DOI QR코드

DOI QR Code

TLD dose variation of Magnetic Resonance Imaging Equipment

자기공명영상장비에서 열형광선량계의 선량 변화

  • Je, Jaeyong (Department of Radiological Technology, Dong-Eui Institute of Technology) ;
  • Kang, Eunbo (Department of Radiological Technology, Dong-Eui Institute of Technology)
  • 제재용 (동의과학대학교 방사선과) ;
  • 강은보 (동의과학대학교 방사선과)
  • Received : 2012.10.12
  • Accepted : 2012.12.21
  • Published : 2012.12.30

Abstract

The PET-MRI which has been installed and being managed recently uses both magnetic field and radiation. Most radiation workers wear a thermoluminescenct dosimeter (TLD) as a personal radiation dosimeter, and the TLD is affected both by a magnetic field and radiation. In this research, the same amount of X-ray was applied to 36 TLDs, and the changes in the dose of the 32 TLDs exposed to magnetic field at the location where its intensity of the magnetic resonance imaging (MRI) was about 5000 Gauss for eight hours with one-hour unit and that of the four TLDs not exposed to magnetic field were compared and checked. The measurement result showed that exposure dose of the TLD attached to the MRI changed irregularly depending on the amount of exposure time. Therefore, the TLD whose amount of changes little in the environment of a MRI is demanded to be developed.

최근 설치되어 운영되어지고 있는 PET-MRI는 자기장과 방사선을 함께 사용하고 있다. 방사선 작업종사자는 대부분 개인피폭선량계로 열형광선량계(TLD)를 착용하고 있고, TLD는 자기장과 방사선 영향을 동시에 받는다. 본 연구에서는 36개의 TLD에 동일 선량의 X-선을 조사하고, 자기공명영상장비의 자기장 세기가 5000 Gauss 정도인 위치에 1시간 단위로 8시간 동안 노출시킨 32개의 TLD와 자기장에 노출되지않은 4개의 선량변화를 확인하였다. 측정 결과 자기공명영상장비에 부착된 TLD의 피폭선량은 노출시간에 따라 불규칙한 선량 변화를 나타내었다. 따라서 자기공명영상장비 환경에서 선량변화가 작은 열형광선량계의 개발이 요구되어진다.

Keywords

References

  1. Oberhofer M and Scharmnn, "Applied Thermolminescence Dosimetry", Pergamon Press, Oxford, 1981.
  2. Wiedemann E and Schmit GC, Ann. Phys. Chem. Neue Folge, Vol. 54, pp.604, 1895.
  3. Randall JT and Wilkins MHF,"Phosphorescence and electron traps I. The study of trap distribution", Proc, Roy. Soc, Vol. A184, pp.366, 1945.
  4. Garlick GFJ and Gibson AF, "The Electron Trap Mechanism of Luminescence in Sulphide and Silicate Phosphors", Proc. Phys. Soc. Vol. 60, pp.574, 1948. https://doi.org/10.1088/0959-5309/60/6/308
  5. 제재용, "자기장이 심부선량과 표층선량에 미치는 영향", 한국방사선학회 춘계종합학술대회 초록집. Vol. 5, No. 1, pp.105-107, 2001.
  6. Tiwari B, Rawat NS, Desai DG, Singh SG, Tyagi M, Ratna P, Gadkari SC, Kulkarni MS, "Thermoluminescence studies on CI-doped Li2B4O7 single crystals", Journal of luminescence, Vol. 103, No. 11, pp.2076-2083, 2010.
  7. 권경재, "CaSo4(Dy, Tm) 열형광체의 제작과 특성에 관한 연구", 경남대학교 석사학위논문, pp.19, 1993.

Cited by

  1. The Effects of Magnetic Field on TLD Glow Curve vol.7, pp.6, 2013, https://doi.org/10.7742/jksr.2013.7.6.415
  2. 시간 경과에 따른 자기장 노출 유·무 환경에서 열형광선량계의 글로우 곡선 및 피폭 방사선량 분석 vol.16, pp.8, 2012, https://doi.org/10.5392/jkca.2016.16.08.419