogram Affirmed Iterative Reconstruction and Automatic Tube Voltage Modulation (CARE kV) in Abdominal CT vol.14, pp.6, 2013, https://doi.org/10.3348/kjr.2013.14.6.886
  3. Low-Dose Chest Computed Tomography With Sinogram-Affirmed Iterative Reconstruction, Iterative Reconstruction in Image Space, and Filtered Back Projection: Studies on Image Quality vol.37, pp.4, 2012, https://doi.org/10.1097/rct.0b013e31828f4dae
  4. Image quality assessment of ultra low-dose chest CT using sinogram-affirmed iterative reconstruction vol.24, pp.4, 2014, https://doi.org/10.1007/s00330-013-3090-9
  5. Adaptive Iterative Dose Reduction Algorithm in CT: Effect on Image Quality Compared with Filtered Back Projection in Body Phantoms of Different Sizes vol.15, pp.2, 2014, https://doi.org/10.3348/kjr.2014.15.2.195
  6. Effects of Dual-Energy CT with Non-Linear Blending on Abdominal CT Angiography vol.15, pp.4, 2014, https://doi.org/10.3348/kjr.2014.15.4.430
  7. Adaptive Iterative Dose Reduction Using Three Dimensional Processing (AIDR3D) Improves Chest CT Image Quality and Reduces Radiation Exposure vol.9, pp.8, 2012, https://doi.org/10.1371/journal.pone.0105735
  8. Radiation Dose and Imaging Quality of Abdominal Computed Tomography before and after Scan Protocol Adjustment: Single-Institution Experience in Three Years vol.71, pp.6, 2014, https://doi.org/10.3348/jksr.2014.71.6.278
  9. Chest CT using spectral filtration: radiation dose, image quality, and spectrum of clinical utility vol.25, pp.6, 2012, https://doi.org/10.1007/s00330-014-3559-1
  10. Ultra-Low-Dose CT of the Thorax Using Iterative Reconstruction: Evaluation of Image Quality and Radiation Dose Reduction vol.204, pp.6, 2012, https://doi.org/10.2214/ajr.14.13629
  11. The Impact of Iterative Reconstruction on Computed Tomography Radiation Dosimetry: Evaluation in a Routine Clinical Setting vol.10, pp.9, 2012, https://doi.org/10.1371/journal.pone.0138329
  12. Combined use of iterative reconstruction and monochromatic imaging in spinal fusion CT images vol.58, pp.1, 2012, https://doi.org/10.1177/0284185116631182
  13. Radiation dose reduction in chest dual-energy computed tomography: effect on image quality and diagnostic information vol.51, pp.6, 2012, https://doi.org/10.1590/0100-3984.2017.0136