DOI QR코드

DOI QR Code

Fenestrated Stent Graft Repair of Abdominal Aortic Aneurysm: Hemodynamic Analysis of the Effect of Fenestrated Stents on the Renal Arteries

  • Sun, Zhonghua (Discipline of Medical Imaging, Department of Imaging and Applied Physics, Curtin University of Technology) ;
  • Chaichana, Thanapong (Discipline of Medical Imaging, Department of Imaging and Applied Physics, Curtin University of Technology)
  • 투고 : 2009.05.19
  • 심사 : 2009.07.31
  • 발행 : 2010.02.01

초록

Objective: We wanted to investigate the hemodynamic effect of fenestrated stents on the renal arteries with using a fluid structure interaction method. Materials and Methods: Two representative patients who each had abdominal aortic aneurysm that was treated with fenestrated stent grafts were selected for the study. 3D realistic aorta models for the main artery branches and aneurysm were generated based on the multislice CT scans from two patients with different aortic geometries. The simulated fenestrated stents were designed and modelled based on the 3D intraluminal appearance, and these were placed inside the renal artery with an intra-aortic protrusion of 5.0-7.0 mm to reflect the actual patients' treatment. The stent wire thickness was simulated with a diameter of 0.4 mm and hemodynamic analysis was performed at different cardiac cycles. Results: Our results showed that the effect of the fenestrated stent wires on the renal blood flow was minimal because the flow velocity was not significantly affected when compared to that calculated at pre-stent graft implantation, and this was despite the presence of recirculation patterns at the proximal part of the renal arteries. The wall pressure was found to be significantly decreased after fenestration, yet no significant change of the wall shear stress was noticed at post-fenestration, although the wall shear stress was shown to decrease slightly at the proximal aneurysm necks. Conclusion: Our analysis demonstrates that the hemodynamic effect of fenestrated renal stents on the renal arteries is insignificant. Further studies are needed to investigate the effect of different lengths of stent protrusion with variable stent thicknesses on the renal blood flow, and this is valuable for understanding the long-term outcomes of fenestrated repair.

키워드

참고문헌

  1. Greenhalgh RM, Brown LC, Kwong GP, Powell JT, Thompson SG; EVAR trial participants. Comparison of endovascular aneurysm repair with open repair in patients with abdominal aortic aneurysm (EVAR trial 1), 30-day operative mortality results: randomised controlled trial. Lancet 2004;364:843-848 https://doi.org/10.1016/S0140-6736(04)16979-1
  2. Prinssen M, Verhoeven EL, Buth J, Cuypers PW, van Sambeek MR, Balm R, et al. A randomized trial comparing conventional and endovascular repair of abdominal aortic aneurysms. N Engl J Med 2004;351:1607-1618 https://doi.org/10.1056/NEJMoa042002
  3. Parodi JC, Palmaz JC, Barone HD. Transfemoral intraluminal graft implantation for abdominal aortic aneurysms. Ann Vasc Surg 1991;5:491-499 https://doi.org/10.1007/BF02015271
  4. Buth J, van Marrewijk CJ, Harris PL, Hop WC, Riambau V, Laheij RJ; EUROSTSR Collaborators. Outcome of endovascular abdominal aortic aneurysm repair in patients with conditions considered unfit for an open procedure: a report on the EUROSTAR experience. J Vasc Surg 2002;35:211-221 https://doi.org/10.1067/mva.2002.121050
  5. Lobato AC, Quick RC, Vaughn PL, Rodriguez-Lopez J, Douglas M, Diethrich EB. Transrenal fixation of aortic endografts: intermediate follow-up of a single-center experience. J Endovasc Ther 2000;7:273-278 https://doi.org/10.1583/1545-1550(2000)007<0273:TFOAEI>2.3.CO;2
  6. Bove PG, Long GW, Zelenock GB, Bendick PJ, Khoury MD, Burr MO, et al. Transrenal fixation of aortic stent-grafts for the treatment of infrarenal aortic aneurysmal disease. J Vasc Surg 2000;32:697-703 https://doi.org/10.1067/mva.2000.110051
  7. Green RM. Patient selection for endovascular abdominal aortic aneurysm repair. J Am Coll Surg 2002;194:S67-S73 https://doi.org/10.1016/S1072-7515(01)01080-8
  8. Dillavou ED, Muluk SC, Rhee RY, Tzeng E, Woody JD, Gupta N, et al. Does hostile neck anatomy preclude successful endovascular aortic aneurysm repair? J Vasc Surg 2003;38:657-663 https://doi.org/10.1016/S0741-5214(03)00738-9
  9. Browne TF, Hartley D, Purchas S, Rosenberg M, Van Schie G, Lawrence-Brown M. A fenestrated covered suprarenal aortic stent. Eur J Vasc Endovasc Surg 1999;18:445-449 https://doi.org/10.1053/ejvs.1999.0924
  10. Faruqi RM, Chuter TA, Reilly LM, Sawhney R, Wall S, Canto C, et al. Endovascular repair of abdominal aortic aneurysm using a pararenal fenestrated stent-graft. J Endovasc Surg 1999;6:354-358 https://doi.org/10.1583/1074-6218(1999)006<0354:EROAAA>2.0.CO;2
  11. Stanley BM, Semmens JB, Lawrence-Brown MM, Goodman MA, Hartley DE. Fenestration in endovascular grafts for aortic aneurysm repair: new horizons for preserving blood flow in branch vessels. J Endovasc Ther 2001;8:16-24 https://doi.org/10.1583/1545-1550(2001)008<0016:FIEGFA>2.0.CO;2
  12. Anderson JL, Berce M, Hartley DE. Endoluminal aortic grafting with renal and superior mesenteric artery incorporation by graft fenestration. J Endovasc Ther 2001;8:3-15 https://doi.org/10.1583/1545-1550(2001)008<0003:EAGWRA>2.0.CO;2
  13. Sun Z, Mwipatayi BP, Semmens JB, Lawrence-Brown MM. Short to midterm outcomes of fenestrated endovascular grafts in the treatment of abdominal aortic aneurysms: a systematic review. J Endovasc Ther 2006;13:747-753 https://doi.org/10.1583/06-1919.1
  14. Semmens JB, Lawrence-Brown MM, Hartley DE, Allen YB, Green R, Nadkarni S. Outcomes of fenestrated endografts in the treatment of abdominal aortic aneurysm in Western Australia (1997-2004). J Endovasc Ther 2006;13:320-329 https://doi.org/10.1583/05-1686.1
  15. Sun Z, Allen YB, Nadkarni S, Knight R, Hartley DE, Lawrence-Brown MM. CT virtual intravascular endoscopy in the visualization of fenestrated stent-grafts. J Endovasc Ther 2008;15:42-51 https://doi.org/10.1583/07-2234.1
  16. Sun Z, Allen YB, Mwipatayi BP, Hartley DE, Lawrence-Brown MM. Multislice CT angiography in the follow-up of fenestrated endovascular grafts: effect of slice thickness on 2D and 3D visualization of the fenestration stents. J Endovasc Ther 2008;15:417-426 https://doi.org/10.1583/08-2432.1
  17. Richter GM, Palmaz JC, Noeldge G, Tio F. Relationship between blood flow, thrombus, and neointima in stents. J Vasc Interv Radiol 1999;10:598-604 https://doi.org/10.1016/S1051-0443(99)70090-4
  18. Beythien C, Gutensohn K, Bau J, Hamm CW, Kuhnl P, Meinertz T, et al. Influence of stent length and heparin coating on platelet activation: a flow cytometric analysis in a pulsed floating model. Thromb Res 1999;94:79-86 https://doi.org/10.1016/S0049-3848(98)00198-4
  19. Peacock J, Hankins S, Jones T, Lutz R. Flow instabilities induced by coronary artery stents: assessment with an in vitro pulse duplicator. J Biomech 1995;28:17-26 https://doi.org/10.1016/0021-9290(95)80003-4
  20. Sun Z, Winder RJ, Kelly BE, Ellis PK, Hirst DG. CT virtual intravascular endoscopy of abdominal aortic aneurysms treated with suprarenal endovascular stent grafting. Abdom Imaging 2003;28:580-587 https://doi.org/10.1007/s00261-002-0069-4
  21. Sun Z. 3D multislice CT angiography in post-aortic stent grafting: a pictorial essay. Korean J Radiol 2006;7:205-211 https://doi.org/10.3348/kjr.2006.7.3.205
  22. Sun Z, Mwipatayi BP, Allen YB, Hartley DE, Lawrence-Brown MM. Multislice CT angiography of fenestrated endovascular stent grafting for treating abdominal aortic aneurysms: a pictorial review of the 2D/3D visualizations. Korean J Radiol 2009;10:285-293 https://doi.org/10.3348/kjr.2009.10.3.285
  23. Frauenfelder T, Lotfey M, Boehm T, Wildermuth S. Computational fluid dynamics: hemodynamic changes in abdominal aortic aneurysm after stent-graft implantation. Cardiovasc Intervent Radiol 2006;29:613-623 https://doi.org/10.1007/s00270-005-0227-5
  24. Borghi A, Wood NB, Mohiaddin RH, Xu XY. Fluid-solid interaction simulation of flow and stress pattern in thoracoabdominal aneurysms: a patient-specific study. J Fluids Struct 2008;24:270-280 https://doi.org/10.1016/j.jfluidstructs.2007.08.005
  25. Li Z, Kleinstreuer C. Analysis of biomechanical factors affecting stent-graft migration in an abdominal aortic aneurysm model. J Biomech 2006;39:2264-2273 https://doi.org/10.1016/j.jbiomech.2005.07.010
  26. Chong CK, How TV. Flow patterns in an endovascular stentgraft for abdominal aortic aneurysm repair. J Biomech 2004;37:89-97 https://doi.org/10.1016/S0021-9290(03)00236-7
  27. Deplano V, Knapp Y, Bertrand E, Gaillard E. Flow behaviour in an asymmetric compliant experimental model for abdominal aortic aneurysm. J Biomech 2007;40:2406-2413 https://doi.org/10.1016/j.jbiomech.2006.11.017
  28. Fillinger MF, Raghavan ML, Marra SP, Cronenwett JL, Kennedy FE. In vivo analysis of mechanical wall stress and abdominal aortic aneurysm rupture risk. J Vasc Surg 2002;36:589-597 https://doi.org/10.1067/mva.2002.125478
  29. Li Z, Kleinstreuer C. Fluid-structure interaction effects on sacblood pressure and wall stress in a stented aneurysm. J Biomech Eng 2005;127:662-671 https://doi.org/10.1115/1.1934040
  30. Liffman K, Lawrence-Brown MM, Semmens JB, Sutalo ID, Bui A, White F, et al. Suprarenal fixation: effect on blood flow of an endoluminal stent wire across an arterial orifice. J Endovasc Ther 2003;10:260-274 https://doi.org/10.1583/1545-1550(2003)010<0260:SFEOBF>2.0.CO;2
  31. Carlier SG, van Damme LC, Blommerde CP, Wentzel JJ, van Langehove G, Verheye S, et al. Augmentation of wall shear stress inhibits neointimal hyperplasia after stent implantation: inhibition through reduction inflammation? Circulation 2003;107:2741-2746 https://doi.org/10.1161/01.CIR.0000066914.95878.6D
  32. Prati F, Di Mario C, Moussa I, Reimers B, Mallus MT, Parma A, et al. In-stent neointimal proliferation correlates with the amount of residual plaque burden outside the stent: an intravascular ultrasound study. Circulation 1999;99:1011-1014 https://doi.org/10.1161/01.CIR.99.8.1011
  33. Gaillard E, Bergeron P, Deplano V. Influence of wall compliance on hemodynamics in models of abdominal aortic aneurysm. J Endovasc Ther 2007;14:593-599 https://doi.org/10.1583/1545-1550(2007)14[593:IOWCOH]2.0.CO;2
  34. Li Z, Kleinstreuer C. Computational analysis of type II endoleaks in a stented abdominal aortic aneurysm model. J Biomech 2006;39:2573-2582 https://doi.org/10.1016/j.jbiomech.2005.09.002
  35. Lawrence-Brown MM, Sun Z, Semmens JB, Liffman K, Sutalo ID, Hartley DB. Type II endoleaks: when is intervention indicated and what is the index of suspicion for types I or III? J Endovasc Ther 2009;16:I106-I118

피인용 문헌

  1. Endovascular stent graft repair of abdominal aortic aneurysms: Current status and future directions vol.1, pp.1, 2010, https://doi.org/10.4329/wjr.v1.i1.63
  2. Computation of hemodynamics in the left coronary artery with variable angulations vol.44, pp.10, 2010, https://doi.org/10.1016/j.jbiomech.2011.04.033
  3. Computational Fluid Dynamics Analysis of the Effect of Plaques in the Left Coronary Artery vol.2012, pp.None, 2010, https://doi.org/10.1155/2012/504367
  4. A multimodality vascular imaging phantom of an abdominal aortic aneurysm with a visible thrombus. vol.40, pp.6, 2013, https://doi.org/10.1118/1.4803497
  5. Free Tools and Strategies for the Generation of 3D Finite Element Meshes: Modeling of the Cardiac Structures vol.2013, pp.None, 2010, https://doi.org/10.1155/2013/540571
  6. Impact of plaques in the left coronary artery on wall shear stress and pressure gradient in coronary side branches vol.17, pp.2, 2014, https://doi.org/10.1080/10255842.2012.671308
  7. Delaunay Tetrahedralization of the Heart Based on Integration of Open Source Codes vol.490, pp.1, 2010, https://doi.org/10.1088/1742-6596/490/1/012131
  8. Commentary: Computational Modeling of Contemporary Stent-Grafts vol.22, pp.4, 2010, https://doi.org/10.1177/1526602815590370
  9. Does Lower Limb Exercise Worsen Renal Artery Hemodynamics in Patients with Abdominal Aortic Aneurysm? vol.10, pp.5, 2010, https://doi.org/10.1371/journal.pone.0125121
  10. Hemodynamic Functions of Fenestrated Stent Graft under Resting, Hypertension, and Exercise Conditions vol.3, pp.None, 2010, https://doi.org/10.3389/fsurg.2016.00035
  11. Effect of a Flared Renal Stent on the Performance of Fenestrated Stent-Grafts at Rest and Exercise Conditions vol.23, pp.5, 2010, https://doi.org/10.1177/1526602816651425
  12. Hemodynamic Performance of Multilayer Stents in the Treatment of Aneurysms with a Branch Attached vol.9, pp.None, 2010, https://doi.org/10.1038/s41598-019-46714-7
  13. The added value of CT virtual angioscopy to MDCT angiography in the evaluation of aortic diseases vol.45, pp.8, 2010, https://doi.org/10.1007/s00261-020-02607-2
  14. Review of the Development of Hemodynamic Modeling Techniques to Capture Flow Behavior in Arteries Affected by Aneurysm, Atherosclerosis, and Stenting vol.144, pp.4, 2022, https://doi.org/10.1115/1.4053082