Impact of Lumen Size and Helical Coil Place Change in Spring on Orthodontic Force

Spring의 lumen size와 helical coil 형성 위치 변화가 교정력에 미치는 영향

  • Lee, Gyu-Sun (Department of Dental Technology, Dongnam Health College) ;
  • Lee, Sun-Kyoung (Department of Preventive and Social Dentistry & Institute of Oral Biology, College of Dentistry, Kyung Hee University) ;
  • Kim, Bok-Dong (Jin Sung IND. CO., LTD.)
  • 이규선 (동남보건대학교 치기공과) ;
  • 이선경 (경희대학교 치과대학 예방사회치과학교실 및 구강생물연구소) ;
  • 김복동 (진성기업(주) 특수선사업부)
  • Received : 2011.10.31
  • Accepted : 2011.12.28
  • Published : 2011.12.30

Abstract

Purpose: The purpose of this study was to impact of force system change in finger spring that add helical coil one round on orthodontic force. Methods: The following conclusions were drawn from the experiment conducted after bending 90 samples with a CNC wire forming machine while changing the height and lumen size to 1mm - 3mm - 5mm and 2mm - 3mm - 4mm respectively in the coil of the force system in finger spring added with one wheel of helical coil of 18-8 stainless steel round wire (${\Phi}0.5mm$, spring hard) from Jinsung Co. in domestic market under the following conditions: Laboratory name = Instron 5942; Temperature($deg^{\circ}C$) = 18.00; Humidity(%) = 50.00; Rate 1 = 10.00000 mm/min; Compressive extension = 5.0mm. Results: When Coil height is 1, 3, 5mm and lumen size is 2, 3, 4mm reduce finger spring as mean value of compressive extension occasion maximum load(mN) increases as coil height rises, and lumen size grows to 5.0mm. And was expose that compressive load(mN) increases as coil position of finger spring rises and increase as lumen size is decrescent. Conclusion: As the adherence height of coil was raised from 1mm through 3mm to 5mm, compressive load increased. As the lumen size increased from 2mm through 3mm to 4mm, compressive load decreased. Therefore, these results suggest that it is desirable to lower the coil height and enlarge the lumen size to enhance the biomechanical efficiency of finger spring when manufacturing the finger spring for removable orthodontic devices.

Keywords

References

  1. 강창은. 교정용 closed coil spring의 기계적 성질에 관한 비교 연구. 원광대학교 대학원, 1998.
  2. 최광철, 김경호, 박영철. 교정용 loop 스프링의 force system에 영향을 주는 요소. 대한치과교정학회지, 29(5), 511-519, 1999.
  3. Born HS. some facts concerning the open coil spring. Am J Orthod, 41, 917-925, 1955. https://doi.org/10.1016/0002-9416(55)90128-7
  4. Boshart BF. Load-deflection rate measurements of activated open and closed coil spring. Angle Orthod, 60, 27-32, 1989.
  5. Chanconas SJ, Caputo AA, Harvey K. Orthodontic force characteristics of open coil spring. Am J Orthod, 85, 494-497, 1984. https://doi.org/10.1016/0002-9416(84)90089-7
  6. Webb RI, Caputo AA, Chaconas SJ. Orthodontic force producation by closed coil spring. Am J Orthod, 74, 405-409, 1978. https://doi.org/10.1016/0002-9416(78)90062-3