DOI QR코드

DOI QR Code

Analysis of changes in dose distribution due to respiration during IMRT

  • Shin, Jung-Suk (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Shin, Eun-Hyuk (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Han, Young-Yih (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Ju, Sang-Gyu (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Kim, Jin-Sung (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Ahn, Sung-Hwan (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Kim, Tae-Gyu (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Jeong, Bae-Kwon (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Park, Hee-Chul (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Ahn, Young-Chan (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Choi, Doo-Ho (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine)
  • Received : 2011.07.04
  • Accepted : 2011.08.31
  • Published : 2011.09.30

Abstract

Purpose: Intensity modulated radiation therapy (IMRT) is a high precision therapy technique that can achieve a conformal dose distribution on a given target. However, organ motion induced by respiration can result in significant dosimetric error. Therefore, this study explores the dosimetric error that result from various patterns of respiration. Materials and Methods: Experiments were designed to deliver a treatment plan made for a real patient to an in-house developed motion phantom. The motion pattern; the amplitude and period as well as inhale-exhale period, could be controlled by in-house developed software. Dose distribution was measured using EDR2 film and analysis was performed by RIT113 software. Three respiratory patterns were generated for the purpose of this study; first the 'even inhale-exhale pattern', second the slightly long exhale pattern (0.35 seconds longer than inhale period) named 'general signal pattern', and third a 'long exhale pattern' (0.7 seconds longer than inhale period). One dimensional dose profile comparisons and gamma index analysis on 2 dimensions were performed. Results: In one-dimensional dose profile comparisons, 5% in the target and 30% dose difference at the boundary were observed in the long exhale pattern. The center of high dose region in the profile was shifted 1 mm to inhale (caudal) direction for the 'even inhale-exhale pattern', 2 mm and 5 mm shifts to exhale (cranial) direction were observed for 'slightly long exhale pattern' and 'long exhale pattern', respectively. The areas of gamma index >1 were 11.88 %, 15.11%, and 24.33% for 'even inhale-exhale pattern', 'general pattern', and 'long exhale pattern', respectively. The long exhale pattern showed largest errors. Conclusion: To reduce the dosimetric error due to respiratory motions, controlling patient's breathing to be closer to even inhaleexhale period is helpful with minimizing the motion amplitude.

Keywords

References

  1. Hogle WP. The state of the art in radiation therapy. Semin Oncol Nurs 2006;22:212-20. https://doi.org/10.1016/j.soncn.2006.07.004
  2. Galvin JM, Chen XG, Smith RM. Combining multileaf fi elds to modulate fluence distributions. Int J Radiat Oncol Biol Phys 1993;27:697-705. https://doi.org/10.1016/0360-3016(93)90399-G
  3. Mott JH, Hounsell AR, Budgell GJ, Wilkinson JM, Williams PC. Customised compensation using intensity modulated beams delivered by dynamic multileaf collimation. Radiother Oncol 1999;53:59-65. https://doi.org/10.1016/S0167-8140(99)00125-5
  4. Chang SX, Cullip TJ, Deschesne KM. Intensity modulation delivery techniques: "step & shoot" MLC auto-sequence versus the use of a modulator. Med Phys 2000;27:948-59. https://doi.org/10.1118/1.598989
  5. Keall PJ, Mageras GS, Balter JM, et al. The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Med Phys 2006;33:3874-900. https://doi.org/10.1118/1.2349696
  6. Li XA, Keall PJ, Orton CG. Point/counterpoint: respiratory gating for radiation therapy is not ready for prime time. Med Phys 2007;34:867-70. https://doi.org/10.1118/1.2514027
  7. Zygmanski P, Kung JH, Jiang SB, Choi N, Chen G. Is IMRT of lung possible without respiratory gating? Clinical evaluation of a true dose received by a moving target volume. Int J Radiat Oncol Biol Phys 2001;51(3 suppl 1):27. https://doi.org/10.1016/S0360-3016(01)01871-5
  8. Kung JH, Zygmanski P, Kooy H, Chen G. Dose error in IMRT treatment of lung cancer without respiratory gating. Int J Radiat Oncol Biol Phys 2000;48(3 suppl 1):344
  9. Li X, Zhang P, Helminen H, Gewanter R, Kutcher GJ. Comparison of different respiratory motion compensation methods for lung cancer IMRT treatment. Int J Radiat Oncol Biol Phys 2005;63 Suppl 1:S564. https://doi.org/10.1016/j.ijrobp.2005.07.949
  10. Duan J, Shen S, Fiveash JB, Popple RA, Brezovich IA. Dosimetric and radiobiological impact of dose fractionation on respiratory motion induced IMRT delivery errors: a volumetric dose measurement study. Med Phys 2006;33:1380-7. https://doi.org/10.1118/1.2192908
  11. Lujan AE, Larsen EW, Balter JM, Ten Haken RK. A method for incorporating organ motion due to breathing into 3D dose calculations. Med Phys 1999;26:715-20. https://doi.org/10.1118/1.598577
  12. Shin E, Park HC, Han Y, Ju SG, Shin JS, Ahn YC. Effi cacy of a respiratory training system on the regularity of breathing. J Korean Soc Ther Radiol Oncol 2008;26:181-8. https://doi.org/10.3857/jkstro.2008.26.3.181
  13. Cuijpers JP, Underberg RW, Lagerwaard FJ, Senan S, Sletman BJ. Penumbra broadening due to respiratory motion: an analysis of 4D clinical datasets. Radiat Oncol 2004;73(Suppl 1):S128.
  14. Bortfeld T, Jokivarsi K, Goitein M, Kung J, Jiang SB. Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation. Phys Med Biol 2002;47:2203-20. https://doi.org/10.1088/0031-9155/47/13/302

Cited by

  1. Dosimetric evaluation of a moving tumor target in intensity-modulated radiation therapy (IMRT) for lung cancer patients vol.63, pp.1, 2013, https://doi.org/10.3938/jkps.63.67
  2. Evaluation of radiation dose to organs during kilovoltage cone‐beam computed tomography using Monte Carlo simulation vol.15, pp.2, 2011, https://doi.org/10.1120/jacmp.v15i2.4556
  3. Optimization of beam orientation and virtual organ delineation for lung IMRT vol.64, pp.7, 2014, https://doi.org/10.3938/jkps.64.1047
  4. Combination effects of tissue heterogeneity and geometric targeting error in stereotactic body radiotherapy for lung cancer using CyberKnife vol.16, pp.5, 2011, https://doi.org/10.1120/jacmp.v16i5.5397
  5. Feasibility study of the neutron dose for real-time image-guided proton therapy: A Monte Carlo study vol.67, pp.1, 2015, https://doi.org/10.3938/jkps.67.142
  6. Secondary Neutron Dose Measurement for Proton Line Scanning Therapy vol.27, pp.3, 2011, https://doi.org/10.14316/pmp.2016.27.3.162
  7. Monte Carlo simulation of secondary neutron dose for scanning proton therapy using FLUKA vol.12, pp.10, 2011, https://doi.org/10.1371/journal.pone.0186544
  8. Dosimetric analysis on the effect of tumor motion in IMRT for liver cancer: comparison of TomoTherapy and VMAT using the Delta4 Hexa-Motion system vol.79, pp.5, 2011, https://doi.org/10.1007/s40042-021-00150-x