Development of Oil Flowmeter and Preliminary Evaluation for Establishing National Leak Standard

국가 리크표준기용 오일유량계 개발 및 예비 평가

  • Hong, S.S. (Center for Mass Relative Quantities, Korea Research Institute of Standards and Science) ;
  • Song, H.W. (Center for Mass Relative Quantities, Korea Research Institute of Standards and Science) ;
  • Choi, I.M. (Center for Mass Relative Quantities, Korea Research Institute of Standards and Science) ;
  • Park, Y.K. (Center for Mass Relative Quantities, Korea Research Institute of Standards and Science)
  • 홍승수 (한국표준과학연구원 기반표준본부) ;
  • 송한욱 (한국표준과학연구원 기반표준본부) ;
  • 최인묵 (한국표준과학연구원 기반표준본부) ;
  • 박연규 (한국표준과학연구원 기반표준본부)
  • Received : 2012.09.25
  • Accepted : 2012.11.08
  • Published : 2012.11.30

Abstract

We preliminarily evaluated the performance of a flowmeter by the measurement of the pressure of a ionization gauge and the produced throughput (Q) using a vacuum system and pistons with the radius of 10 mm and 20 mm. The throughput range with the later piston was from $1{\times}10^{-1}Pa{\cdot}L/s$ to $1.2Pa{\cdot}L/s$ and that with the former piston was from $5.3{\times}10^{-1}Pa{\cdot}L/s$ to $1.05Pa{\cdot}L/s$. In these ranges, both pistons showed good repeatability and linearity. The developed oil flowmeter as the leak standard will be used in the calibration of standard leak.

지름이 각각 10 mm와 20 mm인 피스톤과 진공장치를 이용해서 발생된 유량과 이온게이지의 압력을 측정하여 비교하는 방법으로 개발 중인 오일유량계의 성능을 예비 평가하였다. 큰 피스톤의 유량 발생 범위는 $1{\times}10^{-1}Pa{\cdot}L/s{\sim}1.2Pa{\cdot}L/s$이었고 작은 피스톤의 유량발생 범위는 $5.3{\times}10^{-1}Pa{\cdot}L/s{\sim}1.05Pa{\cdot}L/s$이었으며, 이 구간에서의 반복성과 직선성은 비교적 좋았다. 개발 중인 오일유량계는 국가 리크표준기로서 앞으로 표준리크를 교정하는데 활용할 계획이다.

Keywords

References

  1. S. S. Hong, Y. H. Shin, and K. H. Chung, Handbook of Vacuum Measurements (Chungmoongak, 2010), p. 151.
  2. C. D. Ehrlich and S. A. Tison, NIST Leak Calibration Service, NIST Special Publication 250-38, 3 (1992).
  3. AVS 2.1, Calibration of Leak Detectors of the Mass Spectrometer Type 568 (1973).
  4. W. Jitschin, G. Grosse, and D. Wandrey, Vacuum 38, 883 (1988). https://doi.org/10.1016/0042-207X(88)90485-X
  5. Y. H. Shin, S. S. Hong, I. T. Lim, and K. H. Chung, Key Engineering Materials 270-273, 1674 (2004). https://doi.org/10.4028/www.scientific.net/KEM.270-273.1674
  6. S. H. Bae, S. Y. In, K. H. Chung, Y. B. Lee, and Y. H. Shin, Korea Economy Newspaper, Vacuum Engineering 540 (2000).
  7. K. H. Chung, S. S. Hong, Y. H. Shin, J. Y. Lim, S. K. Lee, and S. Y. Woo, Metrologia 36, 675 (1999). https://doi.org/10.1088/0026-1394/36/6/37
  8. K. H. Chung and S. S. Hong, Establishment of ultrahigh vacuum calibration system down to Pa range in Korea (XIII IMEKO World Congress, Torino, Italy, 1994) p. 1951.
  9. S. S. Hong, I. T. Lim, Y. H. Shin, and K. H. Chung, J. Korean Vac. Soc. 12, 7 (2003).
  10. S. S. Hong, I. T. Lim, and Y. H. Shin, J. Korean Vac. Soc. 15, 347 (2006).