Abstract

The selection of the temperature sensor in the cryogenic system depends on the temperature range, shape and accuracy. The accuracy of the temperature sensor is essential to improve the reliability of experiment. We have developed the variable temperature cryostat using a two-stage cryocooler. In order to reduce heat load, thermal shield is installed at the first stage with MLI (Multiple layer insulation). We have also developed the sensor holder calibrating more than twenty sensors at the same time for saving time and money. The system can calibrate sensor at variable temperature by controlling electric heater. In this paper, we present design and fabrication of variable temperature cryostat and representative result of Cernox sensor calibration.

Keywords

References

  1. A. Szmyka-Grzebyk, et. Al., "Measuring Systems for Thermometer Calibration in Low-Temperature Range," SpringerLink Int J Thermlphys 32, pp. 2466-2476, 2011. https://doi.org/10.1007/s10765-011-1055-2
  2. B. W. A Rickeston and R. E. J Watkins, "The 27${\Omega}$Rhodium-iron ceramic sensors,"Cryogenics, Vol. 49, pp. 320-325, 2009. https://doi.org/10.1016/j.cryogenics.2009.03.006
  3. Kamran Souri and Kofi Makinwa, "Ramp Calibration of Temperature Sensors" IEEE 4th international workshop, pp. 67-70, 2011.
  4. F. P. Incropera and D. P. Dewitt, Fundamentals of Heat and Mass Transfer, John Wiley & Sons, 1996.
  5. G. K. White, Experimental Technique in Low-temperature Physics, Clarendon Press, 1989.
  6. Thomas. M. Flynn, Ph. D and P. E, Cryogenic Engineering, Marcel Dekker, pp. 359-429, 1997
  7. SRDK-408D2 4 K-Cryocooler User Manual, [Online] available: http://www.shicryogenics.com.
  8. Temperature measurement and control catalog, [Online] available: http://www.lakeshore.com.

Cited by

  1. Development of an experimental variable temperature set-up for a temperature range from 2.2 K to 325 K for cost-effective temperature sensor calibration vol.28, pp.5, 2017, https://doi.org/10.1088/1361-6501/aa6464
  2. A New Outer-Rotor Hybrid-Excited Flux-Switching Machine Employing the HTS Homopolar Topology vol.12, pp.14, 2012, https://doi.org/10.3390/en12142654
  3. A Flux-Controllable NI HTS Flux-Switching Machine for Electric Vehicle Applications vol.10, pp.5, 2020, https://doi.org/10.3390/app10051564