The Relationship of Radiation Dose and Image Quality According to the Condition of Chest PA

  • Son, Jin-Hyun (Department of Radiological Technology, Shingu University) ;
  • Min, Jung-Whan (Department of Radiological Technology, Shingu University) ;
  • Kang, Byung-Sam (Department of Radiological Technology, Shingu University) ;
  • Dong, Kyung-Rae (Department of Radiological Technology, Gwangju Health College University)
  • Received : 2011.09.27
  • Accepted : 2011.11.12
  • Published : 2011.12.31

Abstract

The purpose of this study is to compare the measurement result of radiation dose by using standard thoracic phantom and ionization chamber to advice proposal in the shooting condition of chest PA projection at hospitals recently. And to understand the change between radiation dose and resolution in different conditions. The period this study was from August 2010 to September 2010 and the subjects of the study was 3 general hospitals, 4 personal hospitals and 1 laboratory at the college. Finally we study with 6 DR, 1 CR, and 4 F/S equipments. Most hospitals met advice proposal, but some of the hospitals exceed advice dose from the result of our study. We can lower radiation dose about 25% when kVp is lowered about 20% in DR equipment. And we can lower radiation dose about 50% when mAs is lowered about 35%. The image quality was similar to the original in the study. Most hospitals which exceed advice dose were personal hospitals. The reason why it happened is that radiation dose for chest PA projection at personal hospitals is higher than general hospitals and the personal hospitals' equipments are older than general hospitals' equipments. We guess that patients' radiation dose of chest PA projection can be lowered from the result.

Keywords