DOI QR코드

DOI QR Code

Usefulness of Myotonometer for Measurement of Tissue Compliance on Medialis Gastrocnemius in Patients with Stroke

뇌졸중 환자에서 내측 비복근의 조직탄성 측정을 위한 Myotonometer의 유용성

  • Bae, Sea-Hyun (Dept. of Physical Therapy, Gwangju Heemang Hospital) ;
  • Lee, Jeong-In (Dept. of Physical Therapy, Dongshin University Oriental Hospital) ;
  • Kim, Kyung-Yoon (Dept. of Physical Therapy, Dongshin University)
  • 배세현 (광주희망병원 물리치료실) ;
  • 이정인 (동신대학교 목포한방병원 물리치료실) ;
  • 김경윤 (동신대학교 물리치료학과)
  • Received : 2012.01.05
  • Accepted : 2012.03.08
  • Published : 2012.03.31

Abstract

The aim of this study was to find useful parameters of the between myotonometer and surface electromyography(sEMG) on the medialis gastrocnemius of stroke patients and investigate between these parameters and modified Ashworth scale(MAS) relationship. 5 years clinical experience physical therapist using the Modified Ashworth Scale(MAS) was selected 15 patients with ankle spasticity and divided randomly MAS2, MAS3, MAS4 groups. Myotonometer and sEMG was measured during relaxed and maximum voluntary contractions of the gastrocnemius muscle. The results of this study, the higher MAS score was the lowered the relaxation and contraction state tissue compliance and muscle activity and in the correlation analysis the higher MAS score during voluntary contraction in the cylinder receiving low-intensity correlation could see that increased than relaxation. Therefore, the myotonometer is a useful clinical and research tool with spasticity muscle and can provide objective quantitative data about the efficacy of physical therapy interventions.

본 연구의 목적은 뇌졸중 환자의 족관절 저측굴근 경직에 대하여 Myotonometer와 surface Electromyography(sEMG)를 이용하여 경직의 정량적 평가에 유용한 지표를 알아보고, 이 지표들과 임상적으로 평가되는 modified Ashworth scale(MAS)과의 관계를 알아보고자 하였다. 족관절 저측굴근의 경직평가는 물리치료사 5명(임상경력 5년이상)이 MAS를 이용하여 MAS 2, 3, 4 해당 군에 각 5명씩을 무작위로 15명씩 배정하였다. 각 군의 조직탄성과 근활성 측정은 Myotonometer와 sEMG로 이완(relaxed)상태와 최대 수의적 수축(contracted)상태에서 측정하였다. 연구 결과, MAS 등급이 높아짐에 따라 이완과 수축상태 간의 조직저항도 전위차와 근활성도의 차이는 작아졌고, 상관관계 분석에서도 MAS 등급이 높아질수록 이완 시보다 수의적 수축 시 실린더가 받는 낮은 강도에서의 상관성이 더 높아짐을 확인할 수 있었다. 따라서 Myotonometer는 경직을 평가하는 방법으로 다양한 인체의 관절에 비교적 쉽게 적용할 수 있으며 경직의 변화를 민감하게 반영할 수 있어 보다 객관적이고 정량적인 경직의 평가 도구로 사용될 수 있을 것이다.

Keywords

References

  1. K. L. Harburn, K. M Hill, A. A. Vandervoort, A. Helewa, C. H. Goldsmith, A. Kertesz, R. W. Teasell. "Spasticity measurement in stroke: a pilot study", Can J Public Health, Vol. 2, pp.41-45, 1992.
  2. W. Z. Rymer and R. T. Katz. "Mechanisms of spastic hypertonia", Phys Med Rehabil, Vol. 8, pp. 441-454, 1994.
  3. S. J. Lee, T. R. Han. "A quantitative assessment of spasticity in hemiplegic patients using isokinetic dynamometer", J Korean Acad Rehab Med, Vol. 22(4), pp.784-792, 1998.
  4. C. T. Leonard. "Motor behavior and neural changes following perinatal and adult-onset brain damage: implications for therapeutic interventions", Phys Ther, Vol. 74(8), pp. 753-767, 1994. https://doi.org/10.1093/ptj/74.8.753
  5. V. Dietz, W. Berger. "Normal and impaired regulation of muscle stiffness in gait: a new hypothesis about muscle hypertonia", Exp Neurol, Vol. 79(3), pp. 680-687, 1983. https://doi.org/10.1016/0014-4886(83)90032-8
  6. D. H. Sung, J. M. Kim, K. W. Lee, S. Y. Kim, Y. J Ra, J. Y. Jung, H. D. Park, S. J. Yoo. "A quantitative assessment of knee extensor spasticity using isokinetic dynamometer", J Korean Acad Rehab Med, Vol. 26(3), pp. 321-326, 2002.
  7. B. M. Haas, E. Bergstrom, A. Jamous, A. Bennie, "The interrater reliability of the original and of the modified Ashworth scale for the assessment of spasticity in patients with spinal cord injury", Spinal Cord, Vol. 34 pp. 560-564, 1996. https://doi.org/10.1038/sc.1996.100
  8. C. T. Leonard, J. U. Stephens, S. L. Stroppel. "Assessing the spastic condition of individuals with upper motoneuron involvement: validity of the myotonometer", Arch Phys Med Rehabil, Vol. 82(10), pp. 1416-1420, 2001. https://doi.org/10.1053/apmr.2001.26070
  9. R. T. Katz, W. Z. Rymer. "Spastic hypertonia: mechanisms and measurement", Arch Phys Med Rehabil, Vol. 70(2), pp. 144-155, 1989.
  10. J. S. Worley, W. Bennett, G. Miller, M. Miller, B. Walker, C. Harmon. "Reliability of three clinical measures of muscle tone in the shoulders and wrists of post stroke patients", Am J Occup Ther, Vol. 45(1), pp. 50-58, 1991. https://doi.org/10.5014/ajot.45.1.50
  11. E. W. Jones, G. P. Mulley. "The measurement of spasticity In: Rose FC, editor. Advances in stroke therapy", New York: Raven Pr; pp. 187-195, 1982.
  12. C. T. Leonard, J. S. Brown, T. R. Price, S. A. Queen, E. L. Mikhailenok. "Comparison of surface electromyography and myotonometer measurements during voluntary isometric contractions", J Electromyogr Kinesiol, Vol. 14(6), pp. 709-714, 2004. https://doi.org/10.1016/j.jelekin.2004.06.001
  13. C. T. Leonard, W. P. Deshner, J. W. Romo, E. S. Suoja, S. C. Fehrer, E. L. Mikhailenok. "Myotonometer intra- and interrater reliabilities", Arch Phys Med Rehabil, Vol. 84(6), pp. 928-932, 2003. https://doi.org/10.1016/S0003-9993(03)00006-6
  14. A. A. Fischer. "Pressure algometry over normal muscles", Standard values, validity and reproducibility of pressure threshold. Pain, Vol. 30(1), pp. 115-126, 1987b. https://doi.org/10.1016/0304-3959(87)90089-3
  15. F. Sakai, S. Ebihara, M. Akiyama, M. Horikawa. "Pericranial muscle hardness in tension-type headache. A non-invasive measurement method and its clinical application", Brain, Vol. 118(2), pp. 523-531 1995. https://doi.org/10.1093/brain/118.2.523
  16. S. J. Rydahl, B.J. Brouwer. "Ankle stiffness and tissue compliance in stroke survivors; A validation of Myotonometer measurements. Arch Phys Med Rehabil 85(10) pp. 1631-1637, 2004. https://doi.org/10.1016/j.apmr.2004.01.026
  17. S. Y. Kim. "Intra-Rater and Inter-Rater Reliability of the Myotonometer in the Assessment of Biceps Brachii and Quadriceps" Korean Academy of university trained physical therapists, 14(2), pp. 29-35, 2007.
  18. A. Shumway-Cook, M. Woollacott. Motor Control: Theory and practical applications. Baltimore: Williams & Wilkins, 1995.
  19. Neurogenic Technologies Inc,. The$myotonometer^{(R)}$. A device for objective measurement of muscle tone. 5000 Pattee Canyon, St, Missoula, MT 59803.
  20. I. J. Baquley, K. M. Bailey, S. Slewa-Younan. "Prolonged anti-spasticity effects of bolus intrathecal baclofen", Brain Inj, Vol. 19(7), pp. 545-548, 2005. https://doi.org/10.1080/02699050400025075
  21. M. Murayama, K. Nosaka, T. Yoneda, K. Minamitani. "Changes in hardness of the human elbow flexor muscles after eccentric exercise" Eur J Appl Physiol, Vol. 82, pp. 361-367, 2000. https://doi.org/10.1007/s004210000242
  22. A. Dario, and G. Tomei. "A benefit-risk assessment of baclofen in severe spinal spasticity", Drug Saf, Vol. 27(11), pp. 799-818, 2004. https://doi.org/10.2165/00002018-200427110-00004
  23. A. D. Pandyan, M. Gregoric, M. P. Barnes, D. Wood, F. Van Wijck, J. Burridge, H. Hermens, and G. R. Johnson, "Spasticity: clinical perceptions, neurological realities and meaningful measurement", Disabil Rehabil, Vol. 27(1-2), pp. 2-6, 2005. https://doi.org/10.1080/09638280400014576

Cited by

  1. A Comparison of The Effect of Stretching Technique on Hamstring Muscle for Flexibility, Strength, Pressure Pain Threshold Value and Muscle Tone vol.12, pp.4, 2017, https://doi.org/10.13066/kspm.2017.12.4.39