QPS 프로그램에서 Myocardial extract 적용에 따른 임상적 유용성 평가

The Evaluation of Clinical Usefulness on Application of Myocardial Extract in Quantitative Perfusion SPECT

  • 윤종준 (부산대학교병원 핵의학과) ;
  • 임영현 (부산대학교병원 핵의학과) ;
  • 이무석 (부산대학교병원 핵의학과) ;
  • 송현석 (부산대학교병원 핵의학과) ;
  • 정지욱 (부산대학교병원 핵의학과) ;
  • 박세윤 (부산대학교병원 핵의학과) ;
  • 김재환 (부산대학교병원 핵의학과) ;
  • 김정욱 (부산대학교병원 핵의학과)
  • Yun, Jong-Jun (Department of nuclear medicine, Pusan National University Hospital) ;
  • Lim, Yeong-Hyeon (Department of nuclear medicine, Pusan National University Hospital) ;
  • Lee, Mu-Seok (Department of nuclear medicine, Pusan National University Hospital) ;
  • Song, Hyeon-Seok (Department of nuclear medicine, Pusan National University Hospital) ;
  • Jeong, Ji-Uk (Department of nuclear medicine, Pusan National University Hospital) ;
  • Park, Se-Yun (Department of nuclear medicine, Pusan National University Hospital) ;
  • Kim, Jae-Hwan (Department of nuclear medicine, Pusan National University Hospital) ;
  • Kim, Jeong-Uk (Department of nuclear medicine, Pusan National University Hospital)
  • 투고 : 2011.06.28
  • 심사 : 2011.07.05
  • 발행 : 2011.11.05

초록

게이트 심근관류 SPECT의 정량적 평가를 위해 사용되는 자동화된 정량화 소프트웨어(AutoQUANT)는 획득 자료 처리과정에 따라 측정치 오차가 발생할 것으로 예상된다. 동일한 환자를 대상으로 다른 조건의 변화 없이 QPS (Quantitative Perfusion SPECT) 프로그램에서 Myocardial extract 전후에 따른 정량적 지표를 구하여 평가하고자 한다. 2010년 6월부터 2010년 12월까지 부산대학교병원 핵의학과에서 Tc-99m MIBI 게이트 심근관류 SPECT를 시행한 환자 중 51명(남:여=34:17, 평균나이 $66.45{\pm}9.92$)을 대상으로 하였다. 심근의 단축단면상을 이용하여 원형 관심영역은 배후영역을 제외한 심근영역에서 설정한 자료를 획득하였다. AutoQUANT를 이용한 QPS 프로그램에서 일과성허혈확장(Transient Ischemic Dilation, TID), 부하기 관류결손 범위율의 합(Extent of Stress Defect, ESD)과 부하기 관류점수 합(Summed Stress Score, SSS)을 구하여 비교 분석하였다. QPS 프로그램에서 Myocardial extract 적용 전과 적용 후 정량적 지표의 상관계수는 TID (r=0.98), ESD (r=0.99), SSS (r=0.99) 모두 좋은 상관관계를 보였다. 95% 신뢰수준에서 TID (p=0.78), ESD (p=0.31), SSS (p=0.19) 모두 통계적으로 유의한 차이가 없었다. 육안분석에서는 판독에 영향을 줄 수 있을 만한 영상의 차이는 발견할 수 없었다. QPS 프로그램에서 Myocardial extract 적용 전과 적용 후 정량적 지표(TID, ESD, SSS)는 모두 좋은 상관관계가 있다는 것을 확인할 수 있었고, 통계적으로 유의한 차이를 보이지 않았다. Myocardial extract 영상처리기법이 비교적 간단하여 심근관류 SPECT 영상의 질을 올릴 수 있을 것으로 사료된다.

Purpose: As to analytical method of data, the AutoQUANT software in which it is used quantitative rating of the myocardial perfusion SPECT are reported that there is a difference. Therefore the measured value error of the mutual program is expected to be generated even if the quantitative analysis is made data of the same patient. The purpose of this study is to offer the comparative analysis of myocardial extract method in Quantitative Perfusion SPECT. Materials and methods: We analyzed the 51 patients who were examined by Tc-99m MIBI gated myocardial SPECT in nuclear medicine department of Pusan National University Hospital from June to December 2010(34 men, 17 women, mean age $66.5{\pm}9.9$). We acquired the extracted image in myocardial extract protocol. QPS program that uses the AutoQUANT software measured TID(Transient Ischemic Dilation), ESD(Extent of Stress Defect), SSS(Summed Stress Score). Then analyzed the results. Results: The correlation of appyling myocardial extract is TID(r=0.98), ESD(r=0.99), SSS(r=0.99). In the 95% confidence limit, there was no satistically significant difference(TID p=0.78, ESD p=0.31, SSS p=0.19). After blinding test with a physician for making a qualitative analysis, there was no difference. Conclusion: Quantitative indices in QPS program showed good correlation and the results showed no statistically signigicant difference. The variance between method was small. therefore, the functional parameters by each method can be used interchangeably. Also, we expect patient's satisfaction.