Development of Prepolarization Coil Current Driver in SQUID Sensor-based Ultra Low-field Magnetic Resonance Apparatuses

SQUID 센서 기반의 극저자장 자기공명 장치를 위한 사전자화코일 전류구동장치 개발

  • Hwang, S.M. (Korea Research Institute of Standards and Science) ;
  • Kim, K. (Korea Research Institute of Standards and Science) ;
  • Kang, C.S. (Korea Research Institute of Standards and Science) ;
  • Lee, S.J. (Korea Research Institute of Standards and Science) ;
  • Lee, Y.H. (Korea Research Institute of Standards and Science)
  • Received : 2011.11.01
  • Accepted : 2011.11.25
  • Published : 2011.12.31

Abstract

SQUID sensor-based ultra low-field magnetic resonance apparatus with ${\mu}T$-level measurement field requires a strong prepolarization magnetic field ($B_p$) to magnetize its sample and obtain magnetic resonance signal with a high signal-to-noise ratio. This $B_p$ needs to be ramped down very quickly so that it does not interfere with signal acquisition which must take place before the sample magnetization relaxes off. A MOSFET switch-based $B_p$ coil driver has current ramp-down time ($t_{rd}$) that increases with $B_p$ current, which makes it unsuitable for driving high-field $B_p$ coil made of superconducting material. An energy cycling-type current driver has been developed for such a coil. This driver contains a storage capacitor inside a switch in IGBT-diode bridge configuration, which can manipulate how the capacitor is connected between the $B_p$ coil and its current source. The implemented circuit with 1.2 kV-tolerant devices was capable of driving 32 A current into a thick copper-wire solenoid $B_p$ coil with a 182 mm inner diameter, 0.23 H inductance, and 5.4 mT/A magnetic field-to-current ratio. The measured trd was 7.6 ms with a 160 ${\mu}F$ storage capacitor. trd was dependent only on the inductance of the coil and the capacitance of the driver capacitor. This driver is scalable to significantly higher current of superconducting $B_p$ coils without the $t_{rd}$ becoming unacceptably long with higher $B_p$ current.

Keywords

References

  1. John Clarke, Michael Hatridge, and Michael Mossle, "SQUID-Detected Magnetic Resonance Imaging in Microtesla Fields", Annu. Rev. Biomed. Eng. 9, 389- 413 (2007).
  2. Seung Kyun Lee, Michael Mossle, Whittier Myers, Nathan Kelso, Andreas H. Trabesinger, Alexander Pines, and John Clarke, "SQUID-Detected MRI at $132{\mu}T$ with T1-Weighted Contrast Established at $10{\mu}T$-300 mT", Magn. Reson. in Med. 53, 9-14 (2005).
  3. Robert McDermot, SeungKyun Lee, Bennie ten Haken, Andreas H. Trabesinger, Alexander Pines, and John Clarke, "Microtesla MRI with a superconducting quantum interference device", Proc. Natl. Acad. Sci. U.S.A. 101(21), 7857-7861 (2004).
  4. Vadim S. Zotev, Andrei N. Matlashov, Petr L. Volegov, Algis V. Urbaitis, Michelle A. Espy, and Robert H. Kraus, Jr ,"SQUID-based instrumentation for ultra-low-field MRI", Supercond. Sci. Tech, 20(11) S367 (2007).
  5. Nathaniel I. Matter, Greig C. Scott, Thomas Grafendorfer, Albert Macovski, and Steven M. Conolly, "Rapid Polarizing Field Cycling in Magnetic Resonance Imaging", IEEE Trans. Med. Imag. 25(1), 84-93 (2006).
  6. Martin N. Wilson, "NbTi superconductors with low ac loss: A review", Cryogenics 48, 381-395 (2008).