단층촬영영상을 이용한 T.O.D Calibration의 정확성과 유용성에 관한 비교연구

Comparative Study on Accuracy and Usefulness of Calibration Using CT T.O.D

  • 서정범 (건국대학교병원 영상의학과) ;
  • 김동현 (건국대학교병원 영상의학과) ;
  • 이정범 (건국대학교병원 영상의학과)
  • Seo, Jeong-Beom (Department of Radiology, Konkuk University Medical Center) ;
  • Kim, Dong-Hyeon (Department of Radiology, Konkuk University Medical Center) ;
  • Lee, Jeong-Beom (Department of Radiology, Konkuk University Medical Center)
  • 투고 : 2011.01.26
  • 심사 : 2011.03.07
  • 발행 : 2011.03.31

초록

Uses a Tomographic scan image and Table Object Distance(TOD) price after measuring, uses accuracy and usability of blood vessel diameter(Vessel Diameter) measurement under comparison evaluating boil TOD Calibration. The patient who enforces Prosecuting Attorney abdomen Tomographic scan in the object the superior mesentery artery uses PACS View from abdomen fault image and from blood vessel diameter and the table measures the height until of the blood vessel. Uses Angio Catheter from Angiography(5 Fr.) and enforces is measured from PACS View the height until of the table which and the blood vessel at TOD Calibration price and the size of the superior mesentery artery inputs measures an superior mesentery artery building skill. Catheter Calibration input Agnio Catheter where uses in Angiography the size of the superior mesentery artery at Catheter Calibration price and they measure. Produced an accuracy from monitoring data and comparison evaluated. The statistical program used SPSS. TOD Calibration accuracy was 96.53%, standard deviation is 0.03829. Catheter Calibration accuracy of 92.91%, standard deviation is 0.05085. Represents a statistically significant difference(p = 0). According to age and gender was not statistically significant(p > 0.05). TOD Calibration correlation coefficient R-squared of 88.8%, Catheter Calibration of the R-squared is 75.5%. High accuracy of both methods. Through this study, CT images using the measured distance between the table and the Object, TOD Calibration accuracy higher than two Catheter Calibration was measured. TOD and Catheter Calibration represents a statistically significant difference(p = 0).

키워드